‘Non-BOLD fMRI’ data acquired at non-zero echo time (TE) suffer from contamination by the Blood Oxygenation Level Dependent (BOLD) signal due to the unavoidable signal decay caused by transverse relaxation. This contamination further reduces their already low inherent functional sensitivities and makes their correction essential. The Slice-Saturation Slab-Inversion Vascular Space Occupancy (SS-SI–VASO), for instance, cancels out BOLD contributions from VASO data, reflecting cerebral blood volume (CBV) changes, via a dynamic division approach. Alternatively, multi-echo (ME) data provide the possibility of extrapolating to TE=0. Acquisitions at very short TE would minimize the need for such corrections. The center-out EPI variant (‘DEPICTING’) is one such readout which allows for short TE. The ME 2D DEPICTING was compared here against a traditional ME 2D EPI for its sensitivity to functional changes in the VASO signal. The two BOLD-correction schemes were also evaluated. Clear differences in functional sensitivity were observed for the uncorrected VASO data obtained from the first echo, TE1, of the two readouts. VASO data corrected by ME extrapolation were, however, found to be almost identical in their sensitivity for detecting CBV changes for both readouts. An excessively high increase in VASO signal sensitivity observed with the dynamic division correction for both readouts revealed a near-perfect linear dependence on TE of VASO signal changes. This could be attributed to the substantial intravascular BOLD contributions at 3 T. In the present data, extravascular ΔR2* fraction was found to be around ~50–60%. ME extrapolation is, hence, recommended to avoid overestimation of functional CBV changes at commonly used TEs.

VAscular Space Occupancy (VASO) functional magnetic resonance imaging (fMRI) provides indirect measures of changes in cerebral blood volume (CBV; in units of ml blood per ml of tissue) by nulling all blood signal at the time of acquisition (Lu et al., 2003), such that the VASO signal is related to the CBV according to (Huber, Ivanov, et al., 2014):

(1)

Since its inception in 2003, the technique has evolved into many variants, including Multiple Acquisitions with Global Inversion Cycling (MAGIC) VASO, which allowed for multi-slice and whole brain imaging (Lu, Van Zijl, et al., 2004; Scouten & Constable, 2007); inflow-based VASO (iVASO) for arterial and arteriolar quantitative CBV measurements (Hua et al., 2011); or Slab-selective Inversion (SI) VASO (Jin & Kim, 2008) and Slice-Saturation Slab-Inversion (SS-SI) VASO (Huber, Ivanov, et al., 2014) for improved sensitivity. Of these, SS-SI-VASO has caught the most traction in the past few years due to its applicability in high-resolution high-field (≥7 T) layer-fMRI studies (Huber et al., 2015). The higher sensitivity and temporal resolution of SS-SI-VASO can also be exploited at lower field strengths. Corresponding implementations, however, were delayed due to the lower demand for layer-fMRI studies at 3 T. There are at present only three published SS-SI-VASO studies at 3 T (Guidi et al., 2023; Huber, Kronbichler, et al., 2023; Knudsen et al., 2023).

Alternatives to blood oxygenation level-dependent (BOLD) fMRI techniques, such as VASO fMRI, also serve as proxies for neuronal activation and help supplement a more holistic understanding of BOLD fMRI. The irony, however, is in the fact that at non-zero echo time (TE), these non-BOLD methods themselves suffer from contamination by the BOLD response (Hetzer et al., 2011; Huber et al., 2019). The functional sensitivity and accuracy of these measurements could then benefit greatly from short-TE readouts. One such readout is the Double-shot Echo Planar Imaging with Center-out Trajectories and Intrinsic NaviGation or DEPICTING (Hetzer et al., 2011). The multi-echo (ME) version of DEPICTING with very short first TE (TE1) and inter-echo time was recently found to substantially reduce BOLD contamination in pseudo-Continuous Arterial Spin Labeling (pCASL) (Alsop et al., 2015; Dai et al., 2008; Lorenz et al., 2018) measurements of cerebral blood flow (CBF) changes, while providing reliable measures of the simultaneous BOLD response (Devi et al., 2022). SS-SI-VASO generally takes care of the inherent BOLD contamination at non-zero TE by employing a dynamic division strategy, wherein the blood-nulled image is divided by its consecutive non-nulled (BOLD-weighted) control image. A complete correction for BOLD contamination, however, relies on a number of factors, most importantly being the assumption of a similar BOLD contribution in both the nulling and control condition, that is, extravascular (EV) BOLD contributions being equivalent to those of the combined extravascular plus intravascular (IV) BOLD response in the non-nulled parenchyma (Huber, Ivanov, et al., 2014).

In the present study, the feasibility of ME-DEPICTING as a potential readout for SS-SI-VASO at 3 T was investigated. Owing to its shorter TEs, a higher functional sensitivity is expected, as compared to the ME variant of the commonly used Echo Planar Imaging (EPI). All assessments of ME-DEPICTING-based SS-SI-VASO were, hence, contrasted to those of ME-EPI-based SS-SI-VASO. Apart from the provision for simultaneous BOLD and CBV measurements, the ME readouts also allowed for a comparison of the BOLD-correction strategies. In particular, a correction provided by ME extrapolation to TE=0 was tested against the correction provided by the original dynamic-division strategy.

2.1 Participants

Sixteen healthy volunteers (30 ± 5 years, 9 female) gave written informed consent before undergoing the experiments that had been approved by the Ethics Committee at the Medical Faculty of Leipzig University. All participants were right-handed and had normal or corrected-to-normal vision.

2.2 Functional paradigm

A full-field 8-Hz flickering black-and-white radial checkerboard was used for visual stimulation. Each functional cycle lasted 20 repetitions, starting with a rest block, which consisted of a blank gray screen of 12 repetitions (24 s), followed by the task block of 8 repetitions (16 s). The paradigm was programmed using Presentation (v17.2, Neurobehavioral Systems, Berkeley, CA, USA). A central, colored fixation point was present throughout the experiment. Subjects were instructed to focus on this dot and press a button whenever it changed color. Their attention was monitored by visually tracking their responses.

2.3 Magnetic resonance acquisitions

SS-SI-VASO was implemented on a 3-T MAGNETOM Skyrafit scanner (Siemens Healthineers, Erlangen, Germany) equipped with a 32-channel receive head coil. Data were obtained from 10 slices of 4 mm thickness (no slice gap, nominal in-plane resolution 3 mm × 3 mm, field of view 192 mm, matrix 64 × 64, bandwidth 2232 Hz/Px) located along the calcarine sulcus and acquired in descending slice order. Selective inversion for the VASO scans (nulling condition) was achieved over a 34-cm slab centered at the middle of the slice package through a hyperbolic secant inversion radiofrequency (RF) pulse of 8 ms duration, bandwidth-time product of 10, and RF field peak amplitude of B1 = 14 μT. The blood-nulling condition, corresponding to an inversion time, TI = 1153 ms, for an assumed longitudinal relaxation time of blood of T1,b=1664ms (Lu, Clingman, et al., 2004), was fulfilled for the 6th slice. A schematic of the SS-SI-VASO implementation is given in Figure 1 along with the sequence diagrams for ME-DEPICTING and ME-EPI. The specifications for the two two-dimensional (2D) ME readout modules were as follows:

Fig. 1.

SS-SI-VASO implementation with ME-DEPICTING and ME-EPI as imaging readouts. For the slice package centered at the calcarine sulcus, the blood nulling condition at 3 T was fulfilled for the 6th slice (in orange) of the VASO images. The interleaved acquisition of VASO and Control images was accomplished with both ME-DEPICTING and ME-EPI readouts for n = 3 echoes. The sequence diagrams of the two readouts are provided. The double-shot nature of DEPICTING results in a shorter TE when compared to ME-EPI.

Fig. 1.

SS-SI-VASO implementation with ME-DEPICTING and ME-EPI as imaging readouts. For the slice package centered at the calcarine sulcus, the blood nulling condition at 3 T was fulfilled for the 6th slice (in orange) of the VASO images. The interleaved acquisition of VASO and Control images was accomplished with both ME-DEPICTING and ME-EPI readouts for n = 3 echoes. The sequence diagrams of the two readouts are provided. The double-shot nature of DEPICTING results in a shorter TE when compared to ME-EPI.

Close modal
  • ME-DEPICTING: TE1/TE2/TE3 = 1.7 ms/10.7 ms/19.7 ms; repetition time, TR = 2 s; GRAPPA (GeneRalized Autocalibrating Partial Parallel Acquisition) factor 2.

  • ME-EPI: TE1/TE2/TE3 = 7.5 ms/20.7 ms/33.9 ms; TR = 2 s; GRAPPA factor 2; partial-Fourier factor 6/8.

Functional runs consisted of ten functional cycles (200 repetitions, i.e., 100 nulling/control condition pairs; effective TR, TReff = 4 s) and were of approximately 7 min duration. ME-EPI and ME-DEPICTING acquisitions were recorded during the same session. Their order was shuffled across participants to avoid primacy bias.

In addition to these two functional runs, two auxiliary functional runs of similar duration were recorded in one of the subjects, with ME-EPI and ME-DEPICTING specifications different from those stated above. They were:

  • (i)

    ME-EPI with 5 echoes: TE1/TE2/TE3/TE4/TE5 = 10 ms/25.3 ms/40.6 ms/55.9 ms/71.2 ms; TR = 2 s; GRAPPA factor 2; partial-Fourier factor 7/8.

  • (ii)

    High-resolution ME-DEPICTING: 15 slices, 1.7 mm isotropic nominal resolution; TE1/TE2/TE3 = 2.8 ms/15.1 ms/27.4 ms; TR = 2 s; GRAPPA factor 2.

For co-registration purposes, a 2D spoiled gradient-recalled echo (GRE) scan (1.5 mm nominal in-plane resolution; TE = 3.1 ms; TR = 1300 ms; flip angle 90°) was obtained at the start of each session, with slice geometry identical to the functional scans.

2.4 Data preprocessing and analysis

Data were preprocessed and analyzed using Statistical Parametric Mapping (SPM12; Wellcome Trust Centre for Neuroimaging, UCL, London, UK) implemented in MATLAB 2022b (The MathWorks, Natick, MA, USA), FMRIB Software Library (FSL, (Jenkinson et al., 2012)), and additional scripts written in Interactive Data Language (IDL 8.1, Exelis Visual Information Solutions, Boulder, CO, USA).

2.4.1 Preprocessing

All functional data were preprocessed in an identical manner. The data were split into VASO and BOLD datasets using ‘fslsplit’ and ‘fslmerge’. The series acquired at TE1 of each dataset was realigned, and the resulting realignment parameters applied to the corresponding remaining echoes. To allow for a meaningful comparison of functional sensitivity between the two readouts, only participants who exhibited minimal motion, defined as translation <1 mm and rotation <0.015 radians in both scans, were included in the analysis. The mean correlations between the realignment parameters of the VASO and BOLD scans in these subjects were also evaluated. The realigned time series were temporally high-pass filtered at a cut-off frequency equal to two effective functional cycles [1/(20 ×TReff)] and 3D-Gaussian filtered at a full width at half maximum (FWHM) equal to the nominal voxel size (3 mm or 1.7 mm). Extrapolated ME signals at TE = 0, Sbn(TE0) (please refer1) and Sctr(TE0), as well as the effective transverse relaxation rates, R2,bn* and R2,ctr*, of the VASO and BOLD datasets, respectively, were then extracted from the preprocessed ME image volumes acquired at TEi with voxel intensity S(TEi) via linear regression (LINFIT) of the expression lnS(TEi)=TEi×R2*+lnS(TE0).

2.4.2 Parameters

Percent VASO signal changes were then evaluated from: (i) uncorrected first echo of the blood-nulled dataset, Sbn(TE1), (ii) blood-nulled data of all echoes corrected by dynamic division, such that Sdd(TEi)=Sbn(TEi)/Sctr(TEi), and (iii) the ME-extrapolated intercept, Sbn(TE0). The simultaneously acquired BOLD signal was evaluated (i) in terms of ΔR2* from the R2,bn* and R2,ctr* regression results, and (ii) in terms of percent BOLD signal changes from the weighted summation of all echoes, Sctr(sum) (Poser et al., 2006; Posse et al., 1999). The weights were computed from Sctr(TEi) and the underlying BOLD model according to the fitted R2,ctr* of each voxel (see Eq. 6 of Poser et al., 2006).

