Optically-pumped magnetometers (OPMs) are compact and lightweight sensors that can measure magnetic fields generated by current flow in neuronal assemblies in the brain. Such sensors enable construction of magnetoencephalography (MEG) instrumentation, with significant advantages over conventional MEG devices, including adaptability to head size, enhanced movement tolerance, lower complexity, and improved data quality. However, realising the potential of OPMs depends on our ability to perform system calibration—which means finding sensor locations, orientations, and the relationship between the sensor output and magnetic field (termed sensor gain). Such calibration is complex in OPM-MEG since, for example, OPM placement can change from subject to subject (unlike in conventional MEG where sensor locations/orientations are fixed). Here, we present two methods for calibration, both based on generating well-characterised magnetic fields across a sensor array. Our first device (the HALO) is a head mounted system that generates dipole-like fields from a set of coils. Our second (the matrix coil (MC)) generates fields using coils embedded in the walls of a magnetically shielded room. Our results show that both methods offer an accurate means to calibrate an OPM array (e.g., sensor locations within 2 mm of the ground truth) and that the calibrations produced by the two methods agree strongly with each other: reconstructed positions, orientations, and gains differ on average by 2.0 mm; 1.2° and 1.3% between HALO and MC. When applied to data from human MEG experiments, both methods offer improved signal-to-noise ratio after beamforming, suggesting that they give calibration parameters closer to the ground truth than presumed physical sensor coordinates and orientations. Both techniques are practical and easy to integrate into real-world MEG applications. This advances the field significantly closer to the routine use of OPMs for MEG recording.

Optically-Pumped Magnetometers (OPMs) have emerged as a useful means to measure magnetic fields generated by current flow in neural assemblies in the brain (see Brookes et al. (2022) for a review). These assessments of brain activity—termed magnetoencephalography (MEG)—have traditionally been made using sensors based on Superconducting Quantum Interference Devices (SQUIDs) (Baillet, 2017; Hämäläinen et al., 1993). However, SQUIDs operate at cryogenic temperatures, leading to several limitations. First, sensors must be fixed in position in an array that cannot be adapted to individual head size and results in a greater brain-to-sensor distance (and lower signal) for smaller heads (e.g., infants). Even in subjects who fit the array (i.e., adults), sensors must be distal to the head to accommodate thermal insulation, again limiting sensitivity. Second, movement of the head relative to the fixed array degrades the quality of data, meaning subjects must remain still for long periods. Finally, cryogenic systems are complex, requiring either regular filling with liquid helium and/or a helium reliquefier. In contrast, OPMs exploit the quantum properties of alkali atoms to measure the neuromagnetic field (see Schofield et al. (2023) and Tierney et al. (2019) for reviews). They do not require cryogenics and can be microfabricated into small (sugar cube-sized) packages (Schwindt et al., 2004, 2007; V. K. Shah & Wakai, 2013; V. Shah et al., 2007). This offers the possibility of a simpler sensor array that is lightweight, adapts to head shape/size (Corvilain et al., 2024; Feys et al., 2022; Hill et al., 2019; Rier et al., 2024), and enables movement during a scan (Boto et al., 2018; Holmes et al., 2023b; S. Mellor et al., 2023). Because the sensors do not require low temperatures, they can be sited closer to the head, ostensibly improving sensitivity and spatial precision (Hill et al., 2024; Iivanainen et al., 2017). These advantages suggest that OPMs could overtake SQUIDs as the fundamental building block of MEG instrumentation.

A typical OPM contains a glass cell housing a vapour of alkali atoms. Laser light is passed through the cell at a wavelength resonant with the D1 energy transition. This, coupled with circular polarisation of the laser light, enables atoms to be ‘pumped’ into single quantum state, such that their atomic magnetic moments align (Happer, 1972). In this way, the vapour gains a bulk magnetisation, effectively becoming ‘magnetic’, with coherence (across atoms) maintained via operation of the system in a spin exchange relaxation free (SERF) regime (Allred et al., 2002). Assuming the sensor is in zero magnetic field, the atoms remain trapped in their single state and can no longer absorb photons. This means the transmission of light through the cell is maximised. However, if there is an interaction between the bulk magnetisation and an external magnetic field, then the atoms change state and can once again absorb photons, causing a change in light transmission through the cell. Specifically, the polarisation of the vapour (which relates to the opacity of the cell) can be described by the Bloch equations. In this simple case, the intensity of the light passing through the cell becomes a Lorentzian function of magnetic field. When a sinusoidal ‘modulation’ field is also applied to the cell in a direction perpendicular to the laser beam (e.g., in the x-direction) at a frequency of around 1 kHz, the solution to the Bloch equations is modified such that the intensity of the light passing through the cell oscillates at the modulation frequency, with a modulation amplitude that is a linear function of the ambient x-oriented magnetic field, for a range of fields close to zero (~±1 nT) (Cohen-Tannoudji et al., 1970). When the modulation of the transmitted light intensity is measured (e.g., by lock-in detection at the modulation frequency), the system becomes a magnetometer with directional sensitivity. The use of two sinusoidally varying fields oriented at right angles to one another (both perpendicular to the laser beam) and applied in quadrature at the same frequency allows two Cartesian components of the vector field (e.g., x and y) to be measured simultaneously. The field component along the direction of the laser cannot be measured with equivalent accuracy. However, the use of two orthogonal laser beams allows simultaneous measurement of all three field components (Boto et al., 2022; V. Shah et al., 2020). OPMs are sensitive to small fields, with sensitivities ~10-25 fT/√Hz (this is higher than the ~3-5 fT/√Hz commonly achieved using SQUIDs but is typically compensated by increased signal due to the closer proximity of sensors to the brain).

Despite their promise, OPMs still have fundamental challenges to overcome prior to becoming mainstream MEG technology. One key challenge is array calibration, which for MEG applications includes defining the location of sensors (relative to each other), the precise orientation of the sensitive axis, and the relationship between the (voltage) output of the sensor and the magnetic field that passes through the cell (the sensor gain). All three must be known with high accuracy for MEG signals to be accurately recorded. In a SQUID-based MEG system, these calibration coefficients are set during system construction/installation, typically via the use of known magnetic fields from electromagnetic coils placed either inside or outside (Li et al., 2016) the system. Calibration accuracy depends on, for example, the tolerance of superconducting windings and their (fixed) placement within a cryogenic dewar, as well as deviations between the mathematical model of the field produced by the calibration coil and the actual field produced by the coil. Techniques such as harmonic model-refinements (Chella et al., 2012; Taulu et al., 2005), taking X-Ray CT scans of coil wire paths (Oyama et al., 2022), and using larger coils to reduce field model errors (Adachi et al., 2023) can be used to improve accuracy.

