The purpose of this study was to optimize and validate a multi-contrast, multi-echo fMRI method using a combined spin- and gradient-echo (SAGE) acquisition. It was hypothesized that SAGE-based blood oxygen level-dependent (BOLD) functional MRI (fMRI) will improve sensitivity and spatial specificity while reducing signal dropout. SAGE-fMRI data were acquired with five echoes (2 gradient-echoes, 2 asymmetric spin-echoes, and 1 spin-echo) across 12 protocols with varying acceleration factors, and temporal SNR (tSNR) was assessed. The optimized protocol was then implemented in working memory and vision tasks in 15 healthy subjects. Task-based analysis was performed using individual echoes, quantitative dynamic relaxation times T2* and T2, and echo time-dependent weighted combinations of dynamic signals. These methods were compared to determine the optimal analysis method for SAGE-fMRI. Implementation of a multiband factor of 2 and sensitivity encoding (SENSE) factor of 2.5 yielded adequate spatiotemporal resolution while minimizing artifacts and loss in tSNR. Higher BOLD contrast-to-noise ratio (CNR) and tSNR were observed for SAGE-fMRI relative to single-echo fMRI, especially in regions with large susceptibility effects and for T2-dominant analyses. Using a working memory task, the extent of activation was highest with T2*-weighting, while smaller clusters were observed with quantitative T2* and T2. SAGE-fMRI couples the high BOLD sensitivity from multi-gradient-echo acquisitions with improved spatial localization from spin-echo acquisitions, providing two contrasts for analysis. SAGE-fMRI provides substantial advantages, including improving CNR and tSNR for more accurate analysis.

Functional MRI (fMRI) is widely used to map synchronous fluctuations in brain activity via sensitivity to the blood oxygen level-dependent (BOLD) effect. The most common sequence used for fMRI leverages a gradient-echo (GRE) acquisition with echo planar imaging (EPI) readout, which can provide full brain coverage, adequate signal-to-noise ratio (SNR), good temporal resolution, and high sensitivity to BOLD T2* relaxation effects. The drawbacks of standard fMRI methods include susceptibility-induced signal dropouts near air-tissue interfaces, sensitivity to large draining vessels, and sub-optimal T2* sensitivity across the brain. This has led to the development of both multi-gradient-echo (MGE) fMRI (Cohen, Jagra, et al., 2021; Cohen et al., 2018; Evans et al., 2015; Kundu et al., 2013; Poser et al., 2006; Posse et al., 1999; Speck & Hennig, 1998) and spin-echo (SE) fMRI methods (Binney et al., 2010; Embleton et al., 2010; Glielmi et al., 2010; Lee et al., 1999), as well as ultra-high-resolution fMRI methods (Knudsen et al., 2023).

MGE-fMRI acquisitions have become more widely adopted in studies of task-based activation and functional connectivity in recent years. MGE-fMRI methods overcome many of the challenges associated with single GRE-EPI. In particular, they are less sensitive to susceptibility effects due to the inclusion of shorter echo times (TE) (Kundu et al., 2013). Additionally, given that the relaxation time T2* varies greatly across the brain (Hagberg et al., 2002) and the BOLD contrast is maximized when the TE equals the local tissue T2*, acquiring multiple TEs increases BOLD sensitivity (Poser et al., 2006). Moreover, further signal-to-noise (SNR) benefits are gained by optimally combining echo signals (Kundu et al., 2013). Similarly to a single GRE acquisition, MGE-fMRI is sensitive to large draining vessels (Boxerman et al., 1995). Because the BOLD-associated metabolic exchange is localized in the microvasculature, GRE acquisitions are limited in their spatial specificity. A multi-contrast acquisition could be a logical extension of multi-echo fMRI, pairing the benefits of MGE-fMRI with spatial specificity.

Specifically, spin-echo (SE) EPI could provide a complementary analysis to MGE-fMRI given its sensitivity to the microvasculature (Boxerman et al., 1995). Due to the reduced sensitivity to macrovascular signal contributions that are downstream of neuronal activation, SE methods may improve spatial localization of fMRI activation. In addition, SE acquisitions allow for refocusing of susceptibility-induced signal dropouts, which may be particularly advantageous for characterizing activation in regions near susceptibility interfaces such as inferior frontal and temporal lobes (Norris et al., 2002). SE methods are limited by lower BOLD sensitivity and thus suffer from diminished contrast-to-noise ratio (CNR) (Parkes et al., 2005), in large part preventing SE-fMRI from becoming the gold-standard. Therefore, a combined MGE and SE acquisition could provide complementary information that is useful for comprehensive functional analyses.

Combined MGE and SE acquisitions have been proposed for quantification of relaxation times (Jin et al., 2013; Küppers et al., 2022; Ni et al., 2017; Schmiedeskamp, Straka, Newbould, et al., 2012; Stokes et al., 2014), as well as for simultaneous assessment of cerebral blood flow and single-echo BOLD fMRI task-based activation (Glielmi et al., 2010). Additionally, T2-weighted gradient- and spin-echo (GRASE) acquisitions have been implemented in laminar fMRI (De Martino et al., 2013; Olman et al., 2012; Oshio & Feinberg, 1991). To our knowledge, a combined MGE and SE acquisition has not been evaluated for use in relaxation time quantification and multi-echo analysis for task-based BOLD fMRI. Presently, we propose the implementation of a combined spin- and gradient-echo (SAGE) sequence, with multiple echoes and contrasts, to unite the advantages of MGE- and SE-fMRI. The SAGE sequence has been previously developed and applied in perfusion MRI, most prominently in neuro-oncological and neuro-vascular pathologies (Schmiedeskamp, Straka, Newbould, et al., 2012; Schmiedeskamp et al., 2013; Skinner et al., 2014; Stokes et al., 2014). For perfusion MRI, SAGE provides advantages over standard single GRE sequences, including reduced sensitivity to T1 effects (due to contrast leakage in perfusion MRI (Stokes, Semmineh, et al., 2016)), optimal TEs for different tissues (Bell et al., 2017), and both macro- and microvascular sensitivity (Schmiedeskamp et al., 2013). Using the SAGE sequence, both relaxation rates R2* (=1/T2*) and R2 (=1/T2) can be quantified (Schmiedeskamp, Straka, & Bammer, 2012; Stokes & Quarles, 2016; Stokes et al., 2014). In the context of fMRI, quantitative R2* (and R2) may be more directly related to neuronal activation via the susceptibility differences between oxygenated and deoxygenated blood than a single T2*-weighted GRE given the ability to parse BOLD and non-BOLD components of a signal with multiple echoes (Buxton, 2013; Poser et al., 2006; Speck & Hennig, 1998). Specifically, quantitative R2* (and R2) may permit separation of the complex factors (both BOLD-based and non-BOLD-based, such as cardiac and respiratory noise and head motion) that contribute to fMRI signals (Kundu et al., 2012). Moreover, multi-echo and multi-contrast combinations using weighted summation strategies, which involve combination of signals across multiple TEs for optimized analysis, may provide further benefit for fMRI analysis (Poser et al., 2006; Posse et al., 1999). While multi-echo combinations offer many advantages, acquisitions require acceleration in order to meet the spatiotemporal resolution demands of specific research questions. Acceleration must be balanced with changes in signal-to-noise ratio, as well as ideal TEs based on tissue sensitivities.

The purpose of this study was two-fold: 1) to optimize the acquisition protocol by quantitatively assessing acceleration factors for SAGE in fMRI, as prior implementations have focused on perfusion MRI, and 2) validate SAGE-fMRI in task-based fMRI paradigms, comparing results to single-echo methods. For this purpose, we acquired dynamic SAGE-fMRI data with five echoes (2 GRE, 2 asymmetric spin-echoes (ASE), and 1 SE), first in 6 healthy subjects for optimization, then in 15 healthy subjects for validation using working memory and visual stimulus tasks. Single-echo analyses using the second and fifth echoes (i.e., analogous to standard GRE- and SE-fMRI, respectively) were compared to multi-echo SAGE analyses, which included quantitative dynamic T2* and T2 maps and relative relaxation-weighted dynamic signals. We hypothesized that SAGE-based BOLD fMRI (Schmiedeskamp et al., 2014) would improve sensitivity and CNR, while reducing signal dropout through the inclusion of multiple echoes and contrasts.

2.1 SAGE-fMRI theory

The standard SAGE sequence includes five echoes, including two GRE, two ASE, and one SE, as shown in Supplementary Materials Figure S1A. The SAGE signals at each TE depend on relaxation rates R2* and/or R2, as follows (Stokes et al., 2014):

[1]

where S0I and S0II represent the signal intensities following the excitation and refocusing pulses, respectively, and TESE is the TE of the SE. Signals before TESE2 are T2*-weighted, whereas signals after TESE2 are of mixed contrast (i.e., ASE TE3,4) or T2-weighted. The S0 terms vary due to slice profile mismatch and pulse imperfections (Schmiedeskamp, Straka, & Bammer, 2012). The GREs depend on R2*, the ASEs depend on both R2* and R2, and the SE depends only on R2.

For fMRI, SAGE data can be processed using either the individual echo signals (five echoes with signals S1 - S5) or multi-echo combinations. For multi-echo sequences, dynamic relaxation rates R2* and R2 can be quantified and utilized in subsequent fMRI analyses. Additionally, previous MGE studies have described a weighted multi-echo combination method; this approach has shown the benefit of weighting the signal by the measured relaxation rate, which is effectively the relative contribution to BOLD contrast (Poser et al., 2006; Posse et al., 1999). More specifically, the relaxation-weighting factors can be determined from the derivative of the signal by the specific relaxation rate. Relaxation-weighting produces two separate signals for analysis, one each for T2*-weighting and T2-weighting, by combining the five SAGE-fMRI signals with specific weighting factors. Thus, the relative relaxation-weighting factors for T2* and T2 (wT2* and wT2, respectively) are given by the derivative of Eq. 1:

[2]
[3]

where contributions to each weighting factor are determined by each TE’s contrast (i.e., T2*, T2, or mixed). For simplicity, the signal intensities (S0I and S0II) and normalization factors (Σnwx(TEn)), where x = T2* or T2, have been omitted from Eqs. 2 to 3. The voxel-wise weighting factors for each SAGE signal are shown in Supplementary Materials Figure S1B.