2.4.3 Analysis

A general linear model (GLM) with the canonical hemodynamic response function (double gamma function) was implemented in IDL for statistical analyses and applied to all data.

2.4.4 Quantification

Activation-related percent signal changes were quantified based on the resulting β coefficients, such that δs=[(β1β0)/β0]100%. Percent VASO signal changes were converted into relative changes in cerebral blood volume by assuming a resting value of CBVrest = 0.05 ml/ml and negligible contribution from cerebrospinal fluid (CSF) (Scouten & Constable, 2007). The imperfect blood nulling in the remaining slices was then compensated based on Eq. 2b therein, rewritten here as:

(2)

with Q=(1Mz,IV(TI)/Mz,EV(TI))1 for non-nulled slices; Q=1 for the nulled slice; and water densities of blood, ρIV= 0.86 ml water/ml blood, and parenchyma, ρEV+IV= 0.89 ml water/ml parenchyma. Mz,IV(TI) and Mz,EV(TI) are the longitudinal magnetizations per unit of water in blood and parenchyma, respectively, at the time of acquisition; and ΔCBV is the absolute CBV change in ml blood/ml of parenchyma. For calculation of the quantitative rate changes ΔR2* in units of s−1, R2*,rest was taken directly from the estimated β0 parameter of the respective R2* time series.

2.5 Sensitivity of VASO and BOLD signal changes

Significant regions of VASO and BOLD activation were identified based on a voxel-based significance threshold of p<106, except for the high-resolution data, which owing to the lower signal to noise ratio (SNR) was thresholded at p<104. Slices cropped during the realignment process due to inter-session motion were disregarded from all scans of the particular session.

The sensitivity of the VASO response was evaluated from Sbn(TE1), Sdd(TEi) and Sbn(TE0), while that of the BOLD response was evaluated from R2,ctr* and Sctr(sum). The comparison of Sdd(TEi) and Sbn(TE0) VASO responses, inadvertently, allowed a comparison of the two BOLD correction strategies. The functional sensitivities were assessed in terms of (i) number of suprathreshold voxels, and (ii) the more reliable metric of temporal CNR (Geissler et al., 2007). The latter was evaluated from the common region of interest (ROI) between the ‘VASO activation’ obtained from the ME-extrapolated Sbn(TE0) and the ‘BOLD activation’ taken from the R2,ctr* images of both readouts. Comparisons of relative VASO or BOLD signal changes across TEs were also based on this ROI.

2.6 Extravascular and intravascular BOLD contributions

The parenchymal and extravascular BOLD contributions were estimated to evaluate the BOLD correction by dynamic division, which relies on the assumption T2,EV+IV*T2,EV*. The intravascular BOLD contribution was then estimated in terms of ΔR2*, written as ΔR2,IV*. These analyses were based on two ROIs in the nulled slice (i.e., the 6th/8th slice for standard-/high-resolution data):

  • (i)

    ROI-1 was based on common BOLD activation in the R2,ctr* data of the two readouts. ΔR2* evaluated from R2,ctr* was expected to represent parenchymal ΔR2* and is henceforth written as Δ R2,EV+IV*.

  • (ii)

    ROI-2 was based on common BOLD activation in the R2,bn* data of the two readouts. Δ R2* evaluated from R2,bn* was expected to represent extravascular ΔR2* and is henceforth written as Δ R2,EV*.

The related effective transverse relaxation times during the resting condition, T2,EV+IV*,rest and T2,EV*,rest, were also extracted. Prior to the extraction of the blood-nulled slice from the visual cortex masks (-fslslice), the multi-slice mask was multiplied by a brain mask that had been extracted from the structural image (-bet with -f 0.2) (Smith, 2002) and linearly registered with the Sbn(TE0) images of each readout (-flirt with -applyxfm, -usesqform & -noresampblur) (Jenkinson et al., 2002; Jenkinson & Smith, 2001). The relative extravascular rate change was evaluated from ROI-1 and ROI-2, such that

(3)

The effective intravascular BOLD contribution of both readouts was estimated solely from ROI-2 according to:

(4)

An expression for E2,IVactexp(TE×R2,IV*,act) is derived in the Appendix A1 (Eq. A6). Experimental values of T2,EV*,rest and T2,EV*,act were taken from mean values in ROI-2. Similarly, CBVact was calculated from δsVASO,bn(TE0) and the VASO signal from the dynamically divided middle echo δsVASO,dd(TEmid) with TEmid= TE2 = 10.7 ms and 20.7 ms for ME-DEPICTING and ME-EPI, respectively, and TEmid=TE3 = 40.6 ms for the ME-EPI with 5 echoes. The other relevant parameters had the following assumed values: T1,EV = 1330 ms (Wansapura et al., 1999); relative proton densities, ρIV = 0.87 ml water/ml blood; ρEV+IV = 0.89 ml water/ml parenchyma (Donahue et al., 2009; Herscovitch & Raichle, 1985; Lu et al., 2002). ΔR2,IV* was estimated for T2,IV*,rest varying from 15 ms to 40 ms in steps of 5 ms, that is from a more venous to a more arterial regime (Zhao et al., 2007). The influence of resolution and longer TE on ΔR2,IV* and its dependence on T2,IV*,rest was also examined for a single subject with standard versus high-resolution ME-DEPICTING data and 3-echoes versus 5-echoes ME-EPI data.

Four participants were excluded from the evaluation based on the motion criterion and another one due to minimal activation. Supplementary Figure S1 shows examples of plots of the translational and rotational displacements of an excluded and an included subject. The mean correlation between the realignment parameters of VASO and control scans over the remaining 11 participants averaged at 0.86 ± 0.07 and 0.86 ± 0.09 for EPI and DEPICTING scans, respectively (Supplementary Table S1).

3.1 Sensitivity comparison of the readouts

Significant VASO and BOLD activations were identified in all remaining participants. Examples are presented in Figure 2. The higher sensitivity of Sbn(TE1) of DEPICTING at the subject level (Fig. 2A) was confirmed for the group-averaged data (Fig. 3A). The number of significant voxels showing VASO, BOLD and their common activation (VASO BOLD, from Sbn(TE0) and R2,ctr*, respectively) for each participant can be assessed from Supplementary Table S2. The temporal VASO CNR values are shown in Figure 3B and Supplementary Table S3. Interestingly, after extrapolation to zero TE, the activated areas of both readouts became rather similar and the CNR was almost equal.

Fig. 2.

Simultaneously acquired VASO and BOLD responses in participant P1. (A) VASO activation maps obtained from the BOLD-uncorrected Sbn(TE1), BOLD-corrected Sbn(TE0), and Sdd(TE1). Note the larger activation areas obtained with ME-DEPICTING. (B) BOLD activation maps obtained from Sctr(sum) and fitted R2,ctr* data. The δsVASO (%), δsBOLD (%), and ΔR2* (s–1) in the activation maps, shown here for 3 slices, have been overlaid over the corresponding slices of the pre-processed Sbn(TE0) of the two readouts.

Fig. 2.

Simultaneously acquired VASO and BOLD responses in participant P1. (A) VASO activation maps obtained from the BOLD-uncorrected Sbn(TE1), BOLD-corrected Sbn(TE0), and Sdd(TE1). Note the larger activation areas obtained with ME-DEPICTING. (B) BOLD activation maps obtained from Sctr(sum) and fitted R2,ctr* data. The δsVASO (%), δsBOLD (%), and ΔR2* (s–1) in the activation maps, shown here for 3 slices, have been overlaid over the corresponding slices of the pre-processed Sbn(TE0) of the two readouts.

Close modal
Fig. 3.

Functional sensitivity comparison of the two readouts (group-averaged data). Results obtained with the ME-DEPICTING readout are presented in orange and those with the ME-EPI readout in blue. Differences between the two readouts are given in % (orange and blue text boxes) and their significance is marked with asterisks (*, **, and *** for p<0.05, 0.01, and 0.001, respectively). Red and blue arrows indicate sensitivity gains for ME-DEPICTING and ME-EPI, respectively. Error bars denote standard deviations across participants. (A) Sensitivity for VASO signal detection: At minimum TE, DEPICTING yielded 72% more voxels than EPI. Correction for BOLD contamination with Sbn(TE0) led to 30% and approx. 100% increases in the activated area for ME-DEPICTING and ME-EPI, respectively. Dynamic division yielded unexpectedly greater boosts of 78% and 300%, respectively. (B) VASO CNR in commonly activated regions of both readouts: Sbn(TE1) revealed a 27% higher CNR for DEPICTING compared to EPI (1.4 ± 0.3 vs. 1.1 ± 0.2). The CNR values obtained from the dynamically divided data were once again higher than those of the extrapolated data, yielding gains of 35% and 100% for ME-DEPICTING and ME-EPI, respectively. (C) BOLD CNRs based on Sctr(sum) and R2,ctr*: The CNR of ME-EPI was only a 12% higher than that of ME-DEPICTING. (D) Sctr(TEi) showing the expected increase of the BOLD sensitivity with TE for both readouts.

Fig. 3.

Functional sensitivity comparison of the two readouts (group-averaged data). Results obtained with the ME-DEPICTING readout are presented in orange and those with the ME-EPI readout in blue. Differences between the two readouts are given in % (orange and blue text boxes) and their significance is marked with asterisks (*, **, and *** for p<0.05, 0.01, and 0.001, respectively). Red and blue arrows indicate sensitivity gains for ME-DEPICTING and ME-EPI, respectively. Error bars denote standard deviations across participants. (A) Sensitivity for VASO signal detection: At minimum TE, DEPICTING yielded 72% more voxels than EPI. Correction for BOLD contamination with Sbn(TE0) led to 30% and approx. 100% increases in the activated area for ME-DEPICTING and ME-EPI, respectively. Dynamic division yielded unexpectedly greater boosts of 78% and 300%, respectively. (B) VASO CNR in commonly activated regions of both readouts: Sbn(TE1) revealed a 27% higher CNR for DEPICTING compared to EPI (1.4 ± 0.3 vs. 1.1 ± 0.2). The CNR values obtained from the dynamically divided data were once again higher than those of the extrapolated data, yielding gains of 35% and 100% for ME-DEPICTING and ME-EPI, respectively. (C) BOLD CNRs based on Sctr(sum) and R2,ctr*: The CNR of ME-EPI was only a 12% higher than that of ME-DEPICTING. (D) Sctr(TEi) showing the expected increase of the BOLD sensitivity with TE for both readouts.

Close modal

It is to be noted that even though the ROI selection relied on BOLD activation from R2,ctr* (Supplementary Table S2), the sizes of the ROI were similar even when the significant BOLD area was based on Sctr(sum) (not shown). The sensitivity difference for the detection of BOLD signal changes was found to be negligible (~7%) between the Sctr(sum) and R2,ctr* for both readouts (Fig. 3C and Supplementary Table S4). The CNR of BOLD data obtained from each Sctr(TEi) for the two readouts from the ROI is plotted in Figure 3D.

3.2 TE dependence of the BOLD correction by dynamic division

Investigation into the startling gain in VASO sensitivity of the dynamic division-corrected data revealed further increases in both the activation area (Supplementary Table S2) and VASO CNR (Supplementary Table S3) for Sdd(TE2) and Sdd(TE3) for both readouts. Significantly activated areas in two arbitrary participants are presented in Figure 4 along with corresponding maps obtained from Sbn(TE0) and Sbn(TE1). However, in case of the dynamic-division method, along with the increase in activation area with TE, the intensity of the VASO signal change, δsVASO,dd was also found to increase with TE. The increase of δsVASO,dd with TE is also evident in their cycle-averaged time courses (green line with filled squares in Fig. 5).

Fig. 4.

VASO activation maps obtained from the nulled slice in two participants (P1 and P5). The maps based on the Sbn(TE1), Sbn(TE0) and Sdd(TEi) data show BOLD-uncorrected δsVASO,bn(TE1) as well as BOLD-corrected results obtained by extrapolation to zero TE, δsVASO,bn(TE0), and by dynamic division, δsVASO,dd(TEi) for all echo times. The activation maps have been overlaid over the 6th slice of the respective pre-processed VASO images at the specified TE. For both BOLD-correction strategies, the increase of the size and strength of activation compared to uncorrected data is evident. An increase in the intensity of the VASO signal change, δsVASO,dd with TE is indicated by the growing blue/violet area for the Sdd(TE2) and Sdd(TE3) data of both readouts.