In OPMs, however, calibration is more challenging. At the simplest levels, this is because the relative position and orientations of the sensors can vary from subject to subject; for example, when switching sensors between different rigid helmets. However, calibration is further complicated by the physics of OPMs, which can cause changes in both gain and orientation sensitivity. Specifically, calibration of an OPM depends on three key factors: (1) manufacturing tolerances, (2) factory-programmed sensor operational settings, and (3) local operating conditions. Manufacturing tolerances include the placement of the on-sensor coils with respect to the vapour cell, the homogeneity and orthogonality of the fields these coils generate, the propagation direction of the laser beam(s), the position where the vapour cell and the laser beam intersect, the laser beam diameter, polarisation, and intensity, and vapour cell characteristics (such as buffer gas pressure). Operational settings include the relative phase and frequency of the modulation field produced by the on-sensor coils, the exact lock-in demodulation procedure, the vapour cell temperature, the algorithm used to zero the field within the vapour cell, and the laser wavelength offset settings, all of which result in light shifts and other Rubidium density factors that induce asymmetries in the resonance lineshape. Local sensor operating conditions include temporal field drift and varying background magnetic field and gradient field profiles between sites (Hill et al., 2022) which can introduce cross-axis projection errors (Borna et al., 2022) that strongly impact the orientation of the sensitive axis and the sensor gain. Sensor age and intentional (or unintentional) changes in sensor firmware or settings also have an effect. Previous generations of OPMs have used factory-based operational settings to minimise errors. However, we suggest this is inherently unreliable. Sensor positions, orientations, and gains should be independently determined during each MEG experiment to identify any deviations in sensor behaviour. Such an approach also allows for sensor locations and orientations to change from scan-to-scan (e.g., when using a flexible helmet).

In this paper, we take a newly developed 192-channel miniaturised integrated OPM-MEG system (Schofield et al., 2024) featuring sensors operating ‘as is’ directly from the factory floor (i.e., with no attempts made to optimise performance) and we introduce two independent means of calibration—both based on generation of well-characterised magnetic fields over the OPM sensor array. Our first method uses a helmet-mounted planar disc containing dipolar coils called the ‘HALO’ (QuSpin Inc.); our second method uses a matrix coil (MC) magnetic field nulling system (Holmes et al., 2023b) (similar to work by Iivanainen et al. (2022) to perform similar calibrations). We aimed to show that we can accurately determine sensor locations, orientations and gains, and in this way calibrate all sensors in an array.

The calibration was tested by setting up the array in a rigid helmet and quantitatively comparing the HALO- and MC-derived sensor locations, orientations, and gains to those obtained from the computer aided design (CAD) model of the helmet and presumed factory settings. The calibration coefficients are tested for robustness using multiple measurements with the same technique. Further, our independent calibration methods are compared against each other. Finally, we undertake a series of MEG experiments in human participants, testing the hypothesis that the accuracy of mathematical modelling of MEG data will increase following calibration (using both calibration systems).

2.1 OPM-MEG system

We used an OPM-MEG system with 64 triaxial sensors, which can each measure the magnetic field along three (notionally) orthogonal orientations (QuSpin Inc. Colorado, USA). This allows data to be collected from 192 independent channels. Each sensor head (Fig. 1A) incorporates a 87Rb vapour cell, a laser (tuned to 795 nm – the D1 transition for 87Rb) for optical pumping, on-board electromagnetic coils for field control (including generation of modulation fields), and two photodetectors for signal readout. The size of the sensor is 12.4 x 16.6 x 24.4 mm3 and each sensor weighs 4 g. Inside the sensor (Fig. 1B), a beam splitter and associated optics splits the laser output and projects two orthogonal beams through the cell (Boto et al., 2022; V. Shah et al., 2020). As noted above, field measurements are the most accurate perpendicular to the beam, and so beam 1 allows field measurement in the x- and y-orientations, while beam 2 enables measurement in the y- and z-orientations. The two beams combined offer complete assessment of the field vector (for a more complete description, see Boto et al. (2022)). The sensors were mounted in a 3D-printed helmet (Cerca Magnetics Limited, Nottingham, UK) which offers approximately even coverage of the cortex (Fig. 1C). The helmet is designed using CAD and constructed using 3D-printing, so the assumed locations of the vapour cells in each sensor are known (in principle) to an accuracy of ±1 mm. Sensor orientations are assumed to be perpendicular to the OPM housing (x, y, and z in Fig. 1A) and are therefore known (in principle) to ±1°. [In what follows, we use this CAD-derived information to compare our calibration methods; however, we emphasise that none of the assumed information was used for informing either calibration method. This means calibration is not only sensor-specific, but also scan-specific if OPM-MEG data are acquired using a flexible helmet].

Fig. 1.

OPM-MEG system. (A) Photograph of a triaxial OPM sensor. (B) Schematic diagram showing a triaxial sensor. Two orthogonal laser beams intersect the cell and enable measurement of field in three orientations from two photodiodes. (C) 3D-printed helmet in which sensors are mounted. (D) The integrated miniaturised electronics system formed as a backpack which can be worn by the subject. The white square MC panels can be seen on each face of the MSR. (E) System schematic.

Fig. 1.

OPM-MEG system. (A) Photograph of a triaxial OPM sensor. (B) Schematic diagram showing a triaxial sensor. Two orthogonal laser beams intersect the cell and enable measurement of field in three orientations from two photodiodes. (C) 3D-printed helmet in which sensors are mounted. (D) The integrated miniaturised electronics system formed as a backpack which can be worn by the subject. The white square MC panels can be seen on each face of the MSR. (E) System schematic.

Close modal

Control of the OPM requires circuitry and associated software to stabilise the cell and laser temperatures, identify and lock the laser wavelength to the D1 resonance, zero the magnetic field within the cell, generate modulation fields, and read out signals from the photodetectors (which are modulated at the same frequency as the modulation fields, and captured using lock-in-detection). Here, control of these processes was achieved using an integrated miniaturised electronics unit (see (Schofield et al., 2024) (Fig. 1D) (‘NEURO-1’, QuSpin, Colorado, USA)). The electronics unit is 36 x 20 x 6 cm in size, weighs 1.8 kg, and can be worn as a backpack to enable (for example) ambulatory studies (Fig. 1D). Each sensor head was connected to the backpack by a ribbon cable (2.2 gm-1 and 90 cm in length), and each sensor is controlled by a separate electronics card. These cards are grouped together in modules of 8. The digitised output of each sensor/module is sent to a multiplexer and then to a network card which passes all data, via ethernet, to a decoder (DAQ – sbRIO9637, National Instruments). The decoder integrates the sensor outputs with peripheral signals (e.g., those carrying the timings of the experimental paradigm) and passes everything to an acquisition PC, via a second ethernet connection. It is noteworthy that the electronics also enables three-axis closed-loop sensor operation, enabling a large dynamic range (currently ±8 nT). However, closed loop operation was not required for the work described here due to low magnetic field drifts (Hill et al., 2022). The bandwidth of the OPMs is 0–135 Hz (where the upper limit represents a 3 dB loss in sensitivity).

The sensor array and backpack were housed in a magnetically shielded room (MSR) comprising four MuMetal layers and one copper layer, which attenuate DC/low frequency and high-frequency magnetic interference fields respectively (MuRoom, Magnetic Shields Limited, Kent, UK). The MSR walls were equipped with degaussing coils to reduce any remnant magnetisation prior to data collection (Altarev et al., 2015). The MSR was also equipped with the MC (Holmes et al., 2022, 2023b). This comprised coil elements attached to the inner walls of the MSR. These elements can be independently energised using low noise current drivers (Magnetic Shields Limited, Kent, UK) such that the field from all elements sums to flexibly create controlled magnetic field patterns within the MSR (this system will be described in more detail below). A single ‘acquisition’ computer was used for OPM-MEG data acquisition. Experimental paradigms (along with associated triggers) and the MC were controlled by a second ‘stimulus computer’. A schematic diagram of the system is shown in Figure 1E.