2.2.1 Data acquisition: SAGE-fMRI optimization

Dynamic SAGE EPI data were acquired in healthy subjects (n = 6, 33.0 ± 8.7 years old, 1 male) at 3T (Ingenia, Philips) with a 32-channel head coil. The protocol was approved by the local Institutional Review Board (IRB), and all participants provided written informed consent. T1-weighted (T1-w) anatomical images were acquired using a three-dimensional magnetization-prepared rapid acquisition gradient echo (MPRAGE) sequence with the following acquisition parameters: TR/TE: 7.0/3.2 ms; field of view: 256 × 256 mm2; voxel size: 1.0 × 1.0 × 1.0 mm3; sagittal slices: 176; and flip angle: 9°. Dynamic SAGE data were acquired with the following acquisition parameters: TR: 3.0 s, field of view: 240 × 240 mm2, voxel size: 3.0 × 3.0 × 3.0 mm3, and volumes: 40. A reverse phase encoding acquisition was collected for distortion correction. Acceleration factors sensitivity encoding (SENSE) and multiband (MB) influenced the TEs and number of axial slices acquired, while the partial Fourier (PF) factor was held relatively constant (Skinner et al., 2014) (Table 1). The SENSE factor was varied at 2, 2.5, and 3, and the MB factor was varied at 1 (no MB), 2, 3, and 4. For protocol comparisons, TR was held constant at 3.0 s.

Table 1.

Acquisition parameters across MB and SENSE factors.

MB factorSENSE factorTE1-5 (ms)SlicesZ coverage (mm)PF
8.4/27/52/70/89 24 72 0.74 
2.5 7.3/22/44/59/74 24 72 0.74 
6.4/19/36/49/61 24 72 0.74 
2 9.3/31/66/88/110 36 108 0.74 
2.5 8.0/27/59/78/97 36 108 0.74 
8.0/27/59/78/97 36 108 0.74 
8.9/31/66/88/110 36 108 0.74 
2.5 7.5/27/58/77/96 36 108 0.72 
7.4/26/57/76/95 36 108 0.71 
8.9/33/72/95/119 36 108 0.69 
2.5 8.2/30/67/90/112 36 108 0.71 
7.6/29/65/86/107 36 108 0.70 
MB factorSENSE factorTE1-5 (ms)SlicesZ coverage (mm)PF
8.4/27/52/70/89 24 72 0.74 
2.5 7.3/22/44/59/74 24 72 0.74 
6.4/19/36/49/61 24 72 0.74 
2 9.3/31/66/88/110 36 108 0.74 
2.5 8.0/27/59/78/97 36 108 0.74 
8.0/27/59/78/97 36 108 0.74 
8.9/31/66/88/110 36 108 0.74 
2.5 7.5/27/58/77/96 36 108 0.72 
7.4/26/57/76/95 36 108 0.71 
8.9/33/72/95/119 36 108 0.69 
2.5 8.2/30/67/90/112 36 108 0.71 
7.6/29/65/86/107 36 108 0.70 

The optimal sequence is in bolded text. While most MB factors ≥2 permit full brain coverage (≥120 mm), the number of slices was limited to 36 for comparisons across acceleration factors. Wherever possible, all factors outside of MB and SENSE were held constant for consistency. PF = Partial Fourier.

2.2.2 Data acquisition: SAGE-fMRI validation

Data were acquired in healthy subjects (n = 15, 24.4 ± 2.6 years old, 5 males) at 3T (Ingenia, Philips) with a 32-channel head coil. An estimate of minimum sample size necessary to achieve statistical significance (α = 0.05, power = 0.80, effect size = 0.80) was performed a priori using G*Power (Faul et al., 2007) and implemented presently. The protocol was approved by the local IRB, and all participants provided written informed consent. T1-w anatomical images were acquired as described above. T2-weighted anatomical images were acquired using a fluid attenuated inversion recovery (FLAIR) sequence with the following acquisition parameters: TR/TI: 11000/2800 ms; TE: 125 ms; field of view: 240 × 240 mm2; reconstruction matrix: 224 × 224; voxel size: 1.0 × 1.0 × 3.0 mm3; and axial slices: 70. BOLD-fMRI data were acquired with the optimized SAGE acquisition. Specifically, acquisition parameters were as follows: SENSE: 2.5; MB: 2; TE1-5: 8.0/27/59/78/97 ms; TR: 3.0 s; field of view: 240 × 240 mm2; voxel size: 3.0 × 3.0 × 3.0 mm3; axial slices: 40; and volumes: 112 (working memory task) and 72 (visual stimulus task). A reverse phase encoding acquisition was collected for distortion correction.

Two tasks were implemented to validate the use of SAGE-fMRI: working memory and visual stimulus. These tasks were chosen to assess the performance of SAGE-fMRI in a range of brain regions. The N-back task was implemented to evoke activation in regions associated with working memory (Barch et al., 2013; Drobyshevsky et al., 2006; Marshall et al., 2004; Ragland et al., 2002). This task has shown robust activation in the dorsolateral and anterior prefrontal, inferior frontal, anterior cingulate, and dorsal parietal cortices (Barch et al., 2013; Drobyshevsky et al., 2006). These were considered regions of interest (ROI) in the fMRI analysis and results. A block design was used with 2-back versus 0-back blocks, where the 0-back task was the baseline condition. For the 2-back task, subjects were instructed to click the response device when the letter on screen matched the one that appeared two letters previously; there were 2 matches per block. For the 0-back task, subjects were instructed to click the response device when a specified letter appeared on the screen; there were 3 matches per block. Stimuli appeared for 2 s. Blocks were 30 s each, and there were 10 blocks (5 per condition) in the run.

In a single subject, SAGE-fMRI and standard single-echo GRE and SE acquisitions were acquired during the N-back task. All fMRI parameters matched those described above for SAGE-fMRI, except TE and TR were set to 30 ms and 1.5 s, respectively, for GRE, and the SE TE was set to 80 ms (with 3 s TR). Due to the shorter TR, 225 dynamics were acquired for the GRE, yielding the same overall scan time. A reverse phase encoding acquisition was collected for distortion correction. Data were processed as described below, thus enabling direct comparisons between TE2 and GRE and between TE5 and SE. As the GRE acquisition had twice as many dynamic volumes, the data were analyzed using both all volumes and with only the odd volumes, enabling a more direct comparison to SAGE-fMRI.

The visual task entails a flashing (8 Hz) radial checkerboard paradigm, previously shown to robustly activate the visual network (Drobyshevsky et al., 2006; Schneider et al., 1993). Subjects were instructed to keep their eyes open for the duration of the task. The task was organized in a block design, where the baseline condition was a crosshair. Blocks were 21 s each, and there were 10 blocks (5 per condition) in the run.

2.3.1 Data pre-processing: SAGE-fMRI optimization

For the optimization portion of this study, the signals from each SAGE acquisition underwent pre-processing to correct for motion (using TE2 correction for all 5 TEs, 3dvolreg, AFNI (Cox, 1996)) and distortion (topup, applytopup, FSL (Jenkinson et al., 2012)). Root mean square (RMS) was calculated for translational and rotational motion; any acquisitions exceeding RMS = 0.3 were excluded from analysis. FreeSurfer (Fischl, 2012) analysis (recon-all) was performed for all subjects on anatomical T1- and T2-weighted images to extract gray and white matter masks for subsequent analysis.

2.3.2 Data pre-processing: SAGE-fMRI validation

For task-based fMRI, each of the five SAGE echoes underwent standard preprocessing. The first two TRs were removed for the signal to reach steady state (these TRs provided instructions to the subject and were therefore not part of the ON-task or baseline conditions; 3dTcat, AFNI). Data were motion corrected and distortion corrected as described above, followed by de-spiking (3dDespike, AFNI), slice-timing correction (3dTshift, AFNI), and brain-extraction (bet, FSL).

2.4 Echo time combinations

SAGE-fMRI data were processed using both single-echo and multi-echo combinations using an in-house MATLAB script (version R2018b, MathWorks, Natick, MA, United States), with a total of nine combinations (5 single-echo and 4 multi-echo). More specifically, each TE was processed individually (five echoes with signals S1 - S5), where S2 (TE = 27 ms) is the most similar to a standard (single-echo) BOLD-fMRI acquisition and S5 (TE = 97 ms) is similar to single-echo SE-fMRI. SAGE TE2 and TE5, therefore, were included in subsequent statistical analyses for comparison to multi-echo data. For multi-echo analysis, both quantitative fitting of relaxation rates and relaxation-weighted summations were investigated. First, dynamic relaxation rates R2* and R2 were quantified by linear least-squares fitting of the log-transformed signals across all SAGE echoes (Eq. 1) (Sisco et al., 2022). An upper threshold of 500 ms was applied to SAGE T2*- and T2-weighted maps to remove outlier voxels; average T2* and T2 values for the optimized SAGE-fMRI protocol are reported in Supplementary Materials Table S1. Second, weighting strategies (i.e., wT2* and wT2) were implemented only for the validation portion of this study, using Eqs. 2 and 3. Weighting is higher near air-tissue interfaces at shorter TEs, which may improve quantification in these regions (Supplementary Materials Fig. S1B). For T2*-weighting, the non-exponential term (Eq. 2) cancels for S5, yielding zero contribution as expected due to the complete refocusing of T2* effects. For T2-weighting, there is no BOLD contribution prior to application of the refocusing pulse.

2.5.1 Statistical analysis: SAGE-fMRI optimization

To compare the various single- and multi-echo analysis methods, temporal SNR (tSNR) was calculated for each TE across optimization protocols. The tSNR was calculated voxel-wise using the mean signal intensity S¯ over time divided by the standard deviation (SD) of the signal (σS):

[4]

Analysis of variance (ANOVA) was implemented to compare tSNR across MB and SENSE factors. Post-hoc analysis (Tukey’s HSD test) was performed to compare MB and SENSE protocols that exhibited satisfactory acquisition parameters (i.e., whole brain coverage ≥120 mm and TE5 < 100 ms).

To assess loss in SNR due to acceleration, relative g-factor maps were calculated by dividing the product of the SNR of the least accelerated images (in this study, MB1-SENSE2 is the reference protocol) and the square root of its acceleration factor (R) by the product of the SNR of the higher acceleration images and the square root of its R (Robson et al., 2008):

[5]

SNR was calculated using the difference method, as previously described (Dietrich et al., 2007; Reeder et al., 2005), where the mean whole brain signal from the sum of two images (even and odd dynamics) was divided by the standard deviation of the difference of the two images (Krüger et al., 2001).

Therefore, tSNR and g-factor were used, along with qualitative image quality assessment, full brain coverage (≥120 mm), and sufficient TEs (i.e., TE1 < 10 ms and TE5 < 100 ms,), to determine the optimal protocol for whole-brain task-based fMRI.