Fig. 4.

VASO activation maps obtained from the nulled slice in two participants (P1 and P5). The maps based on the Sbn(TE1), Sbn(TE0) and Sdd(TEi) data show BOLD-uncorrected δsVASO,bn(TE1) as well as BOLD-corrected results obtained by extrapolation to zero TE, δsVASO,bn(TE0), and by dynamic division, δsVASO,dd(TEi) for all echo times. The activation maps have been overlaid over the 6th slice of the respective pre-processed VASO images at the specified TE. For both BOLD-correction strategies, the increase of the size and strength of activation compared to uncorrected data is evident. An increase in the intensity of the VASO signal change, δsVASO,dd with TE is indicated by the growing blue/violet area for the Sdd(TE2) and Sdd(TE3) data of both readouts.

Close modal
Fig. 5.

Cycle-averaged time courses of percent VASO signal changes. (A) ME-DEPICTING results, (B) ME-EPI results; both obtained in participant P1 (nulled slice; see top panel in Fig. 4). Red squares: BOLD response from the control data, δsBOLD,ctr(TEi); open green squares: uncorrected VASO response, δsVASO,bn(TEi); filled green squares: VASO response corrected by dynamic division, δsVASO,dd(TEi). Yellow triangles represent the difference [δsVASO,bn(TEi)δsBOLD,ctr(TEi)]. The yellow-shaded region denotes the stimulus duration.

Fig. 5.

Cycle-averaged time courses of percent VASO signal changes. (A) ME-DEPICTING results, (B) ME-EPI results; both obtained in participant P1 (nulled slice; see top panel in Fig. 4). Red squares: BOLD response from the control data, δsBOLD,ctr(TEi); open green squares: uncorrected VASO response, δsVASO,bn(TEi); filled green squares: VASO response corrected by dynamic division, δsVASO,dd(TEi). Yellow triangles represent the difference [δsVASO,bn(TEi)δsBOLD,ctr(TEi)]. The yellow-shaded region denotes the stimulus duration.

Close modal

The time courses in Figure 5 were extracted separately for each readout from a common ROI of VASO activation in the nulled slices of ME-DEPICTING and ME-EPI data. As expected, δsVASO,bn(TEi) switches from a negative to a positive response with increasing TE for both readouts, while δsBOLD,ctr(TEi) grows with TE, as evidenced by the amplitudes of the primary signals and post-stimulus undershoots. Interestingly,δsVASO,dd(TEi) can be seen mirroring the corresponding BOLD signals. This applies to the evolution of both the signal intensities and the shapes of the time courses as a function of TE for both readouts. Another finding is that δsVASO,dd(TEi) agreed rather well with the difference between the uncorrected VASO response during the nulling condition and the BOLD response during the control condition, δsVASO,bn(TEi)δsBOLD,ctr(TEi).

Figures 6 and 7 further demonstrate the TE dependence of the dynamic division-corrected VASO data for both readouts. With a near-perfect linear increase (r20.97), the relation of |δsVASO,dd| bears a striking resemblance to the TE dependence of δsBOLD,ctr. The slopes of the linear fits of |δsVASO,dd| and δsBOLD,ctr averaged across subjects were found to be comparable between the two readouts.

Fig. 6.

Echo-time dependence of dynamic-division corrected VASO signals. Mean δsVASO,dd (A, D), calculated over the common ROI of VASO activation in the nulled slices of ME-DEPICTING (A) and ME-EPI (D), and the corresponding δcbv obtained with Eq. 2 (B, E) are plotted for all individual subjects against the corresponding TEi. δsBOLD,ctr (C, F) serves as references for the TE dependence. The subject-averaged values are shown in red (also see Supplementary Tables S5 and S6).

Fig. 6.

Echo-time dependence of dynamic-division corrected VASO signals. Mean δsVASO,dd (A, D), calculated over the common ROI of VASO activation in the nulled slices of ME-DEPICTING (A) and ME-EPI (D), and the corresponding δcbv obtained with Eq. 2 (B, E) are plotted for all individual subjects against the corresponding TEi. δsBOLD,ctr (C, F) serves as references for the TE dependence. The subject-averaged values are shown in red (also see Supplementary Tables S5 and S6).

Close modal
Fig. 7.

Linear relationships of subject-averaged VASO and BOLD responses with TE. Subject-averaged δsVASO,dd (A, C) and δsBOLD, ctr (B, D) were obtained from Sdd(TEi) and Sctr(TEi) recorded with ME-DEPICTING (A, B) and ME-EPI (C, D). The linear relations and their corresponding goodness of fit, r2, are given within the rectangular boxes. For comparison purposes, the subject-averaged δsVASO,bn(TE0) values are indicated in (A) and (C) by black diamonds on the y-axes. Residual BOLD signals extrapolated to zero TE were 0.02±0.14% and 0.07±0.03% for EPI and DEPICTING, respectively. These residual BOLD-like time-dependent signals and the corresponding δsVASO,bn(TE0) signals (for reference) for each individual participant are shown in Supplementary Figure S2.

Fig. 7.

Linear relationships of subject-averaged VASO and BOLD responses with TE. Subject-averaged δsVASO,dd (A, C) and δsBOLD, ctr (B, D) were obtained from Sdd(TEi) and Sctr(TEi) recorded with ME-DEPICTING (A, B) and ME-EPI (C, D). The linear relations and their corresponding goodness of fit, r2, are given within the rectangular boxes. For comparison purposes, the subject-averaged δsVASO,bn(TE0) values are indicated in (A) and (C) by black diamonds on the y-axes. Residual BOLD signals extrapolated to zero TE were 0.02±0.14% and 0.07±0.03% for EPI and DEPICTING, respectively. These residual BOLD-like time-dependent signals and the corresponding δsVASO,bn(TE0) signals (for reference) for each individual participant are shown in Supplementary Figure S2.

Close modal

Furthermore, the TE dependence of δsVASO,dd was also identified for the auxiliary single-subject data with 5 echoes (Fig. 8). A reduction in the extent of activation can be seen with later echoes (Fig. 8A). The amplitude of the primary signal of VASO responses grows rapidly at smaller TEs before appearing to stabilize at TE ≥ 25.3 ms. A shift in the shape of the time courses can be seen for δsVASO,dd(TEi); specifically, in a faster return to baseline compared to δsVASO,bn(TE1) and δsVASO,bn(TE0). An increase in the amplitude of the post-stimulus overshoot is also evident (Fig. 8B). Figure 9 demonstrates a similar behavior of the dynamic division-corrected data (increase of δsVASO,dd with TE) for higher-resolution data. The shape of the signals from δsVASO,dd(TE1) deviates again from that of uncorrected δsVASO,bn(TE1) and ME-extrapolated δsVASO,bn(TE0). The larger amplitude of VASO signal changes at higher resolution is also evident.

Fig. 8.

SS-SI-VASO acquisition with ME-EPI with 5 echoes. (A) Significant activation observed in the 6th slice (thresholded at p < 10–6) of the uncorrected VASO response from Sbn(TE1), the BOLD-corrected VASO responses from the extrapolated Sbn(TE0), and the dynamically divided echoes Sdd(TEi). The maps have been overlaid over the respective pre-processed VASO images at the specified TE. (B) Corresponding uncorrected (δsVASO,bn(TE1)) and corrected cycle-averaged time courses of δsVASO,bn(TE0) and δsVASO,dd(TEi) for 1i5. The shaded yellow region denotes stimulus duration.

Fig. 8.

SS-SI-VASO acquisition with ME-EPI with 5 echoes. (A) Significant activation observed in the 6th slice (thresholded at p < 10–6) of the uncorrected VASO response from Sbn(TE1), the BOLD-corrected VASO responses from the extrapolated Sbn(TE0), and the dynamically divided echoes Sdd(TEi). The maps have been overlaid over the respective pre-processed VASO images at the specified TE. (B) Corresponding uncorrected (δsVASO,bn(TE1)) and corrected cycle-averaged time courses of δsVASO,bn(TE0) and δsVASO,dd(TEi) for 1i5. The shaded yellow region denotes stimulus duration.

Close modal
Fig. 9.

ME-DEPICTING SS-SI-VASO acquisition at higher resolution. (A) Significant VASO activation (p < 10–4) as seen on the nulled slice (8th slice) of the uncorrected Sbn(TE1) (2nd column, within the black square box), the ME-extrapolated Sbn(TE0) (1st column), and the dynamically divided Sdd(TEi) (columns 3 to 5), overlaid over the corresponding unprocessed images. (B) Comparison of cycle-averaged time courses of VASO signal changes at different spatial resolutions: δsVASO,bn(TE1) (in green), δsVASO,bn(TE0) (in blue), and δsVASO,dd(TE1) (gray broken line). The signals were extracted from the region defined by activation in the Sbn(TE0) data. Shaded yellow region denotes stimulus duration.

Fig. 9.

ME-DEPICTING SS-SI-VASO acquisition at higher resolution. (A) Significant VASO activation (p < 10–4) as seen on the nulled slice (8th slice) of the uncorrected Sbn(TE1) (2nd column, within the black square box), the ME-extrapolated Sbn(TE0) (1st column), and the dynamically divided Sdd(TEi) (columns 3 to 5), overlaid over the corresponding unprocessed images. (B) Comparison of cycle-averaged time courses of VASO signal changes at different spatial resolutions: δsVASO,bn(TE1) (in green), δsVASO,bn(TE0) (in blue), and δsVASO,dd(TE1) (gray broken line). The signals were extracted from the region defined by activation in the Sbn(TE0) data. Shaded yellow region denotes stimulus duration.

Close modal

3.3 Extra- and intravascular BOLD contributions

3.3.1 Extravascular ΔR2* fraction

Averaged functional ΔR2,EV* and ΔR2,EV+IV* values resulted in estimates of the extravascular ΔR2* fraction, over the parenchymal ROI-1 (146 ± 59 voxels), of fEV = 55±15% and 50 ± 12% for ME-EPI and ME-DEPICTING, respectively. A substantial increase in this fraction was obtained in the smaller extravascular ROI-2 (45 ± 31 voxels): fEV = 62 ± 14% and 60 ± 11%, respectively. The sizes of ROI-1 and ROI-2 for each participant and the corresponding ΔR2,EV+IV*, ΔR2,EV* and fEV are provided in Table 1 for both readouts along with the T2,EV+IV*,rest and T2,EV*,rest values. T2,EV*,rest was found to be slightly shorter than their T2,EV+IV*,rest counterparts with both EPI and DEPICTING in both ROIs. An example of the voxel-wise distribution of T2,EV+IV*,rest, T2,EV*,rest, ΔR2,EV+IV*, ΔR2,EV* and the extravascular ΔR2* fraction over ROI-1 (participant P16: average fEV of 51% and 46% for EPI and DEPICTING, respectively) is shown in Figure 10.

Table 1.

Values of T2*,rest and ΔR2* in the EV+IV and EV compartments (obtained from time series of R2,ctr* and R2,bn*, respectively) and the corresponding extravascular ΔR2* fraction. Results obtained with both readouts are shown separately for ROI-1 and ROI-2 located in the blood-nulled slice.