2.2 Calibration with HALO

The ‘HALO’ comprised a circular printed circuit board (PCB) containing 16 independently controllable electromagnetic coils. Not only are PCBs cost-effective, but they can also be manufactured with a high accuracy, critical for use in calibration. Each coil was formed from 6-layers of a 45-turn planar spiral (270 turns total), with an inner radius of 3.68 mm and an outer radius of 10.94 mm. Within a single coil, each spiral layer was connected in series with the other layers by PCB vias. The coil radius was chosen such that when energised at a distance 5–10 cm from a sensor the coil creates a field that closely resembles a magnetic dipole. (Note that the large number of turns and layers means a small current can be applied, and so the relative contribution of stray field from the cables feeding current to the coil is minimised). 12 of these dipolar coils were equidistantly spaced (every 30°) around a circle of radius 12.7 cm; a further 3 dipoles were spaced at 120° intervals around a circle of 6.3 cm radius; and the final dipole was placed at the centre of the circle (see Fig. 2 (left)). This layout was chosen such that the system generated fields approximately evenly across the head. The HALO was electrically coupled to a driving circuit (QuSpin Inc.).

Fig. 2.

HALO calibration system. Left: Aerial view of the HALO. The system contains 16 independently controllable magnetic dipoles which make spatially distinct field patterns across the OPM array. Right: The HALO in position in the rigid OPM-MEG helmet being worn by a participant.

Fig. 2.

HALO calibration system. Left: Aerial view of the HALO. The system contains 16 independently controllable magnetic dipoles which make spatially distinct field patterns across the OPM array. Right: The HALO in position in the rigid OPM-MEG helmet being worn by a participant.

Close modal

The procedure for running the HALO was as follows. First, the HALO was mounted in the OPM helmet on a base which holds the disc 5 cm above the helmet, as shown in Figure 2 (right). Following this, the door to the MSR was closed and the room demagnetised. Each of the 16 dipoles was sequentially energised, to create 16 (spatially different) dipolar field patterns across the sensor array. The driving circuit generated a 10 Hz oscillating current which was applied to one dipole for 1 s before being switched to the next dipole. Once all 16 dipoles had been energised, the current was turned off. One limitation of dipoles is that the magnetic field falls very rapidly with distance, and so while nearby OPMs see a large field, distal OPMs see a small field. To overcome this, the sequence was run 4 times with dipole moments of 4 µAm2, 11.8 µAm2, 28.3 µAm2, and 41.0 µAm2 with a 5-s break between sequences. The procedure lasted ~90 s, and OPM data were collected throughout.

An algorithm was generated to calibrate each OPM based on the model of the field generated by the HALO. Assuming dipole n, is at location pn relative to the centre of the HALO and that a sensor is located at r (also relative to the centre of the HALO) and makes a single measurement along an axis defined by the unit vector o^. The sensor measurement, Vn(r,o^) (i.e., in volts), can be modelled by the equation

[1]

Here, m is the magnetic dipole moment (whose orientation is perpendicular to the plane of the HALO, and magnitude is assumed to be the same for all dipoles), δ=||rp|| is the Euclidean distance from the coil origin to the measurement location, r. g represents the sensor gain (in units of V/T). This equation allows us to model the fields generated by any dipole in the HALO, at any location and orientation across the array. A single sensor makes three measurements of field ([Vno^1,Vno^2,Vno^3] which represent the field magnitude along three axes, o^1, o^2 and o^3). By measuring field from all 16 dipoles, a single sensor makes 48 separate field measurements, and this is repeated 4 times (due to the 4 different dipole amplitudes) making a total of 192 measurements per sensor. By fitting these values to a model generated by the repeated use of Eq. [1], it becomes possible to determine the free parameters comprising the sensor location (r), the 3 measurement orientations (o^1, o^2 and o^3), and the 3 gains (g1, g2 and g3). (In practice, we defined orientation vectors such that their the magnitude represented the gain and the normalised unit vector is the orientation, hence only 12 free parameters are needed not 15). This offers a complete calibration of an OPM.

In practice, for each signal, the amplitudes and direction of the measured signals (relative to the phase of the coil current) were estimated via a fast Fourier transform. To avoid the problem of signals being too small in distal sensors, or so large in proximal sensors that cross-axis projection error (CAPE) effects and sensor gain errors would confound the calibration, only segments of data where the measured 10 Hz signal was between 1 pT and 1000 pT in amplitude were used in the fitting algorithm. The fitting procedure was implemented in MATLAB using the ‘fmincon’ function, which minimises the sum of the squared differences between the dipole model (of fields generated by the HALO) and measured data in such a way to determine the optimum sensor location, orientations, and gains. The initial guess for the sensor position was the same for each sensor and was approximately the centre of mass of the helmet. The initial channel gain was set to the assumed 2.7 V/nT, and each channel orientation was set in the front-back direction of the helmet. [Note – as shown in Fig. 1B the two laser beams, and hence the precise locations of the field measurements for the x and z field components are offset from one another by ~600 µm. However, this small difference was ignored in the fitting algorithm].

2.3 Calibration with matrix coils

The MC comprises 94 independently controllable square coils of side length 450 mm arranged (approximately) symmetrically in 4 x 4 grids on each MSR face. (Due to access requirements, coils on the door face are arranged asymmetrically and 2 coils are removed on the face of the MSR opposite the door because a projector porthole blocks their positioning). We used the MC elements to generate known fields by which to calibrate the OPMs. However, to do this we need to know what field each coil element produces—this is termed the MC forward model.

2.3.1 Determining the MC forward model

One could, in principle, use the positions of the wire paths of each coil to generate a forward model to ensure accurate knowledge of the fields generated by each MC element. However, interactions between the wire paths and the MuMetal walls of the MSR (Hammond, 1960; Roshen, 1990), complexities in monitoring many coil currents, and potential inaccuracies involved in mounting the coils means we chose instead to use a data-driven approach, similar to that originally described by Iivanainen et al. (2022). Briefly, Iivanainen et al. (2022) measured the fields produced from a system of 18 coils placed inside a cylindrical magnetic shield (MSC) over a volume containing an OPM-MEG helmet. By fitting the fields to a vector spherical harmonic model, they obtained a forward model describing the field generated by each coil, within a volume of interest, per unit of applied current. This approach has also been used to measure remnant fields inside MSRs (Holmes et al., 2022; S. J. Mellor et al., 2021; Rea et al., 2021) and to model both neuronal and interference fields measured by MEG systems (Taulu & Kajola, 2005; Tierney et al., 2021, 2022).

Here, to derive the data for the harmonic model, we used a triaxial Bartington Mag-13MSL100 fluxgate magnetometer (Bartington Instruments, Witney, UK) integrated with an optical tracking system (Natural Point Inc., Corvallis, OR, USA) which uses infrared light to determine the position of retroreflective markers placed onto the fluxgate. (The use of multiple markers enables measurement of the sensor orientation) (see Rea et al. (2021) for details). We used a wooden board and marked a 40 x 40 cm2 plane in 10-cm intervals which could be raised or lowered using 10-cm spacers to cover a 40 x 40 x 40 cm3 volume. For each of the 125 grid positions, we placed the magnetometer in position, used the optical tracking system to get its position and orientation, and then drove each of the 94 coils at a different frequency. We chose frequencies between 2 and 11.5 Hz with a separation of 0.1 Hz, recording for 20 s for sufficient frequency resolution (0.05 Hz) to resolve all peaks. We used an NI-cDAQ-9179 (National Instruments) to control both the series of NI-9264 16-bit digital to analogue converters which generated the voltage signals applied to the coil drivers and the NI-9205 analogue to digital converters which collected the data. We used the MATLAB NI-DAQmx package function ‘readwrite’ to simultaneously generate and record signals with a sampling frequency of 10 kHz. Following data collection, we moved the fluxgate to the next position until the whole volume had been sampled. Once data had been collected, we extracted the direction and magnitude of the field generated at each grid point by each coil. We then fitted these data to a 4th-order regular spherical harmonic model. The RMS error between the data and the model was <1% for all coils.