2.5.2 Statistical analysis and data post-processing: SAGE-fMRI validation

For task-based fMRI, six maps were assessed to directly compare single GRE and single SE fMRI to multi-echo methods: TE2, TE5, T2*, T2, wT2*, and wT2; additionally, task-based activation and tSNR results for SAGE TE1,3-4 are provided in the Supplementary Materials.

tSNR was calculated as the mean signal divided by the standard deviation of the noise time series (i.e., unexplained signal from the general linear model; σnoise):

[6]

A t-test with pairwise comparisons was performed between T2*- and T2-dominant tSNR maps; results were corrected by cluster size (3dFWHM, 3dClustSim, AFNI) and for multiple comparisons within each map (False Discovery Rate (FDR) <0.05) and across tests (Bonferroni correction).

In addition to tSNR, CNR was calculated, defined as the difference of the signal during stimulus ON and baseline OFF divided by σnoise (Geissler et al., 2007):

[7]

Effect size, specifically Hedge’s g, was calculated for working memory and visual stimulus tasks (R version 3.6.3, https://www.R-project.org/, RStudio version 1.4.1717, http://www.rstudio.com/. large effect at g ≥ 0.80). Hedge’s g is defined as a standardized effect size for measuring the difference between two group means and is calculated as the difference of the signal during stimulus ON and baseline OFF, divided by the SD of the signal σS:

[8]

For task-based fMRI, each single- and multi-echo combination generated from the SAGE BOLD-fMRI acquisition was co-registered to the subject’s anatomical T1-w MPRAGE through an applied affine transformation performed on TE2 (degrees of freedom = 12, flirt; FSL). Anatomical images were brain-extracted prior to co-registration (ROBEX (Iglesias et al., 2011)). Subsequently, all SAGE-derived images from each subject were linearly normalized to the MNI space using the co-registration matrix obtained from normalization between MPRAGE and the MNI standard image (flirt; FSL). The SAGE-derived images in MNI space were smoothed with a 5.0 mm full-width half-maximum Gaussian blur (3dmerge; AFNI) and scaled to have an average fMRI intensity of 100 (3dTstat, 3dcalc, AFNI). Statistical parametric maps of response to stimuli from SAGE-derived images were obtained (3dDeconvolve, AFNI), and group analysis was performed (3dttest++, AFNI). ROI-based analyses were conducted by using the Harvard Oxford Cortical and Subcortical Atlases (Desikan et al., 2006) and the MNI Structural Atlas (Collins et al., 1995; Mazziotta et al., 2001), using a 50 percent probability threshold. For the working memory task, the following ROIs were included: frontal pole, inferior frontal gyrus (pars triangularis and opercularis), middle frontal gyrus, parietal lobe, cingulate gyrus (anterior and posterior divisions), and frontal medial cortex (Barch et al., 2013; Drobyshevsky et al., 2006). Task-based activation is expected in the frontal pole, inferior frontal gyrus, middle frontal gyrus, and parietal lobe, while default mode network-associated regions (Raichle, 2015) in the cingulate gyrus and frontal medial cortex are expected to be suppressed, showing deactivation. Additionally, the frontal medial cortex ROI allows for comparison of SAGE methods to single-echo fMRI in a region prone to susceptibility effects. For the visual stimulus task, the following ROIs were included: occipital pole, occipital fusiform gyrus, lateral occipital cortex (superior and inferior divisions), and temporal occipital fusiform cortex. To assess differences in GRE- and SE-based functional activation, task-based activation maps were compared in native fMRI space. The Venous Neuroanatomy probabilistic atlas (VENAT) was co-registered into native space and used as a reference for the location of large draining veins in the brain to observe whether macro- and microvascular-weighted methods show significant activation within these veins (Huck et al., 2019).

3.1 SAGE-fMRI: Optimization

Figure 1 shows the SAGE signals in a single slice across various combinations of MB and SENSE factors. Without MB (with SENSE = 2 and PF = 0.74), only 75 mm brain coverage was achieved. MB factors 2, 3, and 4 were implemented to enable full brain coverage (i.e., ≥120 mm), with SENSE factors used to ensure adequate TEs (i.e., TE1 < 10 ms and TE5 < 100 ms, although TE5 < 120 ms was permitted to be able to test relevant acceleration factors; Table 1). Qualitative analysis revealed loss of signal and image quality in MB4 across SENSE factors; brain coverage achieved by each protocol is shown in the bottom right sub-panel, where the four least accelerated protocols did not yield full coverage.

Fig. 1.

Single-subject acquisition for each TE across acceleration factor combinations. T1-weighted (T1-w) sagittal image is included with each maximum brain coverage achievable per protocol listed in the legend.

Fig. 1.

Single-subject acquisition for each TE across acceleration factor combinations. T1-weighted (T1-w) sagittal image is included with each maximum brain coverage achievable per protocol listed in the legend.

Close modal

Figure 2 shows the loss in image quality from minimum to maximum acceleration protocols in a single subject, where the arrows highlight examples of visible artifacts in the most accelerated acquisition.

Fig. 2.

Representative SAGE images for minimum and maximum acceleration protocols per TE. Arrows highlight visible acceleration-related artifacts, apparent in TE1 as bands and in TE5 as loss of definition and blurring.

Fig. 2.

Representative SAGE images for minimum and maximum acceleration protocols per TE. Arrows highlight visible acceleration-related artifacts, apparent in TE1 as bands and in TE5 as loss of definition and blurring.

Close modal

Figure 3 shows mean tSNR trends in gray matter and white matter across echoes and acceleration factors; specific values are provided in Supplementary Materials Table S2. Gray matter tSNR was lower than that in white matter as expected, with decreasing tSNR observed across echoes and with increasing acceleration. Post-hoc comparisons are reported in Supplementary Materials Table S3 (p < 0.05, corrected for multiple comparisons). While not significant for all protocols, the MB2-SENSE2.5 acquisition showed the highest tSNR values from this group. This was further supported by a quantitative assessment of artifacts due to acceleration via g-factor, calculated for each accelerated protocol on TE2 (Fig. 4). Quantitative assessment of T2* and T2 values in gray and white matter ROIs revealed no significant differences between any of the MB2 and MB3 combinations across SENSE factors (p < 0.05, corrected for multiple comparisons, Supplementary Materials Fig. S2). Based on these analyses, the MB2-SENSE2.5 combination was selected for implementation in task-based fMRI due to its minimal loss in tSNR and sufficient improvement in spatial and temporal resolution.

Fig. 3.

Group-level mean tSNR across acceleration factor combinations for TE1-5 in gray (A) and white matter (B). There were no significant differences in tSNR in gray or white matter across TEs between protocols with feasible acquisition parameters (i.e., full brain coverage ≥120 mm, TE1 < 10 ms, TE5 < 100 ms).

Fig. 3.

Group-level mean tSNR across acceleration factor combinations for TE1-5 in gray (A) and white matter (B). There were no significant differences in tSNR in gray or white matter across TEs between protocols with feasible acquisition parameters (i.e., full brain coverage ≥120 mm, TE1 < 10 ms, TE5 < 100 ms).

Close modal
Fig. 4.

TE2 per acquisition and corresponding g-factor analysis in a representative subject. Increasing g-factor values may correspond to acceleration-associated artifacts seen in TE2. Histograms exhibiting the distribution of whole-brain voxelwise g-factor values are provided.

Fig. 4.

TE2 per acquisition and corresponding g-factor analysis in a representative subject. Increasing g-factor values may correspond to acceleration-associated artifacts seen in TE2. Histograms exhibiting the distribution of whole-brain voxelwise g-factor values are provided.

Close modal

3.2 SAGE-fMRI: Validation

Weighted SAGE maps had the highest tSNR values in raw group average maps within their respective vascular scales (Fig. 5; Supplementary Materials Table S4). For both T2*- (i.e., T2*, wT2*, and TE2) and T2-dominant (i.e., T2, wT2, and TE5) analyses, weighted SAGE maps showed significantly higher tSNR across the brain compared to quantitative SAGE maps (wT2* > T2*: maximum t-score (tmax) = 43.1 significant for 99.5% total brain volume; wT2 > T2: tmax = 24.8, significant for 99.3% total brain volume) and single-echo fMRI (wT2* > TE2: tmax = 17.1, significant for 43.5% total brain volume; wT2 > TE5: tmax = 28.9, significant for 70.0% total brain volume). Notably, in the temporal and inferior frontal lobes, there was no significant difference between T2* and TE2, and in portions of these regions, T2* tSNR was significantly higher than TE2 (tmax = 4.7, significant for 1.1% of temporal and inferior frontal lobe volumes). This was not seen between T2 and TE5 except in some inferior frontal lobe white matter; brain regions in which T2 were significantly higher than TE5 were in the thalamus and basal ganglia (tmax = 3.6, significant for 1.49% of thalamus and basal ganglia volumes).

Fig. 5.

Group-level tSNR maps for the working memory task across macro- (TE2, SAGE T2*, and SAGE wT2*) and microvascular (TE5, SAGE T2, and SAGE wT2) acquisitions. Anatomical T1-w image is included for reference.

Fig. 5.

Group-level tSNR maps for the working memory task across macro- (TE2, SAGE T2*, and SAGE wT2*) and microvascular (TE5, SAGE T2, and SAGE wT2) acquisitions. Anatomical T1-w image is included for reference.

Close modal

CNR values were slightly higher for weighted SAGE than quantitative SAGE and single-echo methods, respectively (Fig. 6; Supplementary Materials Table S4). One of the more notable gains occurred in the inferior frontal gyrus, where higher CNR averages and maximum values are found in the weighted SAGE analyses. This is further supported by the t-scores.

Fig. 6.

Group-level CNR maps for the working memory task across macro- (TE2, SAGE T2*, and SAGE wT2*) and microvascular (TE5, SAGE T2, and SAGE wT2) acquisitions. Anatomical T1-w image with task-related regions of interest (ROIs) is included for reference.

Fig. 6.

Group-level CNR maps for the working memory task across macro- (TE2, SAGE T2*, and SAGE wT2*) and microvascular (TE5, SAGE T2, and SAGE wT2) acquisitions. Anatomical T1-w image with task-related regions of interest (ROIs) is included for reference.