ParticipantROI sizeParenchymaGM
T2,EV+IV*,rest(ms)ΔR2,EV+IV* (s–1)T2,EV*,rest(ms)ΔR2,EV* (s–1)Extravascular ΔR2* fraction fEV (%)
EPIDEPICTINGEPIDEPICTINGEPIDEPICTINGEPIDEPICTINGEPIDEPICTING
ROI-1: Slice 6 of functional parenchymal visual cortex mask 
P1 270 41.9 36.3 –0.69 –0.83 40.9 33.9 –0.35 –0.45 50.7 54.3 
P3 195 41.2 40.1 –0.69 –0.96 39.7 38.6 –0.45 –0.53 71.2 56.1 
P4 98 40.5 38.1 –0.81 –0.91 39.7 36.3 –0.50 –0.56 61.0 61.3 
P5 141 48.5 42.4 –0.53 –0.74 46.4 39.8 –0.18 –0.34 33.0 44.6 
P9 157 47.5 42.6 –0.46 –0.67 45.8 39.7 –0.36 –0.46 78.3 68.7 
P10 130 42.4 36.2 –0.40 –0.78 42.3 34.9 –0.21 –0.44 53.6 53.4 
P11 55 47.5 40.1 –0.52 –0.77 49.0 36.7 –0.41 –0.45 78.4 57.1 
P12 163 52.1 46.5 –0.68 –0.58 49.2 42.0 –0.28 –0.15 42.6 25.9 
P13 158 48.9 46.6 –0.50 –0.52 47.5 43.2 –0.25 –0.24 51.0 42.0 
P14 75 47.2 45.0 –0.47 –0.57 45.5 42.7 –0.17 –0.26 38.5 44.6 
P16 164 49.7 43.6 –0.45 –0.51 48.5 40.9 –0.23 –0.25 51.2 45.6 
Mean ± SD 146 ± 59 46.1 ± 3.9 41.6 ± 3.7 –0.56 ± 0.13 –0.71 ± 0.16 45 ± 3.7 39 ± 3.2 –0.31 ± 0.11 –0.38 ± 0.13 55 ± 15 50 ± 12 
ROI-2: Slice 6 of functional extravascular visual cortex mask 
P1 71 41.5 33.4 –0.86 –1.04 41.4 31.4 –0.49 –0.62 59.4 60.7 
P3 97 41.5 42.3 –0.87 –1.16 39.3 39.9 –0.58 –0.73 70.1 63.5 
P4 74 41.7 40.6 –0.88 –0.98 41.2 39.2 –0.58 –0.67 68.7 60.4 
P5 17 50.1 48.1 –0.71 –0.86 46.5 45.8 –0.35 –0.45 49.5 54.0 
P9 73 47.2 40.4 –0.57 –0.76 45.6 38.2 –0.46 –0.61 83.5 83.2 
P10 30 36.6 35.1 –0.61 –1.14 37.6 35.2 –0.37 –0.74 60.2 62.6 
P11 39 45.6 39.9 –0.59 –0.91 48.2 37.0 –0.47 –0.56 78.4 57.1 
P12 48.8 37.1 –1.39 –1.20 48.2 36.0 –0.50 –0.49 35.1 40.6 
P13 45 46.7 42.9 –0.89 –0.81 45.9 39.4 –0.45 –0.43 51.0 52.2 
P14 – – – – – – – – – – – 
P16 43 51.7 46.9 –0.55 –0.66 49.9 44.0 –0.36 –0.40 65.5 61.8 
Mean ± SD 45 ± 31 45.1 ± 4.7 40.7 ± 4.7 –0.79 ± 0.25 -0.95 ± 0.18 44 ± 4.2 38.6 ± 4.2 –0.46 ± 0.08 –0.57 ± 0.12 62 ± 14 60 ± 11 
ParticipantROI sizeParenchymaGM
T2,EV+IV*,rest(ms)ΔR2,EV+IV* (s–1)T2,EV*,rest(ms)ΔR2,EV* (s–1)Extravascular ΔR2* fraction fEV (%)
EPIDEPICTINGEPIDEPICTINGEPIDEPICTINGEPIDEPICTINGEPIDEPICTING
ROI-1: Slice 6 of functional parenchymal visual cortex mask 
P1 270 41.9 36.3 –0.69 –0.83 40.9 33.9 –0.35 –0.45 50.7 54.3 
P3 195 41.2 40.1 –0.69 –0.96 39.7 38.6 –0.45 –0.53 71.2 56.1 
P4 98 40.5 38.1 –0.81 –0.91 39.7 36.3 –0.50 –0.56 61.0 61.3 
P5 141 48.5 42.4 –0.53 –0.74 46.4 39.8 –0.18 –0.34 33.0 44.6 
P9 157 47.5 42.6 –0.46 –0.67 45.8 39.7 –0.36 –0.46 78.3 68.7 
P10 130 42.4 36.2 –0.40 –0.78 42.3 34.9 –0.21 –0.44 53.6 53.4 
P11 55 47.5 40.1 –0.52 –0.77 49.0 36.7 –0.41 –0.45 78.4 57.1 
P12 163 52.1 46.5 –0.68 –0.58 49.2 42.0 –0.28 –0.15 42.6 25.9 
P13 158 48.9 46.6 –0.50 –0.52 47.5 43.2 –0.25 –0.24 51.0 42.0 
P14 75 47.2 45.0 –0.47 –0.57 45.5 42.7 –0.17 –0.26 38.5 44.6 
P16 164 49.7 43.6 –0.45 –0.51 48.5 40.9 –0.23 –0.25 51.2 45.6 
Mean ± SD 146 ± 59 46.1 ± 3.9 41.6 ± 3.7 –0.56 ± 0.13 –0.71 ± 0.16 45 ± 3.7 39 ± 3.2 –0.31 ± 0.11 –0.38 ± 0.13 55 ± 15 50 ± 12 
ROI-2: Slice 6 of functional extravascular visual cortex mask 
P1 71 41.5 33.4 –0.86 –1.04 41.4 31.4 –0.49 –0.62 59.4 60.7 
P3 97 41.5 42.3 –0.87 –1.16 39.3 39.9 –0.58 –0.73 70.1 63.5 
P4 74 41.7 40.6 –0.88 –0.98 41.2 39.2 –0.58 –0.67 68.7 60.4 
P5 17 50.1 48.1 –0.71 –0.86 46.5 45.8 –0.35 –0.45 49.5 54.0 
P9 73 47.2 40.4 –0.57 –0.76 45.6 38.2 –0.46 –0.61 83.5 83.2 
P10 30 36.6 35.1 –0.61 –1.14 37.6 35.2 –0.37 –0.74 60.2 62.6 
P11 39 45.6 39.9 –0.59 –0.91 48.2 37.0 –0.47 –0.56 78.4 57.1 
P12 48.8 37.1 –1.39 –1.20 48.2 36.0 –0.50 –0.49 35.1 40.6 
P13 45 46.7 42.9 –0.89 –0.81 45.9 39.4 –0.45 –0.43 51.0 52.2 
P14 – – – – – – – – – – – 
P16 43 51.7 46.9 –0.55 –0.66 49.9 44.0 –0.36 –0.40 65.5 61.8 
Mean ± SD 45 ± 31 45.1 ± 4.7 40.7 ± 4.7 –0.79 ± 0.25 -0.95 ± 0.18 44 ± 4.2 38.6 ± 4.2 –0.46 ± 0.08 –0.57 ± 0.12 62 ± 14 60 ± 11 
Fig. 10.

T2*,rest, ΔR2* and fEV maps for participant P16 (6th slice, common ROI-1). Voxel-wise results for (A) T2,EV+IV*,rest and T2,EV*,rest, (B) ΔR2,EV+IV* and ΔR2,EV*, and (C) the extravascular ΔR2* fraction, fEV obtained with ME-DEPICTING (top) and ME-EPI (bottom). The maps have been overlaid over the corresponding structural slice.

Fig. 10.

T2*,rest, ΔR2* and fEV maps for participant P16 (6th slice, common ROI-1). Voxel-wise results for (A) T2,EV+IV*,rest and T2,EV*,rest, (B) ΔR2,EV+IV* and ΔR2,EV*, and (C) the extravascular ΔR2* fraction, fEV obtained with ME-DEPICTING (top) and ME-EPI (bottom). The maps have been overlaid over the corresponding structural slice.

Close modal

Increased estimates of the extravascular ΔR2* fraction were obtained from the single-subject high-resolution DEPICTING data with fEV = 80% (ROI-1, 458 voxels) and 95% (ROI-2, 237 voxels). A somewhat higher extravascular ΔR2* fraction was also obtained with the 5-echo EPI data. These results along with those reported in recent literature for the visual cortex are summarized in Table 2.

Table 2.

Summary of results from the EV+IV and EV compartments for T2*,rest and ΔR2*, and the corresponding extravascular ΔR2* fraction, fEV, reported in literature for the visual cortex and from the current work.

PublicationMethodologyTE (ms)Sample sizeVoxel size (mm3)ROIEV+IVEV
T2*,rest (ms)ΔR2* (s–1)T2*,rest (ms)ΔR2* (s–1)Extravasc. ΔR2* fraction fEV (%)
Lu and van Zijl (2005)  ME-VASO 14, 34.5, 55, 75.9 2 × 2 × 5 Single slice 45.5 ± 3.5 –0.58 ± 0.18 47.4 ± 3.1 –0.38 ± 0.10 67 ± 12 
Donahue et al. (2011)  ME GRE-BOLD EPI with/without bipolar crushers 32.7, 44.6, 57.6, 70.7 3.5 × 3.5 × 3.5 Occipitallobe  –0.74 ± 0.13  –0.52 ± 0.19 70 ± 29 
This work SS-SI-VA 1.7, 10.7, 19.7 11 3 × 3 × 4 ROI-1 41.6 ± 3.7 –0.71 ± 0.16 39.0 ± 3.2 –0.38 ± 0.13 50 ± 12 
 SO ME-DEPICTING    ROI-2 40.7 ± 4.7 –0.95 ± 0.18 38.6 ± 4.2 0.57 ± 0.12 60 ± 11 
   1.7 × 1.7 × 1.7 ROI-1 47.5 –1.03 47.2 -0.80 80 
     ROI-2 49.3 –1.13 47.2 -1.0 95 
 SS-SI-VA 7.5, 20.7, 33.9 11 3 × 3 × 4 ROI-1 46.1 ± 3.9 –0.56 ± 0.13 45.0 ± 3.7 –0.31 ± 0.11 55 ± 15 
 SO ME-EPI    ROI-2 45.1 ± 4.7 –0.79 ± 0.25 44.0 ± 4.2 –0.46 ± 0.08 62 ± 14 
  10, 25.3, 40.6, 55.9, 71.2  ROI-1 43.1 –0.54 42.6 –0.31 59 
     ROI-2 43.2 –0.59 42.8 –0.37 67 
PublicationMethodologyTE (ms)Sample sizeVoxel size (mm3)ROIEV+IVEV
T2*,rest (ms)ΔR2* (s–1)T2*,rest (ms)ΔR2* (s–1)Extravasc. ΔR2* fraction fEV (%)
Lu and van Zijl (2005)  ME-VASO 14, 34.5, 55, 75.9 2 × 2 × 5 Single slice 45.5 ± 3.5 –0.58 ± 0.18 47.4 ± 3.1 –0.38 ± 0.10 67 ± 12 
Donahue et al. (2011)  ME GRE-BOLD EPI with/without bipolar crushers 32.7, 44.6, 57.6, 70.7 3.5 × 3.5 × 3.5 Occipitallobe  –0.74 ± 0.13  –0.52 ± 0.19 70 ± 29 
This work SS-SI-VA 1.7, 10.7, 19.7 11 3 × 3 × 4 ROI-1 41.6 ± 3.7 –0.71 ± 0.16 39.0 ± 3.2 –0.38 ± 0.13 50 ± 12 
 SO ME-DEPICTING    ROI-2 40.7 ± 4.7 –0.95 ± 0.18 38.6 ± 4.2 0.57 ± 0.12 60 ± 11 
   1.7 × 1.7 × 1.7 ROI-1 47.5 –1.03 47.2 -0.80 80 
     ROI-2 49.3 –1.13 47.2 -1.0 95 
 SS-SI-VA 7.5, 20.7, 33.9 11 3 × 3 × 4 ROI-1 46.1 ± 3.9 –0.56 ± 0.13 45.0 ± 3.7 –0.31 ± 0.11 55 ± 15 
 SO ME-EPI    ROI-2 45.1 ± 4.7 –0.79 ± 0.25 44.0 ± 4.2 –0.46 ± 0.08 62 ± 14 
  10, 25.3, 40.6, 55.9, 71.2  ROI-1 43.1 –0.54 42.6 –0.31 59 
     ROI-2 43.2 –0.59 42.8 –0.37 67 

3.3.2 Estimates of ΔR2,IV*

Table 3 lists the experimentally obtained data employed in the estimation of intravascular ΔR2,IV* with Eq. 4. Subject-averaged values have been provided for the multi-subject main study with 3-echoes and standard resolution. Values, corresponding to the participant (P1) from this study have also been provided along with their values for the single-subject study with 5 echoes (ME-EPI) and higher resolution (ME-DEPICTING). All values were based on ROI-2. The CBVact (CBVrest+ΔCBV) values were obtained by assuming CBVrest = 0.05 ml/ml and are based on the Sbn(TE0) data. Subject-averaged ΔCBV values of 0.0053 ml/ml and 0.0067 ml/ml corresponded to percent CBV changes of δcbv = 13 ± 3 and 11 ± 4% for EPI and DEPICTING, respectively. The difference between the results obtained with both readouts was insignificant (paired two-tailed t-test, p = 0.08). Similarly, δcbv = 11% was obtained for the 5-echo ME-EPI while the high-resolution DEPICTING data yielded the highest CBV change, ΔCBV= 0.0135 ml/ml or δcbv = 27%.