2.3.2 OPM calibration procedure

To calibrate the OPMs, we used the same signal used to derive the MC forward model, with a maximum field generated of around 1000 pT. We took the FFT of the resulting OPM data, to quantify the field detected by the OPMs from each coil (based on the known coil frequencies) and minimised the difference between the measured data (vn at channel n) and the forward model to determine sensor locations orientations and gains. Mathematically,

[2]

for each channel. Here, Bn(r) denotes the field from each matrix coil from the MC forward model. We reduced the number of coils used from 94 to 12 (2 randomly selected from each face) and set a (empirically derived) threshold level for optimisation and randomised the selected coils in the system until this threshold was met, or 100 iterations were completed (typically this indicated an inactive sensor or a channel with high noise). This was done as using all coils together made the system perform poorly, possibly as the initial search space is difficult to navigate with many coils. We again assumed in our minimisation that each of the channels in an individual OPM sensor shared the same location and optimised the three channels of each sensor simultaneously. The initial guess for OPM position was the centre of the grid used to map the coils, with the channels again oriented along the approximate front-back alignment direction of the helmet with 2.7 V/nT gain. We note that Iivanainen et al. (2022) first used their coils to generate three orthogonal uniform fields and the complete set of five field gradients to obtain an initial guess of the sensor orientation and then position before refining with the harmonic model as above. We found that this initial step was not needed after reducing the number of coils used in the minimisation. This may be due to the larger distance between our MC and the helmet (1.4 m) compared to their experimental setup.

2.4 Validation experiments

To validate our two calibration methods (HALO and MC), we undertook a series of experiments. For each experiment, the OPMs were located in the 3D-printed helmet (Fig. 1C). All sensors were used ‘as is’ from the production line (i.e., with no in factory calibration). This means that gain and orientation sensitivity are subject to errors from manufacturing tolerances and operational settings as well as local conditions. During the sensor start-up procedure, we used the manufacturer’s software to perform a ‘field zeroing’ process, by which the on-board sensor coils were used to offset any remnant DC background field, ensuring a close to zero field within the cell for the experiment. The manufacturer’s software also carries out an additional step in which each of the three on-sensor coils is used to produce a field of known amplitude. Measurements of the amplitude and phase of the variation of transmitted beam intensity at the modulation frequency for the three different applied fields are then used to establish signal weightings that yield values of the field components oriented along the three orthogonal directions defined by the three sensor coil fields (which are approximately equivalent to the orientations of the sides of the sensor casing). These measurements are used to set the assumed sensor gain to 2.7 V/nT for all three axes. Following this, we performed our calibration procedures, as described above.

A single experiment comprised calibration with both the HALO and MC. This was repeated 11 times. For the first 5 repeats, the MSR was opened, and the HALO moved to a different position relative to the helmet (specifically it was positioned on the top, left, right, back, and front of the helmet, as shown in Fig. 3 and remained there throughout the experiment). For the remaining 6 repeats, nothing was changed between experiments (and the HALO was sited on the top of the helmet). The OPMs were always restarted between successive experiments. The order in which the HALO and MC calibration occurred was switched for each experiment.

Fig. 3.

HALO positioning. All five locations the HALO was placed within the helmet.

Fig. 3.

HALO positioning. All five locations the HALO was placed within the helmet.

Close modal

Because the HALO was inserted into one of the sensor slots in the helmet, we had to remove one sensor from the helmet, and hence the calibration process. We also removed any sensors with excessive noise levels. This left us with seven datasets with 63 out of 64 sensors, two datasets with 62 out of 64 sensors, and two datasets with 61 out of 64 sensors. Calibration was also made complicated because moving the location of the HALO (which occupies one sensor slot) meant that we were not able to repeat the calibration for every possible sensor slot in the helmet, for every experiment performed. For this reason, some slots had more repeat experiments (ranging from 11 to 4 repeats).

Following data collection and fitting (using the procedures described above) for each of the 11 runs, we had two independent system calibrations (one from the MC array and one from the HALO), both comprising sensor locations, orientations, and values of the gain for each channel. We quantitatively compared these data in three ways:

  • 1)

    Location: The only ‘ground truth’ available was the locations of the cell for each OPM, which are known from a combination of the helmet CAD file and the design of the OPM. The accuracy of these values is limited by the manufacturing tolerance of the helmet and OPM, the precise location at which the laser passes through the cell (which we assumed to be the centre), and the extent to which all sensors are correctly inserted into the helmet. Nevertheless, we expect these CAD-derived locations to be accurate to within ~1 mm. Comparison of these values to the calibration-derived locations is somewhat complicated since the CAD locations are in the coordinate space of the helmet, and the calibration locations are in the coordinate space of either the HALO or the MC. To quantitatively compare the two, we first computed the Euclidean distances from every sensor in the helmet CAD model to every other sensor. We repeated this calculation for the locations derived using both calibration methods and computed a matrix of residuals by taking the absolute value of the difference between the CAD-derived distances and the calibrated distances. We calculated the mean and standard deviation of these values for each sensor across experiments to form an estimate of the accuracy and repeatability of the location estimate for each sensor. In addition, we also used an iterative closest point (ICP)-style minimisation algorithm to find the rigid body transformation of the calibration-derived sensor locations to the corresponding CAD locations. The algorithm found the transform which minimised the mean Euclidean distance between CAD and calibration-derived locations. We then took the final mean Euclidean distance as an absolute value of the difference between the calibration and CAD-derived locations. (Note that it would also have been possible to generate a similar transform based on the known geometry of the helmet and the HALO; however, this method was not used as it was felt the ICP-like approach would lead to a lower overall error).

  • 2)

    Orientations: The assumed sensitive orientation of OPM measurements relates to the modulation field directions, which, in turn, depend on the orientation of the on-board sensor coils. These orientations should point along the axes of the casing of the OPM (see Fig. 1A). We again used the ICP algorithm to rotate/translate the calibration-derived locations to the CAD locations. We then compared the three orthogonal axes of the sensor casing (from the CAD) to the calibration-derived orientations for each sensor axis. Specifically, for all three axes and for all OPMs we measured the angle between the assumed axis and calibration-derived axis. We then compared both the stability of each method, and the agreement between methods, by calculating and plotting the mean and standard deviation across all available runs from HALO-derived angles against MC-derived angles. We reasoned that if both approaches were stable, the error-bars would be small for all axes (i.e., the angle difference similar for each run). We also reasoned that a linear relationship would suggest that the calibration methods were equivalent (i.e., we find the same sensor orientation with the HALO and the MC calibration).

  • 3)

    Gains: The gain of each OPM channel is assumed to be set using the sensor start-up procedure. We, therefore, expected our recovered gains to be centred around 2.7 V/nT. We produced equivalent plots to those generated for the orientations to compare the stability of and agreement between methods.

2.5 Human experiments: data acquisition

We undertook a series of human experiments with the aim to determine whether calibration of the OPMs leads to an improvement in OPM-MEG performance when assessing data in source space. Seven healthy participants (two identified as female, five identified as male, mean age 32 ± 7 years, all right-handed) took part in the study, which was approved by the University of Nottingham Faculty of Medicine and Health Sciences Research Ethics Committee (approval number H16122016). All participants provided written informed consent.