Close modal

For the working memory task, there was significant activation (p < 0.001, cluster size corrected) in ROIs across T2*- and T2-dominant maps for single-echo and SAGE methods (Fig. 7; Supplementary Materials Table S4); activation and tSNR for the remaining SAGE echoes are provided in Supplementary Materials Figure S3. T2*-dominant maps showed similar activation and deactivation; SAGE wT2* showed more robust deactivation (-tmax) and a larger voxel count for significant activation and deactivation than TE2 and T2* methods. T2-dominant maps showed similar deactivation, except for wT2 maps revealing significant deactivation in the anterior cingulate gyrus and frontal pole where TE5 and T2 did not (Fig. 8; Supplementary Materials Table S4). Both SAGE T2 and wT2 map had higher tmax than TE5; the wT2 map also had a higher voxel count for significant activation than other SE-based methods. As noted with CNR, there were noticeable gains in significant voxels from TE2 to SAGE wT2*, where the former showed unilateral activation in the right inferior frontal gyrus and the latter showed bilateral activation; bilateral activation was also seen for wT2, where neither TE5 nor T2 showed significant activation in this region. Notably, there is an increase in deactivated voxels in the frontal medial cortex for weighted SAGE analyses compared to single-echo counterparts, likely reflecting an improvement in sensitivity and (for GRE) susceptibility-induced dropout in this region.

Fig. 7.

Group-level t-maps for the working memory task across macro- (TE2, SAGE T2*, and SAGE wT2*) and microvascular (TE5, SAGE T2, and SAGE wT2) acquisitions. # vox = voxel count for significant functional activation.

Fig. 7.

Group-level t-maps for the working memory task across macro- (TE2, SAGE T2*, and SAGE wT2*) and microvascular (TE5, SAGE T2, and SAGE wT2) acquisitions. # vox = voxel count for significant functional activation.

Close modal
Fig. 8.

N-back task-based activation histograms for ROIs with negative (A-C; G) and positive (D-G) activation. Each histogram shows the distribution of statistically significant voxels (p < 0.001, cluster size corrected) for each single-echo or SAGE-based method, with the significant voxel count in the legend. FP had both positive (dorsolateral prefrontal cortex) and negative (dorsomedial prefrontal cortex) activation. Voxel clusters of less than 100 were removed for visualization. FMC = Frontal Medial Cortex; CGa = Cingulate Gyrus anterior division; CGp = Cingulate Gyrus posterior division; IFG = Inferior Frontal Gyrus; MFG = Middle Frontal Gyrus; PL = Parietal Lobe; FP = Frontal Pole.

Fig. 8.

N-back task-based activation histograms for ROIs with negative (A-C; G) and positive (D-G) activation. Each histogram shows the distribution of statistically significant voxels (p < 0.001, cluster size corrected) for each single-echo or SAGE-based method, with the significant voxel count in the legend. FP had both positive (dorsolateral prefrontal cortex) and negative (dorsomedial prefrontal cortex) activation. Voxel clusters of less than 100 were removed for visualization. FMC = Frontal Medial Cortex; CGa = Cingulate Gyrus anterior division; CGp = Cingulate Gyrus posterior division; IFG = Inferior Frontal Gyrus; MFG = Middle Frontal Gyrus; PL = Parietal Lobe; FP = Frontal Pole.

Close modal

A preliminary comparison between SAGE-fMRI (TE2 and TE5) and separate single-echo BOLD acquisitions optimized for a single contrast (both GRE and SE) is shown in a representative subject in Supplementary Materials Figure S4. In the single-echo GRE acquisition, the impact of both the slight difference in TE (27 ms for SAGE-TE2 vs. 30 ms for GRE) and the much shorter TR for GRE is assessed. Minimal differences are observed between the SAGE-TE2 and GRE with an equal number of dynamics (i.e., GREalt_dyn), suggesting that the impact of TE is minimal. However, comparisons between SAGE-TE2 and the full (i.e., 225 dynamics) GRE analysis demonstrate the benefits of more volumes enabled by the shorter TR. The optimized SE shows more significant voxels compared to SAGE-TE5, which may be the result of the shorter TE (80 ms vs. 97 ms, respectively). It should be noted that the SAGE-TE4 has a similar TE (78 ms) to the optimized SE protocol, and the subsequent results for SAGE-TE4 more closely matched that of the single-echo SE (data not shown).

Effect size analysis showed similar trends as the t-test results (Supplementary Materials Fig. S5). SAGE wT2* showed a larger maximum effect size for deactivation (-gmax) and a larger count for voxels with at least a large effect size compared to TE2 and T2* methods. To a greater extent, SAGE wT2 had higher counts for voxels with at least a large effect size than TE5 or T2, but the maximum effect sizes were more similar across T2-dominant maps. There was a large effect cluster in SAGE wT2 posterior cingulate cortex, as expected and seen in T2*-dominant maps, that was not found in TE5 or T2 methods.

For the visual task, there was significant activation (p < 0.001, cluster size corrected) in ROIs across single-echo and SAGE methods (Supplementary Materials Fig. S6). SAGE wT2 maps showed significant activation in the temporal occipital fusiform and superior lateral occipital cortices, whereas TE5 and T2 did not. Both SAGE wT2* and wT2 maps had higher tmax and a higher voxel count for significant activation than other T2*- and T2-dominant methods, respectively.

Effect size analysis for the visual task showed that SAGE wT2* and wT2 had higher maximum effect sizes and counts for voxels with at least a large effect size than their single-echo (TE2 and TE5) and quantitative (T2* and T2) counterparts (Supplementary Materials Fig. S7).

Activation for the visual task was observed in the occipital lobe for both GRE (i.e., TE2) and SE (i.e., TE5) fMRI, as well as multi-echo SAGE-fMRI (FDR correction for multiple comparisons, q < 0.05; Fig. 9). When referencing the VENAT atlas, there is significant task-based functional activation that appears to overlap with the draining straight and transverse sinuses in the GRE-based analyses. Conversely, SE-based functional activation appears to be largely localized within the gray matter of the visual cortex.

Fig. 9.

Task-based activation (corrected for multiple comparisons, FDR q < 0.05) in a representative subject for all SAGE macro- (TE2, SAGE T2*, and SAGE wT2*) and microvascular (TE5, SAGE T2, and SAGE wT2) analyses. The VENAT probabilistic atlas, co-registered into native space, is provided for reference corresponding to large draining veins in the brain. VENAT and activation maps are overlaid on T1-w anatomical image co-registered into native space. Macrovascular-weighted activations overlap with the straight, transverse, and sagittal sinuses, whereas microvascular-weighted activations appear to be largely constrained to gray matter of the visual cortex. P(VENAT) = VENAT atlas probability estimate.

Fig. 9.

Task-based activation (corrected for multiple comparisons, FDR q < 0.05) in a representative subject for all SAGE macro- (TE2, SAGE T2*, and SAGE wT2*) and microvascular (TE5, SAGE T2, and SAGE wT2) analyses. The VENAT probabilistic atlas, co-registered into native space, is provided for reference corresponding to large draining veins in the brain. VENAT and activation maps are overlaid on T1-w anatomical image co-registered into native space. Macrovascular-weighted activations overlap with the straight, transverse, and sagittal sinuses, whereas microvascular-weighted activations appear to be largely constrained to gray matter of the visual cortex. P(VENAT) = VENAT atlas probability estimate.

Close modal

In this study, we implemented and optimized a multi-echo and multi-contrast fMRI method based on the SAGE sequence, with proposed fMRI weighting strategies to combine the advantages of MGE- and SE-fMRI. MGE-fMRI methods have several benefits over single GRE EPI, most notably in reduced image artifacts and improved BOLD sensitivity (Cohen et al., 2018; Evans et al., 2015; Kundu et al., 2013; Poser et al., 2006; Posse et al., 1999; Speck & Hennig, 1998). Adding an additional echo with a shorter TE has virtually no effect on any other pulse sequence parameter: the typical BOLD signal (TE2) is preserved, the spatial resolution and slice coverage are maintained, and the temporal resolution and acquisition time are unchanged. With the inclusion of a shorter first TE, the BOLD sensitivity was previously shown to improve by a minimum of 11%, with some sensitivity gains reported up to 67% in regions with high susceptibility-induced signal dropouts (Poser et al., 2006). Other studies have demonstrated that MGE-BOLD improves denoising, yielding higher SNR for functional connectivity (Cohen, Yang, et al., 2021; Kundu et al., 2013) and task-based activation analysis (Cohen, Jagra, et al., 2021; Cohen et al., 2018). However, multi-echo methods suffer many of the same limitations as single GRE methods: sensitivity to large vessels and some (albeit improved) regional signal dropout due to susceptibility. Presently, we observed that GRE-fMRI functional activation analyses may be influenced by sensitivity to large draining veins within the context of a visual task (Fig. 9), in agreement with what we expect based on the literature concerning vessel size sensitivities (Boxerman et al., 1995; Weisskoff et al., 1994).

SE methods have much lower BOLD sensitivity but are not sensitive to susceptibility-induced signal drop-out, and thus these methods have advantages in artifact-prone regions such as the inferior frontal lobe and temporal lobes. An additional benefit of SE-based contrast is improved specificity to the microvasculature (Boxerman et al., 1995; Stables et al., 1998; Weisskoff et al., 1994), due to reduced sensitivity to large vessels. At clinical field strengths (such as 3T), SE-fMRI comprises both intravascular (blood) T2 effects and extravascular (microvessel) effects; this typically necessitates either the use of longer TEs (Ragot & Chen, 2019) or higher field strengths (Duong et al., 2003; Lee et al., 1999) to minimize intravascular effects. Although SE contrast provides higher microvascular specificity (Parkes et al., 2005), this comes at a cost of BOLD sensitivity (Duong et al., 2003; Lowe et al., 2000; Parkes et al., 2005). Other approaches to overcome this limitation could include the use of ultra-high-resolution GRE-fMRI; notably, laminar fMRI has previously shown activation localized within gray matter with sub-millimeter resolution at 3T (Knudsen et al., 2023). In the present study, SE-fMRI showed lower BOLD sensitivity (i.e., smaller volume of clusters with task-based activation), but perhaps higher spatial specificity (i.e., less sensitivity to large draining veins), as activation is mostly within the visual cortex (Fig. 9). Finally, SE-BOLD is less sensitive to physiological noise (Ragot & Chen, 2019) which is a known confounder for fMRI analysis (Birn et al., 2006; Chang et al., 2009). Multi-contrast SAGE-fMRI combines the strengths of MGE- and SE-fMRI.