Table 3.

List of experimental values used to estimate the intravascular ΔR2,IV* for the two readouts based on the results in ROI-2. δsVASO values were extracted from the dynamically divided data at the respective echo time.

TEmid (ms)CBVact (ml/ml)δsVASO,dd(TEmid) (%)T2,EV*,rest (ms)T2,EV*,act (ms)
ME-DEPICTING 
 11 subjects, averaged 10.7 0.0553 –0.9 38.6 39.5 
 P1 (3 × 3 × 4 mm310.7 0.0573 –1.1 31.4 32.1 
 P1 (1.7 × 1.7 × 1.7 mm315.1 0.0635 –1.0 49.0 51.5 
ME-EPI 
 11 subjects, averaged 20.7 0.0567 –1.5 44.4 45.3 
 P1 (3 echoes) 20.7 0.0579 –1.7 41.4 42.3 
 P1 (5 echoes) 40.6 0.0553 –1.3 42.8 43.5 
TEmid (ms)CBVact (ml/ml)δsVASO,dd(TEmid) (%)T2,EV*,rest (ms)T2,EV*,act (ms)
ME-DEPICTING 
 11 subjects, averaged 10.7 0.0553 –0.9 38.6 39.5 
 P1 (3 × 3 × 4 mm310.7 0.0573 –1.1 31.4 32.1 
 P1 (1.7 × 1.7 × 1.7 mm315.1 0.0635 –1.0 49.0 51.5 
ME-EPI 
 11 subjects, averaged 20.7 0.0567 –1.5 44.4 45.3 
 P1 (3 echoes) 20.7 0.0579 –1.7 41.4 42.3 
 P1 (5 echoes) 40.6 0.0553 –1.3 42.8 43.5 

The results for the intravascular ΔR2,IV* estimations are plotted in Figure 11 for assumed T2,IV*,rest values ranging from 15 to 50 ms. A nonlinear relationship between ΔR2,IV* and T2,IV*,rest is evidenced in all the plots. A steeper decline of |ΔR2,IV*| is observed in the region of venous T2,IV*,rest. The differing ΔR2,IV* contributions between the two readouts can also be seen converging at higher resting T2* values (Fig. 11A). The within-sequence comparisons demonstrate a lower intravascular contribution (i.e., smaller |ΔR2,IV*|) with higher spatial resolution (Fig. 11B) and longer TE (Fig. 11C).

Fig. 11.

Dependence of ΔR2,IV* on T2,IV*,rest. (A) Results obtained with subject-averaged input values for ME-DEPICTING (orange) and ME-EPI (blue). (B) High-resolution versus low-resolution data obtained in a single participant (P1) with ME-DEPICTING. (C) Single-subject results (P1) obtained with ME-EPI at different TE. The shaded green area represents the venous range with T2,IV*,rest ≈ 15–30 ms (Donahue et al., 2011).

Fig. 11.

Dependence of ΔR2,IV* on T2,IV*,rest. (A) Results obtained with subject-averaged input values for ME-DEPICTING (orange) and ME-EPI (blue). (B) High-resolution versus low-resolution data obtained in a single participant (P1) with ME-DEPICTING. (C) Single-subject results (P1) obtained with ME-EPI at different TE. The shaded green area represents the venous range with T2,IV*,rest ≈ 15–30 ms (Donahue et al., 2011).

Close modal

Readouts with short TE come with the promise of better sensitivity for non-BOLD contrasts (Hetzer et al., 2011; Huber et al., 2019) as BOLD contamination at TE <2 ms is expected to be minimal. Our recent pCASL study at 3 T had proven this to be the case for functional CBF changes in the visual cortex (Devi et al., 2022). The area of activation obtained with pCASL-prepared ME-DEPICTING at TE1 = 1.7 ms exceeded that obtained with ME-EPI at TE1 = 8 ms by 40%. The present study confirms the same for the VASO contrast. The sensitivity for detecting uncorrected VASO signal changes at TE1 of the blood-nulled dataset, Sbn(TE1=1.7 ms) was found to be substantially higher than that of ME-EPI, Sbn(TE1=7.5 ms). However, comparing the functional sensitivity of data BOLD-corrected by ME extrapolation to TE=0 (Fig. 3B), an almost equivalent VASO CNR was obtained for both readouts. This differs from the results of the pCASL study, wherein a higher CBF-CNR was obtained for the ME-DEPICTING acquisitions extrapolated to zero TE [Fig. 6C and 6E in Devi et al. (2022)] with approximately the same TEs as employed in the current work. Also contrary to the VASO results was a higher CNR for CBF changes obtained with TE1 data for both readouts compared to the corresponding S0 data. An explanation for the improvement of the quality of the fitted data at TE0 in SS-SI-VASO could be the fact that the data were first split into blood-nulling and control images prior to extrapolation, whereas the pCASL data were fit as a whole, with the fluctuations between the control and label data, possibly degrading the quality of the resulting S0 estimate. Nonetheless, as with the pCASL study, the sensitivity for the BOLD response obtained from the R2,ctr* and Sctr(Sum) data was found to be comparable for both readouts (Fig. 3C and Supplementary Table S4). For instance, R2,ctr* yields 1500 ± 448 suprathreshold voxels with ME-DEPICTING vs. 1420 ± 439 voxels with ME-EPI (p = 0.60; CNR 3.2 ± 0.9 vs. 3.7 ± 1.1, p = 0.05). This finding, along with the prediction of a shorter TE1 providing a better approximation of CBV (Genois et al., 2021), makes ME-DEPICTING a promising substitute of the traditional EPI readout for the simultaneous measurement of CBV changes accompanying the BOLD response.

4.1 BOLD correction by dynamic division

A much higher gain in VASO sensitivity was deduced for the TE1 data corrected for BOLD contributions by dynamic division compared to the correction by extrapolation to zero TE for both readouts. Further investigation into this remarkable improvement revealed a linear TE dependence of the BOLD-corrected δsVASO,dd data (Figs. 6A, 6D, 7A and 7C). This TE dependence bears a close resemblance to that of the BOLD data, the difference being that the latter’s intercept is close to zero (Figs. 6C, 6F, 7B and 7D). This suggests a failure of the dynamic division strategy and, hence, the presence of residual BOLD contamination in the δsVASO,dd data, as also hinted by the similar but mirrored shape of their time courses. The VASO measures obtained from extrapolated and even uncorrected data followed temporal dynamics expected for a functional CBV response: a slower return to baseline and less prominent post-stimulus transients (Lu et al., 2003; Mandeville et al., 1998). Dynamic division-corrected data, on the other hand, showed an earlier return to baseline and post-stimulus overshoots that increased in peak amplitude with TE (Figs. 5 and 8B). This finding mimics that of the BOLD response in ME acquisitions (Havlicek et al., 2017) and was even more evident in the ME-EPI data with extended echo train length (TE5 = 71.2 ms) (Fig. 8B).

4.1.1 Intravascular BOLD contribution

Relating VASO signal changes to the CBV response in SS-SI-VASO (Eq. 1) relies on the assumption of equivalent BOLD contributions in the blood-nulling and control condition (Huber, Ivanov, et al., 2014). This assumption holds in situations where the BOLD signal is mostly of extravascular origin. The extent of intravascular BOLD contributions to the BOLD signal, however, relies on a number of factors, including field strength (B0), TE, readout sequence, and diffusion weighting (Duong et al., 2003; Silvennoinen et al., 2003; Uludaǧ et al., 2009). The intravascular ΔR2* fraction is expected to decrease with B0 with predicted intravascular BOLD contributions from the microvasculature of approximately 57%, 36%, 11%, and 5% at the respective field strengths of 1.5 T, 3 T, 4 T, and 4.7 T and TE = T2,EV* (Uludaǧ et al., 2009). Experimentally obtained extravascular ΔR2* fractions are approximately 70% at 3 T (Donahue et al., 2011; Lu et al., 2003) and approximately 90% at 7 T (Cheng et al., 2015; Donahue et al., 2011) in the human visual cortex. The results from the current study are well within the range of what has been reported (Table 2). Our estimated ΔR2,b* values (Fig. 11) agree also with previous in-vitro measurements in bovine blood of varying hematocrit (Hct) levels (ΔR2,b* = –8.16 s–1,—14.3 s–1, and—16.6 s–1 for Hct = 0.21, 0.44, and 0.57, respectively), with the assumption of changes in blood oxygen saturation fraction, Y, from 0.61 to 0.73 (Zhao et al., 2007).

The observed linear dependence on TE of the VASO data corrected by dynamic division align with simulations based on a vascular anatomical model (Genois et al., 2021), wherein the intravascular BOLD contributions were taken into account. The linear dependence of δsVASO,dd on TE ranging from 0 to 30 ms as depicted in Figure 5A of Genois et al. (2021) for 3-T data with an extravascular fraction of 72% resemble our plots in Figure 6A and 6D.

Our results, however, differ from those reported in the original SS-SI-VASO paper (Huber, Ivanov, et al., 2014). The 3-T results therein, with data from two participants, found δsVASO,dd to be independent of TE (14 ms and 30 ms) in the visual cortex. This is remarkable as at a nominal resolution of 3 × 3 × 4 mm3, partial voluming effects are expected to be at play and introduce significant intravascular BOLD contributions. For instance, with the assumption of a 5% increase in oxygen metabolism, 57.5% increase in CBF, TE = 50 ms, and TR = 1s, Zhao et al. (2007) estimated an intravascular BOLD contribution of 13% from pure parenchyma, which increased to a substantial 42% for parenchyma contaminated by 2% veins. Consistently, a reduction in intravascular BOLD contributions from larger veins at higher resolution is evident in our single-subject data (Fig. 11B). Interestingly, despite an estimated 80–90% extravascular ΔR2* fraction for this data, the intravascular fraction was sufficient to bring about a TE dependence of δsVASO,dd.

A TE independence of 3-T ME-EPI data (12 ms and 48 ms) corrected by dynamic division has been reported in a recent layer-fMRI VASO study at sub-millimeter resolution (Huber, Kronbichler, et al., 2023), where contamination by larger veins is expected to be largely reduced. It should be noted that even in the absence of intravascular BOLD signal, any CBV change leads to a TE-dependent SS-SI-VASO signal induced by the TE dependence of the resting signals of corresponding (arterial or venous) blood compartments relative to the tissue signal. Consequently, an SS-SI-VASO signal increasing with TE was also found at 7 T in a study employing a resolution of 1.5 mm (Fig. 5A in Huber, Goense, et al., 2014). Quantification of CBV changes during activation at TE>0 will always be governed by such biases. Under certain circumstances, however, these biases might cancel out for TE>0, because such arterial and venous contributions tend to increase and decrease the SS-SI-VASO signal, respectively.

4.1.2 Relationship between SS-SI-VASO signals

The echo-wise evaluation indicated a remarkable similarity of amplitude and transients of δsVASO,dd and the difference signal, defined as δsVASO,bnδsBOLD,ctr (Fig. 5). Differences in the extravascular ΔR2* fraction and, subsequently, in intravascular BOLD contaminations do not explain this finding. As shown in Appendix A2, the similarity of both quantities is easily explained by the relation δsVASO,dd=(Sctrrest/Sctract)(δsVASO,bnδsBOLD,ctr) (see Eq. A9). Because SctrrestSctract holds (positive BOLD response) and given the small stimulus-related signal change (~1–2%), |δsVASO,dd| is expected to be slightly smaller than |δsVASO,bnδsBOLD,ctr|. This relationship, however, could explain the stabilization of δsVASO,dd at longer TE in our 5-echo example. Since δsBOLD,ctr contains an intravascular signal whereas δsVASO,bn does not, their difference is expected to vary faster at shorter TE, because at 3 T, the intravascular signal increases significantly more than the extravascular (Triantafyllou et al., 2011).