All subjects underwent a visuo-motor experiment designed to elicit oscillatory responses in both the beta and gamma bands, as well as evoked responses. The experimental paradigm incorporated two types of trials:

  • 1)

    Circles trials: A visual stimulus (a central, inwardly moving, maximum-contrast circular grating) was presented. The grating subtended a total visual angle of 14° and was displayed for 1.5 s. This was followed by a baseline period lasting 2 s, during which a central fixation cross was displayed. While the circle was on the screen the subject was asked to make repeated abductions of their right index finger. There were 60 circles trials per experiment. This visual stimulus is known to induce gamma oscillations in the visual cortex (Hoogenboom et al., 2006) while the finger movement is known to modulate beta oscillations in the motor cortex (Pfurtscheller & Lopes da Silva, 1999).

  • 2)

    Faces trials: In these trials, a photograph of a face was displayed on screen for a duration of 0.3 s, followed by a (jittered) rest period of duration 1 ± 0.1 s (during which a central fixation cross was shown). A total of 180 faces trials were used. This task generates evoked responses in primary visual and fusiform areas (Bentin et al., 1996; Halgren et al., 2000; Hill et al., 2022; Taylor et al., 2001).

In all cases, visual stimuli were presented via projection through a waveguide in the MSR onto a back projection screen, positioned ~100 cm in front of the subject (using a ViewSonic PX748-4K data projector).

In an experimental session, the subject was first seated on a patient support at the centre of the MSR. The OPM helmet was placed on the subject’s head, and the HALO was positioned on the top of the helmet (secured via an empty OPM slot – Fig. 3 (top)). Following this, the door to the MSR was closed and the room demagnetised (this procedure takes <1 min and reduces the remnant field inside the MSR to ~2 nT (Rea et al., 2021)). The manufacturer field-zeroing and parameter setting was performed and then calibration of the OPMs was performed using both the HALO and the MC; this process took a total of 2 min. The experimental paradigm was then run, and OPM-MEG data were collected at a sample rate of 375 Hz. The experimental paradigm lasted ~550 s, and the total experimental time (including degaussing, calibration using both techniques, and the experiment itself) was ~13 min.

Immediately following MEG data acquisition, we used an optical scanning technique to determine how the helmet was positioned on the subject’s head. Specifically, a 3D digitisation of the participant’s head (with the helmet in place) was acquired using a 3D structured light scan (Einscan H, SHINING 3D, Hangzhou, China). The 3D surface of the subject’s face was extracted from this scan and matched to the equivalent surface taken from a T1-weighted anatomical magnetic resonance image (MRI). This enabled a co-registration of the helmet relative to brain anatomy (Hill et al., 2020; Zetter et al., 2018).

2.6 Data analysis

OPM-MEG data for every channel were initially inspected by computing the power spectral densities, and any channels with a high noise floor (taken as >30 fT/√Hz in the 60–80 Hz band) or very low signal levels (<7 fT/√Hz) were removed. A trial-by-trial analysis was also carried out, whereby trials with variance greater than 3 standard deviations from the mean trial variance were automatically removed. Data were also inspected visually, and any obvious noisy channels or trials removed. A 1–150 Hz band pass filter was applied along with notch filters at the powerline frequency (50 Hz) and two of its harmonics.

We used a beamformer spatial filter (Robinson & Vrba, 1998) to process data in source space. This requires accurately calibrated data, and we used three separate approaches:

  • 1)

    Assumed calibration: Sensor locations were derived from the helmet CAD; the location of the helmet relative to the brain was derived from the structured light scan and MRI. The combination of these data allowed co-registration of sensor locations relative to the brain. The orientations were assumed to be orthogonal, and parallel to the sides of the sensor casing. The sensor gain was assumed to be 2.7 V/nT.

  • 2)

    HALO calibration: We used the ICP algorithm to co-register the HALO-derived locations to the helmet CAD, and this coupled with the structured light scan and MRI allowed sensor locations relative to the brain to be derived. For every channel, the orientation and gain were taken directly from the HALO fitting algorithm (with orientations rotated to the same space as the MRI scan). Recorded voltage data for each channel were divided by the calculated gain value for each channel to yield field data.

  • 3)

    MC calibration: This was identical to the HALO calibration but using MC-derived coefficients (locations, orientation, and gains).

Notice that for all three calibrations, the processes of co-registration were identical. Having calibrated the system in these three ways, a forward model was constructed using a single-shell volume conductor model (Nolte, 2003). Following this, we undertook three separate analyses:

2.6.1 Gamma modulation

Data collected during the circles trials were segmented to 0 s to 3.5 s windows (relative to the onset of the circle) and filtered to the 35 Hz to 55 Hz band. A covariance matrix and beamformer weights were constructed using these data. To make an image showing the spatial signature of stimulus-induced gamma change, we contrasted 35–55 Hz oscillatory power in the 0.5 s to 1.5 s (active) window to power in the 2.5 s to 3.5 s (control) window, deriving a pseudo-T-statistic (Vrba & Robinson, 2001) for all voxels on a regular 4-mm grid covering the brain. Source orientation for each voxel was taken as that with the largest signal power. Having derived the location of largest gamma change, we then constructed a ‘virtual electrode’ showing the timecourse of electrophysiological change at this peak location. This was done in two ways; first, we generated a gamma amplitude envelope timecourse by taking band-limited data for each trial, applying a Hilbert transform to get the analytic signal, and then taking the absolute value of the analytic signal and averaging over trials (this is termed the Hilbert envelope). Second, we recalculated the beamformer weights using broadband (1–150 Hz) data, and then constructed a time frequency spectrogram (TFS) by sequentially filtering data into overlapping frequency bands, deriving the Hilbert envelope for each band and concatenating these data in frequency. In both cases, an estimate of baseline amplitude was derived (independently for each frequency band) in the control window (3 s to 3.5 s) and this was subtracted; the data were also normalised by these same baseline values to give an estimate of change relative to baseline ‘noise’. Finally, gamma band signal-to-noise ratio (SNR) was measured as the change in mean gamma amplitude (between the active and control window) divided by the standard deviation of the envelope in the control window.

2.6.2 Beta modulation

Stimulus-induced change in beta power (due to finger movement) was mapped in a similar way to the gamma change. Data from Circles trials were segmented into 6-s windows (starting at the onset of the circle stimulus), filtered into the 13 Hz to 30 Hz band, and covariance and beamformer weights calculated. A pseudo-T-statistical image was generated, contrasting active (1 s to 2 s) and control (2.5 s to 3.5 s) windows. A Hilbert envelope of beta amplitude and a TFS were derived for the location of maximum beta modulation as described above (but with a baseline window of 5 s to 6 s), and SNR was calculated as the difference in mean beta amplitude in the active and control windows divided by the standard deviation in the active window.