Another potential advantage for SAGE-fMRI is the direct quantification of dynamic R2* and R2 time courses. Although quantitative T2* and T2 are more directly related to changes in BOLD (Buxton, 2013; Poser et al., 2006; Speck & Hennig, 1998), previous research has shown that T2* is less robust compared to weighted signal combinations (Posse et al., 1999) that aim to maximize BOLD contrast sensitivity by considering local T2* values (Poser et al., 2006). In this study, we observed smaller clusters of significant task-based activation and effect size for both T2* and T2 relative to wT2* and wT2, respectively; by weighting for T2* and T2, higher BOLD contrast sensitivity is expected (i.e., CNR), yielding larger clusters. The T2* and T2 clusters were also smaller than any of the single-echo fMRI clusters. Smaller T2* clusters could be due to separation of BOLD and non-BOLD signal that comes with multi-echo fMRI, possibly improving specificity to the BOLD effect compared to single GRE (Kundu et al., 2012). Smaller SE clusters likely result from both reduced BOLD contrast and from higher spatial specificity (from reduced sensitivity to draining veins) (Duong et al., 2003; Lowe et al., 2000; Parkes et al., 2005). The former is a disadvantage for SE-fMRI, while the latter may be advantageous. In this study, we consistently observed the largest gains across metrics from TE5 and T2 to wT2; by weighting for T2, there is a tradeoff between the loss in microvascular specificity and the gain in BOLD sensitivity. However, as all contrasts can empirically be compared using SAGE-fMRI, we anticipate that the higher spatial specificity associated with TE5 and T2 contrast can provide insight into the underlying sources of activation clusters observed with wT2 contrast.

The SAGE sequence includes multiple echoes and multiple types of contrast, but a drawback of this technique is the necessary trade-offs between sequence parameters. In various iterations of this sequence, GRAPPA, SENSE, PF, and MB methods have been added to accelerate image acquisition (Han et al., 2021; Manhard et al., 2019, 2021; Skinner et al., 2014), yielding better spatial and temporal resolution. Studies have also shown that removing the ASE readouts may enable shorter SE TEs, an approach entitled simplified SAGE (Stokes & Quarles, 2016; Stokes, Skinner, et al., 2016). In this study, various SENSE and MB factors were tested with standard five-echo SAGE, and the requisite parameters included 3 mm isotropic resolution, whole-brain coverage (≥120 mm), TE1 less than 10 ms, TE5 less than 100 ms, and TR equal to 3 s. The TE requirements were in place to ensure adequate signal in each echo. Studies requiring higher spatial resolution would have tradeoffs with TEs, and thus slice coverage may diminish. Conventional GRE-based EPI can typically achieve shorter TRs (on the order of 2 s or less) with whole-brain coverage, but the inclusion of SE contrast necessitates the longer TR for SAGE. In theory, steady-state effects (T1-recovery) decrease the signal intensity at shorter TRs; on the other hand, a longer TR decreases the number of image volumes that can be acquired in a fixed scan length. In general, shorter TRs can give higher detection sensitivity (for a fixed scan time), ignoring serial correlation effects (Sahib et al., 2016). This can be readily seen in the direct comparison between SAGE-fMRI and single-echo GRE, where the shorter TR for the latter enabled improved sensitivity due to the increased number of dynamics. In this study, we held the SAGE-fMRI TR constant at 3 s, although it could have been shortened with increasing acceleration. While the SAGE-fMRI acquisition was optimized here for our specific spatiotemporal requirements, it is important to note that future implementations could further optimize SAGE-fMRI to best address different research questions. Furthermore, other approaches such as multi-shot EPI or EPI with keyhole could be explored to modulate spatial or temporal resolution as needed (Küppers et al., 2022; Wang et al., 2019).

Presently, we have proposed optimized fMRI weighting strategies for SAGE-fMRI that leverage the multiple echoes and multiple types of contrast. Few studies have considered SAGE for fMRI. For example, SAGE-fMRI was previously performed in a single subject using a breath-hold stimulus, demonstrating the feasibility of this method (Schmiedeskamp et al., 2014). Additionally, SAGE-fMRI has been proposed for assessing relative vessel sizes during functional activation (Newton et al., 2014), similar to vessel size imaging methods for perfusion MRI (Kiselev et al., 2005). Glielmi et al. (2010) previously compared simultaneous arterial spin labeling (ASL)-fMRI, GRE-BOLD, and SE-BOLD using a dual-GRE and SE method, similar to simplified SAGE implementations that exclude ASE acquisitions (Stokes & Quarles, 2016; Stokes, Skinner, et al., 2016); more specifically, ASL-based measures of cerebral blood flow (CBF) were obtained from tag-control difference images of the first echo, while GRE-BOLD and SE-BOLD were obtained from averaged images (tag-control) of the second GRE and the SE, respectively. Using a visual stimulus at 3T, they showed that ASL-CBF was the most localized and had the highest specificity, but the lowest sensitivity. GRE-BOLD had the largest activation and the highest sensitivity, while SE-BOLD was a compromise between ASL-CBF and GRE-BOLD (that is, more localized than GRE-BOLD and more sensitive than ASL-CBF) (Glielmi et al., 2010). However, the present study is the first to propose an optimized analysis for SAGE-fMRI based on echo-weighting strategies that leverage the multiple echoes and types of contrast simultaneously.

Comparing SAGE-based T2*-weighted combined images to MGE-fMRI was beyond the scope of the present work but could be an important future direction for SAGE-fMRI. Additionally, while two GREs are implemented in SAGE-fMRI, other studies have utilized three to five to fit for T2* (Cohen et al., 2018; Evans et al., 2015; Kundu et al., 2013; Poser et al., 2006; Posse et al., 1999; Speck & Hennig, 1998). Because T2* values vary across the brain (Hagberg et al., 2002), and gray matter signal decay behaves monoexponentially over a range of TEs (Hagberg et al., 2002; Speck & Hennig, 1998), increasing the number of GREs can enable more accurate estimations of T2*. Moreover, increasing acceleration can reduce EPI readouts, which may enable more optimal TE choices to maximize T2* contrast in various brain regions (Newbould et al., 2007).The range of TEs (∆TE) may also be important to the accurate estimation of T2*; ∆TE should be approximately equal to or greater than the true T2* (Shin et al., 2023). Presently, the SAGE-fMRI acquisition had ∆TE of 19 ms; both the number of echoes and ∆TE are somewhat mitigated by the inclusion of two mixed-contrast ASEs (four echoes, ∆TE = 70 ms). Empirical examination of the difference in T2* estimation between SAGE-fMRI and the more typical multi-GRE approach may be important for understanding the effect of number and contrast of echoes on T2* estimation. Despite this, quantitative T2* values from the optimized SAGE protocol were comparable to those reported in literature for other multi-echo, multi-contrast acquisitions (Küppers et al., 2022), while SAGE-T2 was slightly overestimated in gray matter (Supplementary Materials Table S1). This could be due to differences in acquisitions related to minimum TE achievable or only acquiring one true SE signal with SAGE.

Herein, we have demonstrated that SAGE can be used in fMRI for whole-brain, task-based multi-echo analyses. SAGE-fMRI offers many potential applications, such as resting-state functional connectivity. Because SAGE-fMRI offers improvements and alternative contrasts to those susceptible to signal dropout, it may be applied in pathologies localized within the medial temporal and inferior frontal lobes, such as Alzheimer’s disease and frontotemporal dementia. Moreover, SAGE-fMRI may be of particular interest when pathology is localized in the microvasculature. Additionally, SAGE-fMRI can be used to quantify multiple metrics such as oxygen extraction fraction and cerebrovascular reactivity, thus providing a more comprehensive picture of vascular and metabolic function in the brain on complementary vascular scales.

There are several important limitations to this study and more generally to SAGE-based fMRI. Limitations regarding the optimization study include that only 40 dynamics were acquired and TEs varied across protocols, which could affect tSNR estimation; additionally, g-factor calculations are reported relative to the least accelerated protocol rather than to a non-accelerated protocol. Another limitation is related to the relaxation-weighting factors implemented herein. By including ASE signals with mixed sensitivities and a SE at 3T (where intravascular contributions remain), it is important to acknowledge that the T2-weighted analyses likely include T2* contributions. Additionally, the proposed weighting strategy inherently necessitates analysis with two signals (wT2* and wT2), whereas future studies may prefer a single optimized signal for analysis. Other options to reduce data dimensionality include a simple summation (Posse et al., 1999), CNR-weighting (Poser et al., 2006), vessel-size weighting (Kiselev et al., 2005; Newton et al., 2014; Troprès et al., 2001), or even other relaxation-weightings, such as T2’. Further research should investigate the advantages of these methods in the context of combined contrast across GRE, ASE, and SE readouts. An additional challenge to SAGE-fMRI, specifically fitting for dynamic T2* and T2 values, is the relatively long time required for fitting. This was presently addressed by implementation of a linear least-squares fitting solution (Sisco et al., 2022). Moreover, we previously showed that a simplified SAGE approach may enable real-time quantification of relative T2* and T2 changes (Stokes & Quarles, 2016; Stokes, Skinner, et al., 2016), which may be similarly amenable for SAGE-fMRI.

Finally, one of the challenges for fMRI is the complex origin of the BOLD signal, which depends on physiological and biophysical effects, as well as non-BOLD factors such as pulse sequence parameters (Buxton, 2013). We hypothesize that the inclusion of multiple types of contrast could enable separation of intravascular and extravascular effects at clinical field strengths, though such an investigation is outside the scope of this study. Vessel size specificity may also be attained by tuning the combination of GRE and SE signals (Newton et al., 2014). Moreover, SAGE-fMRI may permit more advanced analysis into spatiotemporal differences between GRE and SE activation, which are critical for understanding neurovascular dynamics in a range of pathologies. A more detailed characterization and quantification of BOLD signal contributions within a single experiment, using the SAGE-fMRI framework developed in this study, may improve our understanding of fMRI mechanisms and interpretation of the origins of functional activation.

SAGE-fMRI enables the simultaneous acquisition of multiple echoes and contrasts, with a wide range of possible fMRI applications. Using SAGE-fMRI, two contrast mechanisms can be feasibly acquired, separated, and enhanced with the relaxation-weighted summation analysis methods. SAGE-fMRI couples the high BOLD sensitivity from MGE acquisitions with the potential microvascular specificity from SE acquisitions. SAGE-fMRI provides several advantages, including improving tSNR and BOLD CNR for more accurate analysis. This combination of multiple echoes and contrast mechanisms bolsters the confidence in the resulting functional maps. SAGE-fMRI may ultimately provide new insight into the complex contributions to BOLD signal and improve spatial localization of brain activation. Overall, SAGE-fMRI provides flexibility for analysis and can be optimized for different contrast mechanisms.

Data and code may be made available upon the construction of a formal data sharing agreement.

Elizabeth G. Keeling: Conceptualization, Data curation, Formal analysis, Methodology, Visualization, and Writing—original draft. Maurizio Bergamino: Formal analysis, Visualization, and Writing—review and editing. Sudarshan Ragunathan: Methodology, Software. C. Chad Quarles: Funding acquisition, Methodology, Software, and Writing—review and editing. Allen T. Newton: Conceptualization, Visualization, and Writing—review and editing. Ashley M. Stokes: Conceptualization, Funding acquisition, Methodology, Project Administration, Supervision, Visualization, and Writing—review and editing.