4.1.3 Imaging sequence-related effects

Differences between the EPI readouts employed in previous studies and the DEPICTING readout, such as different T2*-sensitivities of the specific k-space trajectories (Devi et al., 2022; Patzig et al., 2021), may contribute to the failure of the dynamic-division strategy in the present work. Although this aspect is likely to be of only secondary order because a TE dependence was introduced in our δsVASO,dd data and corresponding δcbf acquired with EPI as well (Fig. 6A, 6B, 6D and 6E), it cannot be ignored. As shown in Appendix A3, additional T2*-weighting due to the k-space trajectory would result in non-zero δsctr(TE0) and may be captured by an offset term TEt leading to an effective echo time, TEeff=TE+TEt. With the intercepts obtained in Figure 7B and 7D and the results from Table 2, Eq. A11 yields estimates of TEt ≈ 0.96 ms and 0.35 ms for DEPICTING and EPI, respectively. The larger offset TEt for DEPICTING is due to the different T2*-sensitivity of its k-space trajectory (Devi et al., 2022; Patzig et al., 2021). A significant deviation from the nominal TE results only at TE1 of DEPICTING (TEt ≈ 0.56 TE1), but no relevant bias at longer TEs and for EPI. This effect leads to the slightly lower δcbv value estimated from Sbn(TE0) of DEPICTING (11 ± 4%) compared to that of EPI (13 ± 3%) and also explains somewhat larger BOLD-like fluctuations in its extrapolated δsctr signal (cf., participants P11 and P12 in Supplementary Fig. S2). The different zero-crossing TE of the δsbn of the two readouts (Supplementary Fig. S3: approximately 11 ms and 17 ms for DEPICTING and EPI, respectively) could also be attributed to this difference in BOLD weighting at equivalent TEs.

A distinction lies in the TEs employed by earlier studies, which were comparatively longer and, hence, less likely to be significantly influenced by intravascular BOLD contributions. A consistent tendency could be the near-plateauing of δsVASO,dd values at longer TE of the ME-EPI data, while the ME-DEPICTING data with shorter TEs appear to follow an increasing trend over the entire range (Fig. 6A and 6D). This is also evident in time courses of later echoes extracted from the nulled slice of ME-EPI in Figures 5 and 9B. Further ME studies might provide additional insight into this matter. Such studies are, however, presently rare, even with SS-SI-VASO being the current workhorse of CBV-based layer-fMRI studies. Apart from the work mentioned above, all other studies used single-echo acquisitions at 7 T with 15 ms ≤ TE ≤ 28 ms (Beckett et al., 2020; de Oliveira et al., 2023; Faes et al., 2023; Guidi et al., 2016, 2020; Huber et al., 2015, 2021; Huber, Kassavetis, et al., 2023; Liu et al., 2022; Oliveira et al., 2021, 2022; Pizzuti et al., 2023) and with 27 ms at 3 T (Knudsen et al., 2023). We note that we do not expect relevant differences between the 2D readouts used here and those of the 2D Simultaneous Multi-Slice (SMS) EPI and 3D EPI variants (Huber et al., 2018), given that evaluations were exclusively based on the nulled slice.

Short TEs can also be achieved with spiral acquisitions. With recent developments in their implementation, artifact correction, and reconstruction, these can be expected to be a promising alternative to EPI sequences with Cartesian trajectories for non-BOLD fMRI (Glover, 2012; Kasper et al., 2022). A preliminary report of a two-fold improvement in temporal SNR (tSNR) compared to 3D-EPI was recently presented for VASO fMRI with spiral readouts at 7 T (Monreal-Madrigal et al., 2023).

4.2 Limitations

SS-SI-VASO was implemented in the present study for TI/TR = 1153 ms/2000 ms. This fulfills the requirements of the original paper [see Fig. 1a of Huber, Ivanov, et al. (2014)] by choosing a TI shorter than the arterial transit time of blood and a ‘period III’ (i.e., 2TR+TI) that prevents contamination with blood that was inverted more than once. Our timing, however, does not null the CSF contribution to the signal. A higher extravascular signal intensity was preferred over CSF nulling by our choice of a shorter TR compared to that required for simultaneous nulling of blood and CSF contributions (TR = 2.75 s). Pilot experiments had revealed a more focal activation with CSF nulling. However, δsVASO,bn(TE0) values within a common ROI were found to be very similar. Nonetheless, partial volume effects due to CSF contribution cannot be disregarded in our results. The TI of 1153 ms was based on a blood T1 value of 1664 ms measured from bovine blood at 3 T (Lu, Clingman, et al., 2004), which is lower than recently published T1 values (~1800 ms) of human blood (Li et al., 2016, 2017). With our approximate slice acquisition time of 71 ms, our choice of a TI might has nulled the 7th rather than the 6th slice. This is expected to result in only minimal blood contributions and, hence in significant over- or underestimations of CBV changes.

The ROI definition in the current study could also have been attributed to deviations from the original report. The voxel-based thresholding employed in the present work could bias the selection to voxels that showed the maximum percent signal changes and, hence, towards voxels with more partial voluming with blood vessels. On the other hand, the more widely employed cluster-based thresholding is sensitive to weaker and more diffuse signal changes, while suffering from lower spatial specificity (Woo et al., 2014).

The very short TE achieved with ME-DEPICTING was of primary interest for the current work and as such, 2D readouts were compared. While the error due to incompletely nulled slices was corrected for evaluating ΔCBV, such corrections are not easily obtained for ΔR2* values. Consequently, estimations of ΔR2,EV*, and hence also ΔR2,IV*, were limited to the single blood-nulled slice. Future work with 2D SMS-EPI (Huber et al., 2016) and SMS-DEPICTING or corresponding 3D readouts could solve this issue. At 7 T, 2D SMS-EPI was found to outperform 3D EPI (Poser et al., 2010) at lower resolutions due to lesser contributions from physiological noise (Huber et al., 2018).

Only a single pilot acquisition at higher resolution, which is likely to be less biased by intravascular BOLD contributions, and another one with more echoes have been provided here. Increasing the number of participants in each could further solidify these findings. Assessing the TE dependence may benefit from a further increase of the number of echoes. Extending the evaluation beyond the visual system, such as the motor cortex, is also warranted. The impact of intravascular BOLD contributions might also be investigated by combining 3-T experiments with those at 7 T, where the extravascular BOLD effect is expected to be dominant. Further experimentation is, indeed, needed to comment on suitable ranges of TE and spatial resolutions, within which the dynamic division method would be feasible at the respective field strengths. This would help identify the impact of the dynamic-division correction in SS-SI-VASO-based layer fMRI studies.

Presently, in the absence of a ground truth, δsVASO,bn(TE0) and, subsequently, δcbv values obtained by extrapolation to zero TE appear to deliver the most reliable estimations, especially, when quantification is the goal. However, the assumed mono-exponential T2* decay with TE is likely to be too simplistic and may introduce some bias.

SS-SI-VASO was implemented at 3 T with two 2D readouts. The feasibility of the ME-DEPICTING sequence as a potentially advantageous implementation compared to ME-EPI was investigated, motivated by the shorter TE1 and inter-echo time capabilities of this double-shot EPI readout. Unwanted BOLD contamination was, in fact, drastically reduced at the very short TE1 of 1.7 ms. The extrapolation to zero TE, however, resulted in equivalent VASO sensitivity for the two readouts. Interestingly, the dynamic division approach was found to bring about a higher gain in VASO sensitivity for ME-EPI than for ME-DEPICTING. VASO signal changes corrected for BOLD with this strategy were, however, found to exhibit a distinct TE- dependence. This is probably due to the influence of intravascular BOLD contributions at 3 T. The correction for BOLD contamination by extrapolation to zero TE using multiple echoes does not rely on a negligible intravascular BOLD signal and is, therefore, recommended, if an overestimation of functional CBV changes is to be avoided. Consequently, in the absence of ME data, the short TE of the DEPICTING readout still provides an alternative to conventional EPI if better sensitivity and accuracy of the VASO data are desired.

Pre-processed data for all echoes, data derived from multi-echo fitting and relevant scripts for BOLD correction by dynamic division and estimation of ΔR2,IV* are available at https://osf.io/cg7sp/.

R.D.: Conceptualization, Data Acquisition, Methodology, Software, Formal analysis, Investigation, Writing—original draft, Writing—review & editing, and Visualization. J.L.: Methodology, Software. T.M.: Conceptualization, Methodology, Software, Investigation, Writing—original draft, and Writing—review & editing. H.E.M.: Conceptualization, Investigation, Writing—original draft, Writing—review & editing, Supervision, Project administration, and Funding acquisition.

The DEPICTING sequence is registered under Hetzer S, Mildner T, and Moeller H. 2014. Magnetic resonance imaging with improved imaging contrast. US Patent 8,664,954 B2, filed March 31, 2009, and issued March 4, 2014; European Patent 2 414 861 B1, filed March 31, 2009, and issued January 28, 2015.

We thank Dr. Laurentius (Renzo) Huber for helpful discussions on the implementation of SS-SI-VASO at 3 T. We also greatly appreciate Anke Kummer, Sylvie Neubert, and Nicole Pampus for their assistance with the volunteers and ascertaining smooth operation despite COVID regulations/restrictions. And we hope to memorialize Torsten Schlumm’s contribution to the development of the ME-DEPICTING project with our dedication.

A preliminary account of this work has been presented in the Proceedings of the 31st Annual Meeting of the ISMRM, London, UK, 2022. This work has been supported by the Max Planck Society and by the International Max Planck Research School on Neuroscience of Communication: Function, Structure, and Plasticity.

Supplementary material for this article is available with the online version here: https://doi.org/10.1162/imag_a_00333.