2.6.3 Evoked responses

To look at evoked responses to faces, data were segmented to -0.24 s to 1.26 s time windows (relative to the presentation of a face) and frequency filtered to the 2 Hz to 40 Hz band. Covariance and weights were constructed using data from all faces trials. To compute the evoked response, we first used a beamformer to reconstruct a virtual electrode at an anatomically defined point in the primary visual cortex (selected according to the automated anatomical labelling (AAL) atlas (Tzourio-Mazoyer et al., 2002)). Beamformed timecourses were averaged across trials, giving the evoked response. For the peak in the evoked response at 108 ms, we generated a pseudo-Z-statistical image (which contrasts beamformer projected source amplitude (at a single point in time) to the noise amplitude (Vrba & Robinson, 2001)). The instantaneous beamformer projected amplitude was normalised by an estimate of the projected noise (taken as (wTCNw)1/2 where w are the beamformer weights and CN is the 2–40 Hz covariance estimated in a control window (1.06 s to 1.26 s)). Evoked response SNR was estimated as the maximum amplitude of the evoked response divided by the same projected noise.

Each of the above three analyses was undertaken (independently) for each of the three calibration types (assumed (CAD), HALO, and MC). We tested statistically (using Wilcoxon signed-rank test) whether SNR was increased when using either the HALO or MC compared to the assumed calibrations. We also explored the source localisation repeatability by comparing the peak location for each participant in each method. To do this, we co-registered each subject to the MNI-152 average brain using FLIRT (FSL) (Jenkinson & Smith, 2001; Jenkinson et al., 2002) and computed the average and standard deviation of the peak location across participants for each case. We then calculated an ‘error ellipsoid’ (one for each calibration method) that was centred on the average location and with each axis of the ellipsoid representing the standard deviation of peak locations in X, Y, and Z across participants. We reasoned that if all three methods localised with similar accuracy, the ellipsoids would overlap with similar volumes (suggesting that errors are dominated by intersubject variation rather than calibration method).

3.1 Validation experiments

Figure 4A shows the basic output of the calibration procedures. In both the left and right panels, the grey data-points and arrows show the sensor locations and sensitive orientations, according to CAD, respectively. In the left-hand plot, the coloured data-points and arrows show the HALO-derived locations and orientations following the ICP process. In the right-hand plot, the coloured data points and arrows show the MC-derived locations and orientations following the ICP process. In both cases, the colour of the arrow represents the derived gain multiplier for that channel (i.e., relative to the assumed value of 2.7 V/nT); blue shows gain scaling values <1 and red shows gain scaling values >1. Note that grey arrows are used for the CAD data as we always assume a gain of 2.7 V/nT. Data from a single representative experimental run are shown. For this specific run, the mean (across sensors) absolute Euclidean distance from the CAD-derived to the HALO-derived sensor location was 2.9 mm; the equivalent distance from the CAD to the MC-derived location was 3.4 mm. Across all experimental runs, the average absolute Euclidean distances were 3.0 ± 0.1 mm for the HALO and 3.7 ± 0.3 mm for the MC (mean ± standard deviation values shown). These values show good agreement between calibration and the CAD (the latter being the closest we have to a ground truth).

Fig. 4.

Quantifying the accuracy of calibration-derived sensor locations. (A) CAD- (grey) and calibration- (coloured) derived sensor locations, orientations, and gains. Left panel shows HALO, and right panel shows MC. In both cases, the data points show the sensor location, the arrows show three sensitive orientations, and the colour shows the gain multiplier. (In the case of the CAD-derived values, the gain multiplier is assumed to be 1). (B) Matrix showing the mean residual sensor-to-sensor distance between CAD- and HALO-derived locations. (C) Matrix showing the mean residual sensor-to-sensor distance between CAD- and MC-derived locations. (D) Matrix showing the mean residual sensor-to-sensor distance between MC- and HALO-derived locations.

Fig. 4.

Quantifying the accuracy of calibration-derived sensor locations. (A) CAD- (grey) and calibration- (coloured) derived sensor locations, orientations, and gains. Left panel shows HALO, and right panel shows MC. In both cases, the data points show the sensor location, the arrows show three sensitive orientations, and the colour shows the gain multiplier. (In the case of the CAD-derived values, the gain multiplier is assumed to be 1). (B) Matrix showing the mean residual sensor-to-sensor distance between CAD- and HALO-derived locations. (C) Matrix showing the mean residual sensor-to-sensor distance between CAD- and MC-derived locations. (D) Matrix showing the mean residual sensor-to-sensor distance between MC- and HALO-derived locations.

Close modal

A limitation with these absolute discrepancies is that they do not allow investigation of which sensor locations are best matched to the CAD-derived locations. This is because our ICP algorithm minimises the average distance between CAD- and calibration-derived locations, and this could mask individual sensors that offer only a poor agreement to CAD. For this reason, we also derived sensor-to-sensor distances across the helmet (which are independent of the ICP algorithm). Figure 4B, C, and D show matrices representing the residuals between the CAD-derived and the calibration-derived sensor-to-sensor distances. Figure 4B shows the comparison between the HALO and CAD. Figure 4C shows the comparison between the MC and CAD. Figure 4D shows the comparison between the HALO and the MC. Note first that the residual matrices in Figure 4B and C appear visually similar (the Pearson correlation coefficient between these matrices is 0.77). This means that sensors in which HALO-derived locations differ most from CAD also tend to be the sensors where the MC-derived locations differ most from CAD. In general, the residuals between the HALO and MC sensor-to-sensor distances were lower, suggesting that the HALO and MC are in better agreement with each other than they are with CAD. This suggests the dominant inaccuracies relate to the CAD itself (i.e., helmet manufacture and sensor placement).

Formalising the above results, we computed the average residual sensor-to-sensor distance across all sensors and all experimental repeats (i.e., the average value of all elements in the corresponding matrix Fig. 4B, C or D) and computed an estimate of how much individual inter-sensor distances changed from repeat-to-repeat by averaging the individual standard deviation values. The final values were 2.08 ± 0.61 mm for HALO versus CAD, 2.62 ± 1.7 mm for MC versus CAD, and 1.97 ± 1.5 mm for HALO versus MC. (All cases show mean ± standard deviation.) Most importantly, these values, and the matrices from which they are derived, show that both of our calibration techniques are able to localise sensors to within a few millimetres of their CAD-derived locations, and that these residual ‘errors’ most likely derive from the CAD.

Figure 5 shows how our calibration-derived estimates of the sensitive orientations of the OPM differ from the CAD models (these differences are measured as angles, so an angle of 10° means the orientation of the sensitive axis in the OPM differs by 10° from the orientation of its outer casing). Figure 5A-C plots the MC-derived angular discrepancy against the HALO-derived angular discrepancy. Panels A, B, and C show the x, y, and z axes of each OPM, respectively. In all three cases, the data points show the mean angular discrepancy, for a single OPM, calculated across experimental runs. The error bar shows the standard deviation across runs, meaning a small error bar denotes good consistency in the calibration-derived orientation across experiments. In all cases, the best fit line is shown in red and the line of equality in black. The fact that these lines fall very close to each other indicates good agreement between methods (i.e., the sensitive orientation derived using the HALO is similar to that derived using the MC).

Fig. 5.

Estimating the accuracy of calibrated sensor orientations. All panels show the MC-derived angular discrepancy (with respect to CAD) against the HALO-derived angular discrepancy (also with respect to CAD), that is, an angle of 10° would mean that the orientation of the sensitive axis in the OPM differs by 10° from the orientation of the outer casing of the sensor. Panels (A), (B), and (C) show the x, y, and z axes of each OPM, respectively. Red lines show a best fit to the data. Black lines show the line of equality. All data points show the mean value across experimental runs, and the error bars represent standard deviation. A close linear relationship suggests that the two calibration methods result in very similar sensitive orientations.

Fig. 5.