This work was supported by the Barrow Neurological Foundation, Philips Healthcare, and the NIH (NINDS 1R01NS12457, NCI 2R01CA158079).

Dr. Stokes has received speaking fees from Eli Lilly.

Supplementary material for this article is available with the online version here: https://doi.org/10.1162/imag_a_00217.

Barch
,
D. M.
,
Burgess
,
G. C.
,
Harms
,
M. P.
,
Petersen
,
S. E.
,
Schlaggar
,
B. L.
,
Corbetta
,
M.
,
Glasser
,
M. F.
,
Curtiss
,
S.
,
Dixit
,
S.
,
Feldt
,
C.
,
Nolan
,
D.
,
Bryant
,
E.
,
Hartley
,
T.
,
Footer
,
O.
,
Bjork
,
J. M.
,
Poldrack
,
R.
,
Smith
,
S.
,
Johansen-Berg
,
H.
,
Snyder
,
A. Z.
, &
Van Essen
,
D. C.
(
2013
).
Function in the human connectome: Task-fMRI and individual differences in behavior
.
NeuroImage
,
80
,
169
189
. https://doi.org/10.1016/J.NEUROIMAGE.2013.05.033
Bell
,
L. C.
,
Does
,
M. D.
,
Stokes
,
A. M.
,
Baxter
,
L. C.
,
Schmainda
,
K. M.
,
Dueck
,
A. C.
, &
Quarles
,
C. C
. (
2017
).
Optimization of DSC MRI echo times for CBV measurements using error analysis in a pilot study of high-grade gliomas
.
AJNR American Journal of Neuroradiology
,
38
(
9
),
1710
1715
. https://doi.org/10.3174/ajnr.A5295
Binney
,
R. J.
,
Embleton
,
K. V.
,
Jefferies
,
E.
,
Parker
,
G. J. M.
, &
Ralph
Lambon
,
A.
M.
(
2010
).
The ventral and inferolateral aspects of the anterior temporal lobe are crucial in semantic memory: Evidence from a novel direct comparison of distortion-corrected fMRI, rTMS, and semantic dementia
.
Cerebral Cortex
,
20
(
11
),
2728
2738
. https://doi.org/10.1093/cercor/bhq019
Birn
,
R. M.
,
Diamond
,
J. B.
,
Smith
,
M. A.
, &
Bandettini
,
P. A
. (
2006
).
Separating respiratory-variation-related fluctuations from neuronal-activity-related fluctuations in fMRI
.
NeuroImage
,
31
(
4
),
1536
1548
. https://doi.org/10.1016/j.neuroimage.2006.02.048
Boxerman
,
J. L.
,
Hamberg
,
L. M.
,
Rosen
,
B. R.
, &
Weisskoff
,
R. M.
(
1995
).
MR contrast due to intravascular magnetic susceptibility perturbations
.
Magnetic Resonance in Medicine
,
34
(
4
),
555
566
. https://doi.org/10.1002/MRM.1910340412
Buxton
,
R. B
. (
2013
).
The physics of functional magnetic resonance imaging (fMRI)
.
Reports on Progress in Physics
,
76
(
9
),
96601
. https://doi.org/10.1088/0034-4885/76/9/096601
Chang
,
C.
,
Cunningham
,
J. P.
, &
Glover
,
G. H.
(
2009
).
Influence of heart rate on the BOLD signal: The cardiac response function
.
NeuroImage
,
44
(
3
),
857
869
. https:/doi.org/10.1016/j.neuroimage.2008.09.029
Cohen
,
A. D.
,
Jagra
,
A. S.
,
Yang
,
B.
,
Fernandez
,
B.
,
Banerjee
,
S.
, &
Wang
,
Y
. (
2021
).
Detecting task functional MRI activation using the multiband multiecho (MBME) echo-planar imaging (EPI) sequence
.
Journal of Magnetic Resonance Imaging
,
53
(
5
),
1366
1374
. https://doi.org/10.1002/jmri.27448
Cohen
,
A. D.
,
Nencka
,
A. S.
, &
Wang
,
Y
. (
2018
).
Multiband multi-echo simultaneous ASL/BOLD for task-induced functional MRI
.
PLoS One
,
13
(
2
),
e0190427
. https://doi.org/10.1371/JOURNAL.PONE.0190427
Cohen
,
A. D.
,
Yang
,
B.
,
Fernandez
,
B.
,
Banerjee
,
S.
, &
Wang
,
Y
. (
2021
).
Improved resting state functional connectivity sensitivity and reproducibility using a multiband multi-echo acquisition
.
NeuroImage
,
225
,
117461
. https://doi.org/10.1016/J.NEUROIMAGE.2020.117461
Collins
,
D.
,
Holmes
,
C.
,
Peters
,
T.
, &
Evans
,
A
. (
1995
).
Automatic 3-D model-based neuroanatomical segmentation
.
Human Brain Mapping
,
3
(
3
),
190
208
. https://doi.org/10.1002/hbm.460030304
Cox
,
R. W
. (
1996
).
AFNI: Software for analysis and visualization of functional magnetic resonance neuroimages
.
Computers and Biomedical Research
,
29
(
3
),
162
173
. https://doi.org/10.1006/cbmr.1996.0014
De Martino
,
F.
,
Zimmermann
,
J.
,
Muckli
,
L.
,
Ugurbil
,
K.
,
Yacoub
,
E.
, &
Goebel
,
R
. (
2013
).
Cortical depth dependent functional responses in humans at 7T: Improved specificity with 3D GRASE
.
PLoS One
,
8
(
3
),
e60514
. https://doi.org/10.1371/JOURNAL.PONE.0060514
Desikan
,
R. S.
,
Se
,
F.
,
Fischl
,
B.
,
Quinn
,
B. T.
,
Dickerson
,
B. C.
,
Blacker
,
D.
,
Buckner
,
R. L.
,
Dale
,
A. M.
,
Maguire
,
R. P.
,
Hyman
,
B. T.
,
Albert
,
M. S.
, &
Killiany
,
R. J
. (
2006
).
An automated labeling system for subdividing the human cerebral cortex on MRI scans into gyral based regions of interest
.
NeuroImage
,
31
(
3
),
968
980
. https://doi.org/10.1016/j.neuroimage.2006.01.021
Dietrich
,
O.
,
Raya
,
J. G.
,
Reeder
,
S. B.
,
Reiser
,
M. F.
, &
Schoenberg
,
S. O
. (
2007
).
Measurement of signal-to-noise ratios in MR images: Influence of multichannel coils, parallel imaging, and reconstruction filters
.
Journal of Magnetic Resonance Imaging
,
26
(
2
),
375
385
. https://doi.org/10.1002/JMRI.20969
Drobyshevsky
,
A.
,
Baumann
,
S. B.
, &
Schneider
,
W
. (
2006
).
A rapid fMRI task battery for mapping of visual, motor, cognitive and emotional function
.
NeuroImage
,
31
(
2
),
732
744
. https://doi.org/10.1016/J.NEUROIMAGE.2005.12.016
Duong
,
T. Q.
,
Yacoub
,
E.
,
Adriany
,
G.
,
Hu
,
X.
,
Uǧurbil
,
K.
, &
Kim
,
S. G
. (
2003
).
Microvascular BOLD contribution at 4 and 7 T in the human brain: Gradient-echo and spin-echo fMRI with suppression of blood effects
.
Magnetic Resonance in Medicine
,
49
(
6
),
1019
1027
. https://doi.org/10.1002/MRM.10472
Embleton
,
K. V
,
Haroon
,
H. A.
,
Morris
,
D. M.
,
Ralph
,
M. A. L.
, &
Parker
,
G. J. M
. (
2010
).
Distortion correction for diffusion-weighted MRI tractography and fMRI in the temporal lobes
.
Human Brain Mapping
,
31
(
10
),
1570
1587
. https://doi.org/10.1002/hbm.20959
Evans
,
J. W.
,
Kundu
,
P.
,
Horovitz
,
S. G.
, &
Bandettini
,
P. A
. (
2015
).
Separating slow BOLD from non-BOLD baseline drifts using multi-echo fMRI
.
NeuroImage
,
105
,
189
197
. https://doi.org/10.1016/j.neuroimage.2014.10.051
Faul
,
F.
,
Erdfelder
,
E.
,
Lang
,
A.
, &
Buchner
,
A
. (
2007
).
G * Power 3: A flexible statistical power analysis program for the social, behavioral, and biomedical sciences
.
Behavior Research Methods
,
39
(
2
),
175
191
. https://doi.org/10.3758/bf03193146
Fischl
,
B
. (
2012
).
FreeSurfer
.
NeuroImage
,
62
(
2
),
774
781
. https://doi.org/10.1016/j.neuroimage.2012.01.021
Geissler
,
A.
,
Gartus
,
A.
,
Foki
,
T.
,
Tahamtan
,
A. R.
,
Beisteiner
,
R.
, &
Barth
,
M
. (
2007
).
Contrast-to-noise ratio (CNR) as a quality parameter in fMRI
.
Journal of Magnetic Resonance Imaging
,
25
(
6
),
1263
1270
. https://doi.org/10.1002/JMRI.20935
Glielmi
,
C. B.
,
Xu
,
Q.
,
Craddock
,
R. C.
, &
Hu
,
X
. (
2010
).
Simultaneous acquisition of gradient echo/spin echo BOLD and perfusion with a separate labeling coil
.
Magnetic Resonance in Medicine
,
64
(
6
),
1827
1831
. https://doi.org/10.1002/MRM.22554
Hagberg
,
G. E.
,
Indovina
,
I.
,
Sanes
,
J. N.
, &
Posse
,
S
. (
2002
).
Real-time quantification of T(2)(*) changes using multiecho planar imaging and numerical methods
.
Magnetic Resonance in Medicine
,
48
(
5
),
877
882
. https://doi.org/10.1002/MRM.10283
Han
,
M.
,
Yang
,
B.
,
Fernandez
,
B.
,
Lafontaine
,
M.
,
Angela
,
P. A.
,
Brian
,
J.
,
Melanie
,
L. B.
,
Villanueva-meyer
,
J. E.
,
Lupo
,
J. M.
, &
Chang
,
S. M
. (
2021
).
Simultaneous multi-slice spin- and gradient-echo dynamic susceptibility-contrast perfusion-weighted MRI of gliomas
.
NMR in Biomedicine
,
34
(
1
),
e4399
. https://doi.org/10.1002/nbm.4399
Huck
,
J.
,
Wanner
,
Y.
,
Fan
,
A. P.
,
Jäger
,
A. T.
,
Grahl
,
S.