Alsop
,
D. C.
,
Detre
,
J. A.
,
Golay
,
X.
,
Günther
,
M.
,
Hendrikse
,
J.
,
Hernandez-Garcia
,
L.
,
Lu
,
H.
,
Macintosh
,
B. J.
,
Parkes
,
L. M.
,
Smits
,
M.
,
Van Osch
,
M. J. P.
,
Wang
,
D. J. J.
,
Wong
,
E. C.
, &
Zaharchuk
,
G.
(
2015
).
Recommended implementation of arterial spin-labeled Perfusion mri for clinical applications: A consensus of the ISMRM Perfusion Study group and the European consortium for ASL in dementia
.
Magn Reson Med
,
73
,
102
116
. https://doi.org/10.1002/mrm.25197
Beckett
,
A. J. S.
,
Dadakova
,
T.
,
Townsend
,
J.
,
Huber
,
L.
,
Park
,
S.
, &
Feinberg
,
D. A.
(
2020
).
Comparison of BOLD and CBV using 3D EPI and 3D GRASE for cortical layer functional MRI at 7 T
.
Magn Reson Med
,
84
,
3128
3145
. https://doi.org/10.1002/mrm.28347
Cheng
,
Y.
,
van Zijl
,
P. C. M.
, &
Hua
,
J.
(
2015
).
Measurement of parenchymal extravascular R2* and tissue oxygen extraction fraction using multi-echo vascular space occupancy MRI at 7 T
.
NMR Biomed
,
28
,
264
271
. https://doi.org/10.1002/nbm.3250
Dai
,
W.
,
Garcia
,
D.
,
De Bazelaire
,
C.
, &
Alsop
,
D. C.
(
2008
).
Continuous flow-driven inversion for arterial spin labeling using pulsed radio frequency and gradient fields
.
Magn Reson Med
,
60
,
1488
1497
. https://doi.org/10.1002/mrm.21790
de Oliveira
,
Í. A. F.
,
Siero
,
J. C. W.
,
Dumoulin
,
S. O.
, &
van der Zwaag
,
W.
(
2023
).
Improved selectivity in 7 T digit mapping using VASO-CBV
.
Brain Topogr
,
36
,
23
31
. https://doi.org/10.1007/s10548-022-00932-x
Devi
,
R.
,
Lepsien
,
J.
,
Lorenz
,
K.
,
Schlumm
,
T.
,
Mildner
,
T.
, &
Möller
,
H. E.
(
2022
).
Multi-echo investigations of positive and negative CBF and concomitant BOLD changes
.
Neuroimage
,
263
,
119661
. https://doi.org/10.1016/j.neuroimage.2022.119661
Donahue
,
M. J.
,
Blicher
,
J. U.
,
Østergaard
,
L.
,
Feinberg
,
D. A.
,
MacIntosh
,
B. J.
,
Miller
,
K. L.
,
Günther
,
M.
, &
Jezzard
,
P.
(
2009
).
Cerebral blood flow, blood volume, and oxygen metabolism dynamics in human visual and motor cortex as measured by whole-brain multi-modal magnetic resonance imaging
.
J Cereb Blood Flow Metab
,
29
,
1856
1866
. https://doi.org/10.1038/jcbfm.2009.107
Donahue
,
M. J.
,
Hoogduin
,
H.
,
Van Zijl
,
P. C. M.
,
Jezzard
,
P.
,
Luijten
,
P. R.
, &
Hendrikse
,
J.
(
2011
).
Blood oxygenation level-dependent (BOLD) total and extravascular signal changes and ΔR2* in human visual cortex at 1.5, 3.0 and 7.0 T
.
NMR Biomed
24
,
25
34
. https://doi.org/10.1002/nbm.1552
Duong
,
T. Q.
,
Yacoub
,
E.
,
Adriany
,
G.
,
Hu
,
X.
,
Uǧurbil
,
K.
, &
Kim
,
S. G.
(
2003
).
Microvascular BOLD contribution at 4 and 7 T in the human brain: Gradient-echo and spin-echo fMRI with suppression of blood effects
.
Magn Reson Med
,
49
,
1019
1027
. https://doi.org/10.1002/mrm.10472
Faes
,
L. K.
,
De Martino
,
F.
, &
Huber
,
L.
(
2023
).
Cerebral blood volume sensitive layer-fMRI in the human auditory cortex at 7T: Challenges and capabilities
.
PLoS One
,
18
,
e280855
. https://doi.org/10.1371/journal.pone.0280855
Geissler
,
A.
,
Gartus
,
A.
,
Foki
,
T.
,
Tahamtan
,
A. R.
,
Beisteiner
,
R.
, &
Barth
,
M.
(
2007
).
Contrast-to-noise ratio (CNR) as a quality parameter in fMRI
.
J Magn Reson Imaging
,
25
,
1263
1270
. https://doi.org/10.1002/jmri.20935
Genois
,
É.
,
Gagnon
,
L.
, &
Desjardins
,
M.
(
2021
).
Modeling of vascular space occupancy and BOLD functional MRI from first principles using real microvascular angiograms
.
Magn Reson Med
,
85
,
456
468
. https://doi.org/10.1002/mrm.28429
Glover
,
G. H.
(
2012
).
Spiral imaging in fMRI
.
Neuroimage
,
62
,
706
712
. https://doi.org/10.1016/j.neuroimage.2011.10.039
Guidi
,
M.
,
Giulietti
,
G.
,
Biondetti
,
E.
,
Wise
,
R.
, &
Giove
,
F.
(
2023
).
Towards high-resolution quantitative assessment of vascular dysfunction
.
Front Phys
,
11
,
1
7
. https://doi.org/10.3389/fphy.2023.1248021
Guidi
,
M.
,
Huber
,
L.
,
Lampe
,
L.
,
Gauthier
,
C. J.
, &
Möller
,
H. E.
(
2016
).
Lamina-dependent calibrated BOLD response in human primary motor cortex
.
Neuroimage
,
141
,
250
261
. https://doi.org/10.1016/j.neuroimage.2016.06.030
Guidi
,
M.
,
Huber
,
L.
,
Lampe
,
L.
,
Merola
,
A.
,
Ihle
,
K.
, &
Möller
,
H. E.
(
2020
).
Cortical laminar resting-state signal fluctuations scale with the hypercapnic blood oxygenation level-dependent response
.
Hum Brain Mapp
,
41
,
2014
2027
. https://doi.org/10.1002/hbm.24926
Havlicek
,
M.
,
Ivanov
,
D.
,
Poser
,
B. A.
, &
Uludag
,
K.
(
2017
).
Echo-time dependence of the BOLD response transients— A window into brain functional physiology
.
Neuroimage
,
159
,
355
370
. https://doi.org/10.1016/j.neuroimage.2017.07.034
Herscovitch
,
P.
, &
Raichle
,
M. E.
(
1985
).
What is the correct value for the brain-blood partition coefficient for water?
J Cereb Blood Flow Metab
,
5
,
65
69
. https://doi.org/10.1038/jcbfm.1985.9
Hetzer
,
S.
,
Mildner
,
T.
, &
Möller
,
H. E.
(
2011
).
A Modified EPI sequence for high-resolution imaging at ultra-short echo time
.
Magn Reson Med
,
65
,
165
175
. https://doi.org/10.1002/mrm.22610
Hua
,
J.
,
Qin
,
Q.
,
Donahue
,
M. J.
,
Zhou
,
J.
,
Pekar
,
J. J.
, &
Van Zijl
,
P. C. M.
(
2011
).
Inflow-based vascular-space-occupancy (iVASO) MRI
.
Magn Reson Med
,
66
,
40
56
. https://doi.org/10.1002/mrm.22775
Huber
,
L.
(
2015
).
Mapping human brain activity by functional magnetic resonance imaging of blood volume
[Dissertation].
Leipzig University
,
227
. https://nbn-resolving.org/urn:nbn:de:bsz:15-qucosa-165252
Huber
,
L.
,
Finn
,
E. S.
,
Chai
,
Y.
,
Goebel
,
R.
,
Stirnberg
,
R.
,
Stöcker
,
T.
,
Marrett
,
S.
,
Uludag
,
K.
,
Kim
,
S. G.
,
Han
,
S. H.
,
Bandettini
,
P. A.
, &
Poser
,
B. A
.
(
2021
).
Layer-dependent functional connectivity methods
.
Progress in Neurobiology
,
207
,
101835
. https://doi.org/10.1016/j.pneurobio.2020.101835
Huber
,
L.
,
Goense
,
J.
,
Kennerley
,
A. J.
,
Ivanov
,
D.
,
Krieger
,
S. N.
,
Lepsien
,
J.
,
Trampel
,
R.
,
Turner
,
R.
, &
Möller
,
H. E.
(
2014
).
Investigation of the neurovascular coupling in positive and negative BOLD responses in human brain at 7T
.
Neuroimage
,
97
,
349
362
. https://doi.org/10.1016/j.neuroimage.2014.04.022
Huber
,
L.
,
Goense
,
J.
,
Kennerley
,
A. J.
,
Trampel
,
R.
,
Guidi
,
M.
,
Reimer
,
E.
,
Ivanov
,
D.
,
Neef
,
N.
,
Gauthier
,
C. J.
,
Turner
,
R.
, &
Möller
,
H. E.
(
2015
).
Cortical lamina-dependent blood volume changes in human brain at 7T
.
Neuroimage
,
107
,
23
33
. https://doi.org/10.1016/j.neuroimage.2014.11.046
Huber
,
L.
,
Ivanov
,
D.
,
Guidi
,
M.
,
Turner
,
R.
,
Uludag
,
K.
,
Moller
,
H. E.
, &
Poser
,
B. A.
(
2016
).
Functional cerebral blood volume mapping with simultaneous multi-slice acquisition
.
Neuroimage
,
125
,
1159
1168
. https://doi.org/10.1016/j.neuroimage.2015.10.082
Huber
,
L.
,
Ivanov
,
D.
,
Handwerker
,
D. A.
,
Marrett
,
S.
,
Guidi
,
M.
,
Uludağ
,
K.
,
Bandettini
,
P. A.
, &
Poser
,
B. A.
(
2018
).
Techniques for blood volume fMRI with VASO: From low-resolution mapping towards sub-millimeter layer-dependent applications
.
Neuroimage
,
164
,
131
143
. https://doi.org/10.1016/j.neuroimage.2016.11.039
Huber
,
L.
,
Ivanov
,
D.
,
Krieger
,
S. N.
,
Streicher
,
M. N.
,
Mildner
,
T.
,
Poser
,
B. A.
,
Möller
,
H. E.
, &
Turner
,
R.
(
2014
).
Slab-selective, BOLD-corrected VASO at 7 tesla provides measures of cerebral blood volume reactivity with high signal-to-noise ratio
.
Magn Reson Med
,
72
,
137
148
. https://doi.org/10.1002/mrm.24916
Huber
,
L.
,
Kassavetis
,
P.
,
Gulban
,
O. F.
,
Hallett
,
M.
, &
Horovitz
,
S. G.
(
2023
).
Laminar VASO fMRI in focal hand dystonia patients
.
Dystonia
,
2
,
10806
. https://doi.org/10.3389/dyst.2023.10806
Huber
,
L.
,
Kronbichler
,
L.
,
Stirnberg
,
R.
,
Ehses
,
P.
,
Stöcker
,
T.
,
Fernández-Cabello
,
S.
,
Poser
,
B. A.
, &
Kronbichler
,
M.
(
2023
).
Evaluating the capabilities and challenges of layer-fMRI VASO at 3T
.
Aperture Neuro
,
1
17
. https://doi.org/10.1101/2022.07.26.501554
Huber
,
L.
,
Uludağ
,
K.
, &
Möller
,
H. E.
(
2019
).
Non-BOLD contrast for laminar fMRI in humans: CBF, CBV, and CMRO2
.
Neuroimage
,
197
,
742
760
. https://doi.org/10.1016/j.neuroimage.2017.07.041
Jenkinson
,
M.
,
Bannister
,
P.
,
Brady
,
M.
, &
Smith
,
S.
(
2002
).
Improved optimization for the robust and accurate linear registration and motion correction of brain images
.
Neuroimage
,
17
,
825
841
. https://doi.org/10.1006/NIMG.2002.1132
Jenkinson
,
M.
,
Beckmann
,
C. F.
,
Behrens
,
T. E. J.
,
Woolrich
,
M. W.
, &
Smith
,
S. M.
(
2012
).
FSL
.
Neuroimage
,
62
,
782
790
. https://doi.org/10.1016/j.neuroimage.2011.09.015
Jenkinson
,
M.
, &
Smith
,
S.
(
2001
).
A global optimisation method for robust affine registration of brain images
.
Med Image Anal
,
5
,
143
156
. https://doi.org/10.1016/S1361-8415(01)00036-6
Jin
,
T.
, &
Kim
,
S. G.
(
2008
).
Improved cortical-layer specificity of vascular space occupancy fMRI with slab inversion relative to spin-echo BOLD at 9.4 T
.
Neuroimage
,
40
,
59
67
. https://doi.org/10.1016/j.neuroimage.2007.11.045
Kasper
,
L.
,
Engel
,
M.
,
Heinzle
,
J.
,
Mueller-Schrader
,
M.
,
Graedel
,
N. N.
,
Reber
,
J.
,
Schmid
,
T.
,
Barmet
,
C.
,
Wilm
,
B. J.
,
Stephan
,
K. E.
, &
Pruessmann
,
K. P.
(
2022
).
Advances in spiral fMRI: A high-resolution study with single-shot acquisition
.
Neuroimage
,
246
,
118738
. https://doi.org/10.1016/j.neuroimage.2021.118738
Knudsen
,
L.
,
Bailey
,
C. J.
,
Blicher
,
J. U.
,
Yang
,
Y.
,
Zhang
,
P.
, &
Lund
,
T. E.
(
2023
).
Improved sensitivity and microvascular weighting of 3T laminar fMRI with GE-BOLD using NORDIC and phase regression
.
Neuroimage
,
271
,
120011
. https://doi.org/10.1016/j.neuroimage.2023.120011
Li
,
W.
,
Grgac
,
K.
,
Huang
,
A.
,
Yadav
,
N.
,
Qin
,
Q.
, &
van Zijl
,
P. C. M.
(
2016
).
Quantitative theory for the longitudinal relaxation time of blood water
.
Magn Reson Med
,
76
,
270
281
. https://doi.org/10.1002/mrm.25875
Li
,
W.
,
Liu
,
P.
,
Lu
,
H.
,
Strouse
,
J. J.
,
van Zijl
,
P. C. M.
, &
Qin
,
Q.
(
2017
).
Fast measurement of blood T1 in the human carotid artery at 3T: Accuracy, precision, and reproducibility
.
Magn Reson Med
,
77
,
2296
2302
. https://doi.org/10.1002/mrm.26325
Liu
,
T. T.
,
Fu
,
J. Z.
,
Chai
,
Y.
,
Japee
,
S.
,
Chen
,
G.
,
Ungerleider
,
L. G.
, &
Merriam
,
E. P.
(
2022
).
Layer-specific, retinotopically-diffuse modulation in human visual cortex in response to viewing emotionally expressive faces
.
Nat Commun
13
,
6302
. https://doi.org/10.1038/s41467-022-33580-7
Lorenz
,
K.
,
Mildner
,
T.
,
Schlumm
,
T.
, &
Möller
,
H. E.
(
2018
).
Characterization of pseudo-continuous arterial spin labeling: Simulations and experimental validation
.
Magn Reson Med
,
79
,
1638
1649
. https://doi.org/10.1002/mrm.26805
Lu
,
H.
,
Clingman
,
C.
,
Golay
,
X.
, &
Van Zijl
,
P. C. M.
(
2004
).
Determining the longitudinal relaxation time (T1) of blood at 3.0 tesla
.
Magn Reson Med
,
52
,
679
682
. https://doi.org/10.1002/mrm.20178
Lu
,
H.
,
Golay
,
X.
,
Pekar
,
J. J.
, &
Van Zijl
,
P. C. M.
(
2003
).
Functional magnetic resonance imaging based on changes in vascular space occupancy
.
Magn Reson Med
,
50
,
263
274
. https://doi.org/10.1002/mrm.10519
Lu
,
H.
,
Golay
,
X.
, &
van Zijl
,
P. C. M.
(
2002
).
Intervoxel heterogeneity of event-related functional magnetic resonance imaging responses as a function of T1 weighting
.
Neuroimage
,
17
,
943
955
. https://doi.org/10.1006/nimg.2002.1206
Lu
,
H.
, &
Van Zijl
,
P. C. M.
(
2005
).
Experimental measurement of extravascular parenchymal BOLD effects and tissue oxygen extraction fractions using multi-echo VASO fMRI at 1.5 and 3.0 T
.
Magn Reson Med
,
53
,
808
816
. https://doi.org/10.1002/mrm.20379
Lu
,
H.
,
Van Zijl
,
P. C. M.
,
Hendrikse
,
J.
, &
Golay
,
X.
(
2004
).
Multiple acquisitions with global inversion cycling (MAGIC): A multislice technique for vascular-space-occupancy dependent fMRI
.
Magn Reson Med
,
51
,
9
15
. https://doi.org/10.1002/mrm.10659
Mandeville
,
J. B.
,
Marota
,
J. J. A.
,
Kosofsky
,
B. E.
,
Keltner
,
J. R.
,
Weissleder
,
R.
,
Rosen
,
B. R.
, &
Weisskoff
,
R. M.
(
1998
).
Dynamic functional imaging of relative cerebral blood volume during rat forepaw stimulation
.
Magn Reson Med
,
39
,
615
624
. https://doi.org/10.1002/mrm.1910390415
Monreal-Madrigal
,
A.
,
Kurban
,
D.
,
Laraib
,
Z.
,
Huber
,
R.
,
Ivanov
,
D.
,
Boulant
,
N.
, &
Poser
,
B. A.
(
2023
).
Combining the benefits of 3D acquisitions and spiral readouts for VASO fMRI at UHF
. In
Proceedings of the International Society for Magnetic Resonance in Medicine
. https://cds.ismrm.org/protected/22MProceedings/PDFfiles/2113.html
Oliveira
,
Í. A. F.
,
Cai
,
Y.
,
Hofstetter
,
S.
,
Siero
,
J. C. W.
,
van der Zwaag
,
W.
, &
Dumoulin
,
S. O.
(
2022
).
Comparing BOLD and VASO-CBV population receptive field estimates in human visual cortex
.
Neuroimage
,
248
,
118868
. https://doi.org/10.1016/j.neuroimage.2021.118868
Oliveira
,
Í. A. F.
,
van der Zwaag
,
W.
,
Raimondo
,
L.
,
Dumoulin
,
S. O.
, &
Siero
,
J. C. W.
(
2021
).
Comparing hand movement rate dependence of cerebral blood volume and BOLD responses at 7T
.
Neuroimage
,
226
,
117623
. https://doi.org/10.1016/j.neuroimage.2020.117623
Patzig
,
F.
,
Mildner
,
T.
,
Schlumm
,
T.
,
Müller
,
R.
, &
Möller
,
H. E.
(
2021
).
Deconvolution-based distortion correction of EPI using analytic single-voxel point-spread functions
.
Magn Reson Med
,
85
,
2445
2461
. https://doi.org/10.1002/mrm.28591
Pizzuti
,
A.
,
Huber
,
L.
(Renzo),
Gulban
,
O. F.
,
Benitez-Andonegui
,
A.
,
Peters
,
J.
, &
Goebel
,
R.
(
2023
).
Imaging the columnar functional organization of human area MT+ to axis-of-motion stimuli using VASO at 7 Tesla
.
Cerebral Cortex
,
33
,
8693
8711
. https://doi.org/10.1093/cercor/bhad151
Poser
,
B. A.
,
Koopmans
,
P. J.
,
Witzel
,
T.
,
Wald
,
L. L.
, &
Barth
,
M.
(
2010
).
Three dimensional echo-planar imaging at 7 Tesla
.
Neuroimage
,
51
,
261
266
. https://doi.org/10.1016/j.neuroimage.2010.01.108
Poser
,
B. A.
,
Versluis
,
M. J.
,
Hoogduin
,
J. M.
, &
Norris
,
D. G.
(
2006
).
BOLD contrast sensitivity enhancement and artifact reduction with multiecho EPI: Parallel-acquired inhomogeneity-desensitized fMRI
.
Magn Reson Med
,
55
,
1227
1235
. https://doi.org/10.1002/mrm.20900
Posse
,
S.
,
Wiese
,
S.
,
Gembris
,
D.
,
Mathiak
,
K.
,
Kessler
,
C.
,
Grosse-Ruyken
,
M. L.
,
Elghahwagi
,
B.
,
Richards
,
T.
,
Dager
,
S. R.
, &
Kiselev
,
V. G.
(
1999
).
Enhancement of BOLD-contrast sensitivity by single-shot multi-echo functional MR imaging
.
Magn Reson Med
,
42
,
87
97
. https://doi.org/10.1002/(SICI)1522-2594(199907)42:1<87::AID-MRM13>3.0.CO;2-O
Scouten
,
A.
, &
Constable
,
R. T.
(
2007
).
Applications and limitations of whole-brain MAGIC VASO functional imaging
.
Magn Reson Med
,
58
,
306
315
. https://doi.org/10.1002/mrm.21273
Silvennoinen
,
M. J.
,
Clingman
,
C. S.
,
Golay
,
X.
,
Kauppinen
,
R. A.
, &
Van Zijl
,
P. C. M.
(
2003
).
Comparison of the dependence of blood R2 and R2* on oxygen saturation at 1.5 and 4.7 Tesla
.
Magn Reson Med
49
,
47
60
. https://doi.org/10.1002/mrm.10355
Smith
,
S. M.
(
2002
).
Fast robust automated brain extraction
.
Hum Brain Mapp
,
17
,
143
155
. https://doi.org/10.1002/hbm.10062
Triantafyllou
,
C.
,
Wald
,
L. L.
, &
Hoge
,
R. D.
(
2011
).
Echo-time and field strength dependence of BOLD reactivity in veins and parenchyma using flow-normalized hypercapnic manipulation
.
PLoS ONE
,
6
(
9
),
e24519
. https://doi.org/10.1371/journal.pone.0024519
Uludaǧ
,
K.
,
Müller-Bierl
,
B.
, &
Uǧurbil
,
K.
(
2009
).
An integrative model for neuronal activity-induced signal changes for gradient and spin echo functional imaging
.
Neuroimage
,
48
,
150
165
. https://doi.org/10.1016/j.neuroimage.2009.05.051
Wansapura
,
J. P.
,
Holland
,
S. K.
,
Dunn
,
R. S.
, &
Ball
,
W. S.
(
1999
).
NMR relaxation times in the human brain at 3.0 Tesla
.
J Magn Reson Imaging
,
9
,
531
538
. https://doi.org/10.1002/(SICI)1522-2586(199904)9:4<531::AID-JMRI4>3.0.CO;2-L
Woo
,
C. W.
,
Krishnan
,
A.
, &
Wager
,
T. D.
(
2014
).
Cluster-extent based thresholding in fMRI analyses: Pitfalls and recommendations
.
Neuroimage
,
91
,
412
419
. https://doi.org/10.1016/j.neuroimage.2013.12.058
Zhao
,
J. M.
,
Clingman
,
C. S.
,
Närväinen
,
M. J.
,
Kauppinen
,
R. A.
, &
Van Zijl
,
P. C. M.
(
2007
).
Oxygenation and hematocrit dependence of transverse relaxation rates of blood at 3T
.
Magn Reson Med
,
58
,
592
597
. https://doi.org/10.1002/mrm.21342