Estimating the accuracy of calibrated sensor orientations. All panels show the MC-derived angular discrepancy (with respect to CAD) against the HALO-derived angular discrepancy (also with respect to CAD), that is, an angle of 10° would mean that the orientation of the sensitive axis in the OPM differs by 10° from the orientation of the outer casing of the sensor. Panels (A), (B), and (C) show the x, y, and z axes of each OPM, respectively. Red lines show a best fit to the data. Black lines show the line of equality. All data points show the mean value across experimental runs, and the error bars represent standard deviation. A close linear relationship suggests that the two calibration methods result in very similar sensitive orientations.

Close modal

To quantify the deviation of the angles from the expected value for each sensor axis, we calculated the mean, standard deviation, minimum, and maximum angular discrepancies across all runs and channels. These were: x-axis: 17.1 ± 8.6° (max/min = 37.1/2.0°), y-axis: 5.4 ± 2.3° (max/min = 10.7/0.3°), and z-axis: 10.3 ± 5.3° (max/min = 23.1/0.8°) using the HALO and x-axis: 17.2 ± 8.5° (max/min = 36.0/2.4°), y-axis: 5.4 ± 2.4° (max/min = 11.3/0.2°), and z-axis: 10.5 ± 5.4° (max/min = 23.7/1.3°) using the MC. The x-axis has the largest error and widest range of values and y-channel the smallest error and narrowest range of values.

The average angles between the HALO and MC recovered orientations were: x-axis: 1.6 ± 1.3° (max/min = 7.6/0°), y-axis 1.0 ± 0.6° (max/min = 4.8/0°), and z-axis 1.0 ± 0.6° (max/min = 3.5/0°). The wide discrepancy across both sensors and channels with the CAD model is maintained across methods, highlighting the importance of calibration.

Figure 6 shows equivalent plots to Figure 5 but for the derived gain values. All values are centred around the expected 2.7 V/nT gain, and again the best-fit line closely follows the line of equality, showing that the two calibration techniques generate similar gain values, for each sensor. We again quantified the deviation of the gain from the expected value across each axis via computation of the mean, standard deviation, and minimum and maximum values across all runs and channels. These were: x-axis: 2.61 ± 0.24 V/nT (max/min = 3.15/2.01 V/nT), y-axis: 2.78 ± 0.19 V/nT (max/min = 3.75/2.39 V/nT), and z-axis: 2.67 ± 0.17 V/nT (max/min = 3.01/2.23 V/nT) for the HALO and x-axis: 2.63 ± 0.25 V/nT (max/min = 3.16/2.13 V/nT), y-axis: 2.80 ± 0.19 V/nT (max/min = 3.71/2.42 V/nT), and z-axis: 2.69 ± 0.16 V/nT (max/min = 3.07/2.29 V/nT) using the MC.

Fig. 6.

The accuracy of the calibrated sensor gains compared with the expected value (of 2.7 V/nT). Each panel plots the HALO-recovered gain against the MC-recovered gain. The values are shown as the mean across repeats, with the standard deviation shown by the error bars. Panels (A), (B), and (C) show angles for the sensor X, Y, and Z channels respectively. The dashed crosshairs show the expected gain value of 2.7 V/nT.

Fig. 6.

The accuracy of the calibrated sensor gains compared with the expected value (of 2.7 V/nT). Each panel plots the HALO-recovered gain against the MC-recovered gain. The values are shown as the mean across repeats, with the standard deviation shown by the error bars. Panels (A), (B), and (C) show angles for the sensor X, Y, and Z channels respectively. The dashed crosshairs show the expected gain value of 2.7 V/nT.

Close modal

The average percentage deviations of the recovered gain values between the HALO and MC were: x-axis: 1.6 ± 1.4% (max/min 8.2/0.0%), y-axis 1.2 ± 0.9% (max/min 4.9/0.0%), and z-axis 1.2 ± 1.0% (max/min 6.3/0.0%). Again, a discrepancy between the recovered values and the expected value and agreement across calibration techniques highlights the importance of calibration and the accuracy of both methods.

3.2 Human experiments

Figure 7 shows the beta band modulation induced by finger abduction. Figure 7A shows the group-averaged pseudo-T-statistical image, derived independently using the CAD- (top), HALO- (middle), and MC- (bottom) derived calibrations. The colour overlay, which shows the spatial signature of task-induced beta power loss during movement, is thresholded to 80% of its minimum value. In all cases, the largest beta modulation is observed in the sensorimotor areas as would be expected. However, the pseudo-T value itself (a proxy metric for SNR) is ‘largest’ (i.e., most negative) for the HALO-derived calibration (-0.41 ± 0.03 – mean ± standard deviation over subjects); the MC-derived calibration resulted in a marginally lower pseudo-T value (-0.37 ± 0.03), and the CAD-derived calibration shows the smallest (least negative) value (-0.29 ± 0.04). This suggests a better reconstruction of the source for both the HALO and MC calibration, compared to the CAD model, with an average increase of 1.42 ± 0.09 and 1.31 ± 0.09 respectively.

Fig. 7.

Beta band effects. (A) Average pseudo-T-statistical image (thresholded at 80% of their minimum value) and a TFS showing the evolution of neural oscillatory activity at the location of peak beta modulation; data averaged across all participants for the CAD- (Top), HALO- (Middle), and MC- (Bottom) derived calibrations. (B) The source reconstructed beta-band envelope for each method (CAD, HALO, and MC in black, red, and blue respectively) averaged across subjects with the standard deviation shown as a shaded area. This was again calculated at the location of largest beta modulation for each subject. (C) Beta-band SNR calculated for each subject for CAD-, HALO-, and MC-derived calibration in black, red, and blue respectively. (D) The location of largest beta change for each participant, using all three calibration methods. CAD peaks are shown as black circles, HALO as red dots, and MC as blue crosses. In addition, the mean position and standard deviation in each axis are shown by shaded ellipsoids.

Fig. 7.

Beta band effects. (A) Average pseudo-T-statistical image (thresholded at 80% of their minimum value) and a TFS showing the evolution of neural oscillatory activity at the location of peak beta modulation; data averaged across all participants for the CAD- (Top), HALO- (Middle), and MC- (Bottom) derived calibrations. (B) The source reconstructed beta-band envelope for each method (CAD, HALO, and MC in black, red, and blue respectively) averaged across subjects with the standard deviation shown as a shaded area. This was again calculated at the location of largest beta modulation for each subject. (C) Beta-band SNR calculated for each subject for CAD-, HALO-, and MC-derived calibration in black, red, and blue respectively. (D) The location of largest beta change for each participant, using all three calibration methods. CAD peaks are shown as black circles, HALO as red dots, and MC as blue crosses. In addition, the mean position and standard deviation in each axis are shown by shaded ellipsoids.

Close modal

Figure 7A also shows the group-averaged TFS for a source at the location of the peak pseudo-T-statistic, for the three calibrations. The three plots are similar, with the expected movement-related beta decrease followed by an increase above baseline (the beta ‘rebound’) immediately following movement cessation. The TFSs reconstructed using either the HALO or MC calibrations show larger relative change from baseline compared to the TFS produced using CAD calibration. This difference is quantified in Figure 7B/C. Figure 7B shows the beta-band envelope averaged over participants, with the standard deviation indicated by the shaded area. Figure 7C shows the reconstructed source-space SNR for each subject. While both calibration methods show significant (p = 0.0156 for both methods, Wilcoxon sign-rank test) increases in SNR compared to the CAD-derived calibration, the HALO-derived calibration provides a greater increase. On average, the SNR increased by a factor of 2.85 ± 0.95 for the HALO and 1.77 ± 0.23 for the MC-derived calibration.