,
Schneider
,
U.
,
Villringer
,
A.
,
Steele
,
C. J.
,
Tardif
,
C. L.
,
Bazin
,
P. L.
, &
Gauthier
,
C. J
. (
2019
).
High resolution atlas of the venous brain vasculature from 7 T quantitative susceptibility maps
.
Brain Structure and Function
,
224
(
7
),
2467
2485
. https://doi.org/10.1007/S00429-019-01919-4
Iglesias
,
J. E.
,
Liu
,
C. Y.
,
Thompson
,
P. M.
, &
Tu
,
Z
. (
2011
).
Robust brain extraction across datasets and comparison with publicly available methods
.
IEEE Transactions on Medical Imaging
,
30
(
9
),
1617
1634
. https://doi.org/10.1109/TMI.2011.2138152
Jenkinson
,
M.
,
Beckmann
,
C.
,
Behrens
,
T.
,
Woolrich
,
M.
, &
Smith
,
S
. (
2012
).
FSL
.
NeuroImage
,
62
(
2
),
782
790
. https://doi.org/10.1016/j.neuroimage.2011.09.015
Jin
,
N.
,
Guo
,
Y.
,
Zhang
,
Z.
,
Zhang
,
L.
,
Lu
,
G.
, &
Larson
,
A. C
. (
2013
).
GESFIDE-PROPELLER approach for simultaneous R2 and R2* measurements in the abdomen
.
Magnetic Resonance Imaging
,
31
(
10
),
1760
1765
. https://doi.org/10.1016/J.MRI.2013.08.003
Kiselev
,
V. G.
,
Strecker
,
R.
,
Ziyeh
,
S.
,
Speck
,
O.
, &
Hennig
,
J
. (
2005
).
Vessel size imaging in humans
.
Magnetic Resonance in Medicine
,
53
(
3
),
553
563
. https://doi.org/10.1002/Mrm.20383
Knudsen
,
L.
,
Bailey
,
C. J.
,
Blicher
,
J. U.
,
Yang
,
Y.
,
Zhang
,
P.
, &
Lund
,
T. E.
(
2023
).
Improved sensitivity and microvascular weighting of 3T laminar fMRI with GE-BOLD using NORDIC and phase regression
.
NeuroImage
,
271
. https://doi.org/10.1016/J.NEUROIMAGE.2023.120011
Krüger
,
G.
,
Kastrup
,
A.
, &
Glover
,
G. H
. (
2001
).
Neuroimaging at 1.5 T and 3.0 T: Comparison of oxygenation-sensitive magnetic resonance imaging
.
Magnetic Resonance in Medicine
,
45
(
4
),
595
604
. https://doi.org/10.1002/MRM.1081
Kundu
,
P.
,
Brenowitz
,
N. D.
,
Voon
,
V.
,
Worbe
,
Y.
,
Vértes
,
P. E.
,
Inati
,
S. J.
,
Saad
,
Z. S.
,
Bandettini
,
P. A.
, &
Bullmore
,
E. T
. (
2013
).
Integrated strategy for improving functional connectivity mapping using multiecho fMRI
.
Proceedings of the National Academy of Sciences of the United States of America
,
110
(
40
),
16187
16192
. https://doi.org/10.1073/PNAS.1301725110
Kundu
,
P.
,
Inati
,
S. J.
,
Evans
,
J. W.
,
Luh
,
W. M.
, &
Bandettini
,
P. A
. (
2012
).
Differentiating BOLD and non-BOLD signals in fMRI time series using multi-echo EPI
.
NeuroImage
,
60
(
3
),
1759
1770
. https://doi.org/10.1016/j.neuroimage.2011.12.028
Küppers
,
F.
,
Yun
,
S. D.
, &
Shah
,
N. J
. (
2022
).
Development of a novel 10-echo multi-contrast sequence based on EPIK to deliver simultaneous quantification of T2 and T2* with application to oxygen extraction fraction
.
Magnetic Resonance in Medicine
,
88
(
4
),
1608
1623
. https://doi.org/10.1002/mrm.29305
Lee
,
S.-P.
,
Silva
,
A. C.
,
Ugurbil
,
K.
, &
Kim
,
S.-G
. (
1999
).
Diffusion-weighted spin-echo fMRI at 9.4 T: Microvascular/tissue contribution to BOLD signal changes
.
Magnetic Resonance in Medicine
,
42
(
5
),
919
928
. https://doi.org/10.1002/(SICI)1522-2594(199911)42:5<919::AID-MRM12>3.0.CO;2-8
Lowe
,
M. J.
,
Lurito
,
J. T.
,
Mathews
,
V. P.
,
Phillips
,
M. D.
, &
Hutchins
,
G. D
. (
2000
).
Quantitative comparison of functional contrast from BOLD-weighted spin-echo and gradient-echo echoplanar imaging at 1.5 Tesla and H2 15O PET in the whole brain
.
Journal of Cerebral Blood Flow and Metabolism : Official Journal of the International Society of Cerebral Blood Flow and Metabolism
,
20
(
9
),
1331
1340
. https://doi.org/10.1097/00004647-200009000-00008
Manhard
,
M. K.
,
Berkin
,
B.
,
Liao
,
C.
,
Han
,
S.
,
Witzel
,
T.
,
Yen
,
Y.-F.
, &
Setsompop
,
K
. (
2019
).
Accelerated whole‐brain perfusion imaging using a simultaneous multislice spin‐echo and gradient‐echo sequence with joint virtual coil reconstruction
.
Magnetic Resonance in Medicine
,
82
(
3
),
973
983
. https://doi.org/10.1002/mrm.27784
Manhard
,
M. K.
,
Stockmann
,
J.
,
Liao
,
C.
,
Park
,
D.
,
Han
,
S.
,
Fair
,
M.
,
Boomen
,
M.
Van Den, Polimeni
,
J.
,
Bilgic
,
B.
, &
Setsompop
,
K
. (
2021
).
A multi-inversion multi-echo spin and gradient echo echo planar imaging sequence with low image distortion for rapid quantitative parameter mapping and synthetic image contrasts
.
Magnetic Resonance in Medicine
,
86
(
2
),
866
880
. https://doi.org/10.1002/mrm.28761
Marshall
,
I.
,
Simonotto
,
E.
,
Deary
,
I. J.
,
Maclullich
,
A.
,
Ebmeier
,
K. P.
,
Rose
,
E. J.
,
Wardlaw
,
J. M.
,
Goddard
,
N.
, &
Chappell
,
F. M
. (
2004
).
Repeatability of motor and working-memory tasks in healthy older volunteers: Assessment at functional MR imaging
.
Radiology
,
233
(
3
),
868
877
. https://doi.org/10.1148/RADIOL.2333031782
Mazziotta
,
J.
,
Toga
,
A.
,
Evans
,
A.
,
Fox
,
P.
,
Lancaster
,
J.
,
Zilles
,
K.
,
Woods
,
R.
,
Paus
,
T.
,
Simpson
,
G.
,
Pike
,
B.
,
Holmes
,
C.
,
Collins
,
L.
,
Thompson
,
P.
,
MacDonald
,
D.
,
Iacoboni
,
M.
,
Schormann
,
T.
,
Amunts
,
K.
,
Palomero-Gallagher
,
N.
,
Geyer
,
S.
, …
Mazoyer
,
B.
(
2001
).
A probabilistic atlas and reference system for the human brain: International Consortium for Brain Mapping (ICBM)
.
Philosophical Transactions of the Royal Society of London. Series B, Biological Sciences
,
356
(
1412
),
1293
1322
. https://doi.org/10.1098/RSTB.2001.0915
Newbould
,
R. D.
,
Skare
,
S. T.
,
Jochimsen
,
T. H.
,
Alley
,
M. T.
,
Moseley
,
M. E.
,
Albers
,
G. W.
, &
Bammer
,
R
. (
2007
).
Perfusion mapping with multiecho multishot parallel imaging EPI
.
Magnetic Resonance in Medicine
,
58
(
1
),
70
81
. https://doi.org/10.1002/mrm.21255
Newton
,
A. T.
,
Skinner
,
J. T.
, &
Quarles
,
C. C.
(
2014
).
Mapping of relative vessel sizes in fMRI with SAGE
. In
Proceedings of the 22nd Annual Meeting of ISMRM, Milan, Italy, 10–16 May
(Abstract #1833). https://cds.ismrm.org/protected/14MPresentations/abstracts/1833.pdf
Ni
,
W. W.
,
Christen
,
T.
, &
Zaharchuk
,
G
. (
2017
).
Benchmarking transverse spin relaxation based oxygenation measurements in the brain during hypercapnia and hypoxia
.
Journal of Magnetic Resonance Imaging : JMRI
,
46
(
3
),
704
. https://doi.org/10.1002/JMRI.25582
Norris
,
D. G.
,
Zysset
,
S.
,
Mildner
,
T.
, &
Wiggins
,
C. J
. (
2002
).
An investigation of the value of spin-echo-based fMRI using a stroop color-word matching task and EPI at 3 T
.
NeuroImage
,
15
(
3
),
719
726
. https://doi.org/10.1006/nimg.2001.1005
Olman
,
C. A.
,
Harel
,
N.
,
Feinberg
,
D. A.
,
He
,
S.
,
Zhang
,
P.
,
Ugurbil
,
K.
, &
Yacoub
,
E
. (
2012
).
Layer-specific fMRI reflects different neuronal computations at different depths in human V1
.
PLoS One
,
7
(
3
),
e32536
. https://doi.org/10.1371/JOURNAL.PONE.0032536
Oshio
,
K.
, &
Feinberg
,
D. A
. (
1991
).
GRASE (gradient-and spin-echo) imaging: A novel fast MRI technique
.
Magnetic Resonance in Medicine
,
20
(
2
),
344
349
. https://doi.org/10.1002/MRM.1910200219
Parkes
,
L. M.
,
Schwarzbach
,
J. V.
,
Bouts
,
A. A.
,
Deckers
,
R. H. R.
,
Pullens
,
P.
,
Kerskens
,
C. M.
, &
Norris
,
D. G
. (
2005
).
Quantifying the spatial resolution of the gradient echo and spin echo BOLD response at 3 Tesla
.
Magnetic Resonance in Medicine
,
54
(
6
),
1465
1472
. https://doi.org/10.1002/MRM.20712
Poser
,
B. A.
,
Versluis
,
M. J.
,
Hoogduin
,
J. M.
, &
Norris
,
D. G
. (
2006
).
BOLD contrast sensitivity enhancement and artifact reduction with multiecho EPI: Parallel-acquired inhomogeneity-desensitized fMRI
.
Magnetic Resonance in Medicine
,
55
(
6
),