Appendix A1. Intravascular BOLD Response

The signal intensities recorded during the blood-nulling (bn) and control (ctr) conditions in a parenchyma (EV+IV) voxel containing extravascular tissue and an intravascular compartment (blood) can be written as (Huber, 2015; Huber, Goense, et al., 2014; Huber, Ivanov, et al., 2014), respectively

(A1)

and

(A2)

with A12exp(TI/T1,EV)+exp(TR/T1,EV), B1exp(TR/T1,EV), C1exp(TR/T1,IV), and E2,iexp(TE/T2,i*) with i{EV+IV, EV,IV}. Vi and ρi are the volumes and the relative proton densities (compared to water), and M0 is the equilibrium magnetization. Due to the preservation of mass, VEV+IV·ρEV+IV=VEV·ρEV+CBV·ρIV, and the absolute activation-related VASO signal change can be written as:

(A3)

with αjCBVj·ρIV/VEV+IV, βj(ρEV+IVαj) and j{act,rest}, such that

(A4)

and

(A5)

Equation A4 permits an estimation of the intravascular BOLD effect (i.e., ΔR2,IV*), as a function of E2,IVrest, from

(A6)

Appendix A2. Relationship Between SS-SI-VASO Signals

The activation-related VASO-signal change with dynamic division is given by:

(A7)

whereas the difference between the uncorrected signal change in the nulling condition and the signal change in the control condition is given by:

(A8)

This leads to a ratio of the two quantities according to:

(A9)

Appendix A3. Effective TE of the Readout Sequence

With both EPI and DEPICTING, each k-space line is acquired at an individual echo time, so the effective T2*-weighting may differ from the first-order approximation based on the TE of the central k-space line. To analyze the influence from the readout trajectory, we define an effective echo time, TEeff=TE+TEt, that considers an additional term TEt due to the readout trajectory. Consideration of TEt and a first-order Taylor expansion of exp(TE×R2*), the BOLD-weighted signal change at TE =0 during the control condition is, therefore, given by

(A10)

which leads to

(A11)

We dedicate this work to the memory of Torsten Schlumm, who passed away in May 2023 at the age of 52.

1

S’ denotes a certain type of data or time series as indicated by an additional subscript (‘bn’ for blood-nulled, ‘ctr’ for control, or ‘dd’ for dynamically divided). The corresponding TE is further specified in parentheses, for example, Sbn(TEi), Sctr(TEi) or Sdd(TEi), where i indicates the echo number. Data obtained by extrapolation to zero TE or a weighted summation of data acquired at multiple TEs are denoted as ‘S(TE0)’ and ‘S(sum)’, respectively. For simplicity, ME datasets acquired during the blood-nulled and control conditions are sometimes referred to as ‘VASO data’ and ‘BOLD data’, respectively. The relative signal change expressed as percent of the resting value is denoted as ‘δs’ plus a subscript indicating the contrast (‘VASO’ or ‘BOLD’) and the data type (as above) as well as the TE information. Percent signal changes between activation and rest are therefore written as δsVASO,bn(TEi), δsBOLD,ctr(TEi) or δsVASO,dd(TEi). A comprehensive list of the notations is given in the Supplementary Material.

Author notes

*These authors contributed equally

This is an open-access article distributed under the terms of the Creative Commons Attribution 4.0 International (CC BY 4.0) license, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. For a full description of the license, please visit https://creativecommons.org/licenses/by/4.0/legalcode.

Supplementary data