1227
1235
. https://doi.org/10.1002/mrm.20900
Posse
,
S.
,
Wiese
,
S.
,
Gembris
,
D.
,
Mathiak
,
K.
,
Kessler
,
C.
,
Grosse-Ruyken
,
M. L.
,
Elghahwagi
,
B.
,
Richards
,
T.
,
Dager
,
S. R.
, &
Kiselev
,
V. G
. (
1999
).
Enhancement of BOLD-contrast sensitivity by single-shot multi-echo functional MR imaging
.
Magnetic Resonance in Medicine
,
42
(
1
),
87
97
. https://doi.org/10.1002/(SICI)1522-2594(199907)42:1<87::AID-MRM13>3.0.CO;2-O
Ragland
,
J. D.
,
Turetsky
,
B. I.
,
Gur
,
R. C.
,
Gunning-Dixon
,
F.
,
Turner
,
T.
,
Schroeder
,
L.
,
Chan
,
R.
, &
Gur
,
R. E
. (
2002
).
Working memory for complex figures: An fMRI comparison of letter and fractal n-back tasks
.
Neuropsychology
,
16
(
3
),
370
. https://doi.org/10.1037/0894-4105.16.3.370
Ragot
,
D. M.
, &
Chen
,
J. J
. (
2019
).
Characterizing contrast origins and noise contribution in spin-echo EPI BOLD at 3 T
.
Magnetic Resonance Imaging
,
57
,
328
336
. https://doi.org/10.1016/j.mri.2018.11.005
Raichle
,
M. E
. (
2015
).
The brain’s default mode network
.
Annual Review of Neuroscience
,
38
,
433
447
. https://doi.org/10.1146/ANNUREV-NEURO-071013-014030
Reeder
,
S. B.
,
Wintersperger
,
B. J.
,
Dietrich
,
O.
,
Lanz
,
T.
,
Greiser
,
A.
,
Reiser
,
M. F.
,
Glazer
,
G. M.
, &
Schoenberg
,
S. O
. (
2005
).
Practical approaches to the evaluation of signal-to-noise ratio performance with parallel imaging: Application with cardiac imaging and a 32-channel cardiac coil
.
Magnetic Resonance in Medicine
,
54
(
3
),
748
754
. https://doi.org/10.1002/MRM.20636
Robson
,
P. M.
,
Grant
,
A. K.
,
Madhuranthakam
,
A. J.
,
Lattanzi
,
R.
,
Sodickson
,
D. K.
, &
McKenzie
,
C. A
. (
2008
).
Comprehensive quantification of signal-to-noise ratio and g-factor for image-based and k-space-based parallel imaging reconstructions
.
Magnetic Resonance in Medicine : Official Journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine
,
60
(
4
),
895
. https://doi.org/10.1002/MRM.21728
Sahib
,
A. K.
,
Mathiak
,
K.
,
Erb
,
M.
,
Elshahabi
,
A.
,
Klamer
,
S.
,
Scheffler
,
K.
,
Focke
,
N. K.
, &
Ethofer
,
T
. (
2016
).
Effect of temporal resolution and serial autocorrelations in event-related functional MRI
.
Magnetic Resonance in Medicine
,
76
(
6
),
1805
1813
. https://doi.org/10.1002/mrm.26073
Schmiedeskamp
,
H.
,
Andre
,
J. B.
,
Straka
,
M.
,
Christen
,
T.
,
Nagpal
,
S.
,
Recht
,
L.
,
Thomas
,
R. P.
,
Zaharchuk
,
G.
, &
Bammer
,
R
. (
2013
).
Simultaneous perfusion and permeability measurements using combined spin- and gradient-echo MRI
.
Journal of Cerebral Blood Flow & Metabolism
,
33
(
5
),
732
743
. https://doi.org/10.1038/jcbfm.2013.10
Schmiedeskamp
,
H.
,
Peterson
,
E.
,
Maclaren
,
J.
,
O’Halloran
,
R.
,
Christen
,
T.
,
Holdsworth
,
S. J.
,
Aboussouan
,
E.
,
Grissom
,
W. A.
, &
Bammer
,
R
. (
2014
).
Multiband spin- and gradient-echo (SAGE) fMRI
. In
Proceedings of the 22nd Annual Meeting of ISMRM
,
Milan, Italy
, 10–16 May (Abstract #1503). https://cds.ismrm.org/protected/14MProceedings/PDFfiles/1503.pdf
Schmiedeskamp
,
H.
,
Straka
,
M.
, &
Bammer
,
R
. (
2012
).
Compensation of slice profile mismatch in combined spin- and gradient-echo echo-planar imaging pulse sequences
.
Magnetic Resonance in Medicine
,
67
(
2
),
378
388
. https://doi.org/10.1002/mrm.23012
Schmiedeskamp
,
H.
,
Straka
,
M.
,
Newbould
,
R. D.
,
Zaharchuk
,
G.
,
Andre
,
J. B.
,
Olivot
,
J. M.
,
Moseley
,
M. E.
,
Albers
,
G. W.
, &
Bammer
,
R
. (
2012
).
Combined spin- and gradient-echo perfusion-weighted imaging
.
Magnetic Resonance in Medicine
,
68
(
1
),
30
40
. https://doi.org/10.1002/MRM.23195
Schneider
,
W.
,
Noll
,
D. C.
, &
Cohen
,
J. D
. (
1993
).
Functional topographic mapping of the cortical ribbon in human vision with conventional MRI scanners
.
Nature
,
365
(
6442
),
150
153
. https://doi.org/10.1038/365150A0
Shin
,
S.
,
Yun
,
S. D.
, &
Shah
,
N. J
. (
2023
).
T2* quantification using multi-echo gradient echo sequences: A comparative study of different readout gradients
.
Scientific Reports
,
13
(
1
),
1
14
. https://doi.org/10.1038/s41598-023-28265-0
Sisco
,
N.
,
Keeling
,
E.
,
Borazanci
,
A.
,
Dortch
,
R.
, &
Stokes
,
A.
(
2022
).
Rapid parameter estimation for combined spin and gradient echo (SAGE) imaging
. In
Proceedings of the 31st Annual Meeting of ISMRM, London, United Kingdom, 7–12 May
(Abstract #4972). https://doi.org/10.58530/2022/4972
Skinner
,
J. T.
,
Robison
,
R. K.
,
Elder
,
C. P.
,
Newton
,
A. T.
,
Damon
,
B. M.
, &
Quarles
,
C. C
. (
2014
).
Evaluation of a multiple spin- and gradient-echo (SAGE) EPI acquisition with SENSE acceleration: Applications for perfusion imaging in and outside the brain
.
Magnetic Resonance Imaging
,
32
(
10
),
1171
1180
. https://doi.org/10.1016/j.mri.2014.08.032
Speck
,
O.
, &
Hennig
,
J
. (
1998
).
Functional imaging by I0- and T2*-parameter mapping using multi-image EPI
.
Magnetic Resonance in Medicine
,
40
(
2
),
243
248
. https://doi.org/10.1002/mrm.1910400210
Stables
,
L. A.
,
Kennan
,
R. P.
, &
Gore
,
J. C
. (
1998
).
Asymmetric spin-echo imaging of magnetically inhomogeneous systems: Theory, experiment, and numerical studies
.
Magnetic Resonance in Medicine
,
40
(
3
),
432
442
. https://doi.org/10.1002/mrm.1910400314
Stokes
,
A. M.
, &
Quarles
,
C. C
. (
2016
).
A simplified spin and gradient echo approach for brain tumor perfusion imaging
.
Magnetic Resonance in Medicine
,
75
(
1
),
356
362
. https://doi.org/10.1002/mrm.25591
Stokes
,
A. M.
,
Semmineh
,
N.
, &
Quarles
,
C. C
. (
2016
).
Validation of a T1 and T2* leakage correction method based on multiecho dynamic susceptibility contrast MRI using MION as a reference standard
.
Magnetic Resonance in Medicine
,
76
(
2
),
613
625
. https://doi.org/10.1002/mrm.25906
Stokes
,
A. M.
,
Skinner
,
J. T.
, &
Quarles
,
C. C
. (
2014
).
Assessment of a combined spin- and gradient-echo (SAGE) DSC-MRI method for preclinical neuroimaging
.
Magnetic Resonance Imaging
,
32
(
10
),
1181
1190
. https://doi.org/10.1016/J.MRI.2014.08.027
Stokes
,
A. M.
,
Skinner
,
J. T.
,
Yankeelov
,
T.
, &
Quarles
,
C. C
. (
2016
).
Assessment of a simplified spin and gradient echo (sSAGE) approach for human brain tumor perfusion imaging
.
Magnetic Resonance Imaging
,
34
(
9
),
1248
1255
. https://doi.org/10.1016/j.mri.2016.07.004
Troprès
,
I.
,
Grimault
,
S.
,
Vaeth
,
A.
,
Grillon
,
E.
,
Julien
,
C.
,
Payen
,
J.-F.
,
Lamalle
,
L.
, &
Decorps
,
M
. (
2001
).
Vessel size imaging
.
Magnetic Resonance in Medicine
,
45
(
3
),
397
408
. https://doi.org/10.1002/1522-2594(200103)45:3<397::aid-mrm1052>3.3.co;2-v
Wang
,
F.
,
Dong
,
Z.
,
Reese
,
T. G.
,
Bilgic
,
B.
,
Manhard
Katherine
, M.,
Chen
,
J.
,
Polimeni
,
J. R.
,
Wald
,
L. L.
, &
Setsompop
,
K
. (
2019
).
Echo planar time-resolved imaging (EPTI)
.
Magnetic Resonance in Medicine
,
81
(
6
),
3599
3615
. https://doi.org/10.1002/MRM.27673
Weisskoff
,
R.
,
Zuo
,
C. S.
,
Boxerman
,
J. L.
, &
Rosen
,
B. R
. (
1994
).
Microscopic susceptibility variation and transverse relaxation: Theory and experiment
.
Magnetic Resonance in Medicine
,
31
(
6
),
601
610
. https://doi.org/10.1002/mrm.1910310605
This is an open-access article distributed under the terms of the Creative Commons Attribution 4.0 International (CC BY 4.0) license, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. For a full description of the license, please visit https://creativecommons.org/licenses/by/4.0/legalcode.

Supplementary data