Figure 7D shows the location of maximum beta change for each subject, after transformation to the MNI-152 average brain. The average reconstructed location and its standard deviation in each axis are represented by the shaded ellipsoids. For all participants, the activity is localised to the precentral gyrus and the error ellipsoids overlap. However, for Subject 2 in the CAD-derived data, the peak location was found in the right precentral gyrus (i.e., an ipsilateral response)—the HALO- and MC-derived calibrations did not show this. The ellipsoids have volumes of 6417 mm3, 829 mm3, and 889 mm3 for the CAD, HALO, and MC calibrations, respectively, suggesting that the CAD-derived peak locations have a greater spatial spread while the HALO and MC are similar. These peak localisations are summarised in Table 1.

Table 1.

Localisation summary.

MethodMean position (mm)Standard deviation (mm)Volume (mm3)
BetaGammaEvokedBetaGammaEvokedBetaGammaEvoked
CAD [-26, -13, 64] [-2, -92, 5] [9, -91, 25] [31, 8, 5] [7, 4, 5] [19, 10, 15] 6417 822 13363 
HALO [-40, -16, 60] [-4, -93, 6] [0, -91, 21] [6, 7, 4] [5, 5, 5] [10, 5, 14] 829 703 3808 
MC [-39, -17, 63] [-4, -92, 5] [0, -91, 23] [7, 7, 3] [4, 5, 7] [10, 7, 16] 889 776 4900 
MethodMean position (mm)Standard deviation (mm)Volume (mm3)
BetaGammaEvokedBetaGammaEvokedBetaGammaEvoked
CAD [-26, -13, 64] [-2, -92, 5] [9, -91, 25] [31, 8, 5] [7, 4, 5] [19, 10, 15] 6417 822 13363 
HALO [-40, -16, 60] [-4, -93, 6] [0, -91, 21] [6, 7, 4] [5, 5, 5] [10, 5, 14] 829 703 3808 
MC [-39, -17, 63] [-4, -92, 5] [0, -91, 23] [7, 7, 3] [4, 5, 7] [10, 7, 16] 889 776 4900 

The mean and standard deviation for peak localisation across participants for each of the three tasks using each of the three calibration methods. The final column shows the volumes of the error ellipsoids plotted in Figures 7D, 8D, and 9D.

Figure 8 shows gamma band (40–60 Hz) oscillations induced by visual stimulation using the circular grating. (The Figure layout is equivalent to Fig. 7.) The pseudo-T statistics at the locations of maximum gamma change were 0.48 ± 0.30 for the CAD calibration, 0.68 ± 0.47 for the HALO, and 0.62 ± 0.42 for the MC. Again, this suggests a better source reconstruction with the HALO and MC, with an average increase of 1.32 ± 0.17 and 1.22 ± 0.15 respectively. The TFS again show larger relative change when calibration is carried out using either the HALO or the MC compared to CAD. On average, the SNR increased by a factor of 1.32 ± 0.35 for the HALO and 1.25 ± 0.41 for the MC-derived calibrations. However, while the improvement in SNR was significant for the HALO at the group level (p = 0.047 - Wilcoxon sign-rank test), it was not significant for the MC (p = 0.38 – Wilcoxon sign-rank test). Data in Figure 8C show that SNR improvements were less pronounced for gamma activity than they were for beta-band activity. Indeed, for some subjects SNR was marginally reduced; this will be addressed further in the Discussion. For all participants, using all three calibrations, the maximum gamma modulation activity was localised to the primary visual cortex, as shown in Figure 8D. The ellipsoids have similar volumes of 822 mm3, 703 mm3, and 776 mm3 for the CAD, HALO, and MC calibrations respectively.

Fig. 8.

Gamma-band effects. (A) Average pseudo-T-statistical images (thresholded at 80% of their maximum value) and a TFS showing the evolution of neural oscillatory activity at the location of peak gamma modulation; data averaged across participants for the CAD- (Top), HALO- (Middle), and MC- (Bottom) derived calibrations. (B) The source reconstructed gamma-band envelope for each calibration method (CAD, HALO, and MC in black, red, and blue respectively) averaged across subjects with the standard deviation shown as a shaded area. This was again calculated at the location of largest gamma modulation for each subject. (C) Gamma-band SNR calculated for each subject for CAD-, HALO-, and MC-derived calibration in black, red, and blue respectively. (D) The location of largest gamma change for each participant, using all three calibration methods. CAD peaks are shown as black circles, HALO as red dots, and MC as blue crosses. In addition, the mean position and standard deviation in each axis are shown by shaded ellipsoids.

Fig. 8.

Gamma-band effects. (A) Average pseudo-T-statistical images (thresholded at 80% of their maximum value) and a TFS showing the evolution of neural oscillatory activity at the location of peak gamma modulation; data averaged across participants for the CAD- (Top), HALO- (Middle), and MC- (Bottom) derived calibrations. (B) The source reconstructed gamma-band envelope for each calibration method (CAD, HALO, and MC in black, red, and blue respectively) averaged across subjects with the standard deviation shown as a shaded area. This was again calculated at the location of largest gamma modulation for each subject. (C) Gamma-band SNR calculated for each subject for CAD-, HALO-, and MC-derived calibration in black, red, and blue respectively. (D) The location of largest gamma change for each participant, using all three calibration methods. CAD peaks are shown as black circles, HALO as red dots, and MC as blue crosses. In addition, the mean position and standard deviation in each axis are shown by shaded ellipsoids.

Close modal

Figure 9 shows the results of recording evoked responses to faces. Figure 9A shows the subject averaged (normalised) pseudo-Z-statistical map at a time t = 108 ms, thresholded to 80% of its maximum value. All three cases show an average localisation in the occipital lobe. The activity derived using the CAD data localises entirely to the calcarine; for the MC-derived calibration, activity localises to the cuneus; and the HALO-derived activity bridges the two regions. Figure 9B shows the reconstructed timecourse for a point at the centre of mass of the calcarine as well as the left and right fusiform (as determined by the AAL atlas for each participant). For the three calibrations, occipital timecourses show similar temporal morphology, with the HALO- and MC-derived positions showing larger peaks at 80 ms and 108 ms compared to the CAD data. Figure 9C shows the occipital SNR for all participants, which increased significantly for both the HALO and MC compared to CAD (p = 0.0156 for both methods – Wilcoxon sign-rank test). On average, the SNR increased by a factor of 1.31 ± 0.19 for the HALO and 1.21 ± 0.10 for the MC-derived calibrations. Figure 9D shows that, for all participants, the peak at 108 ms was localised to the occipital lobe with the mean location (i.e., the average across participants) remaining similar in each case. However, there was a greater standard deviation of peak location in the CAD-derived case. The ellipsoids have volumes of 13,363 mm3, 3808 mm3, and 4900 mm3 for CAD, HALO, and MC calibrations respectively. This again implies a larger scatter of locations for the CAD-derived peak.

Fig. 9.

Evoked responses. (A) Average pseudo-Z-statistical images (normalised by their maximum absolute value) taken at t = 108 ms and thresholded at 80% of the maximum. (B) the source-reconstructed evoked response in the occipital, and left and right fusiform, shown for each calibration method (CAD, HALO, and MC in black, red, and blue respectively). All data are averaged across subjects with the standard deviation shown as a shaded area. The data were reconstructed for a point at the centre of mass of the calcarine, identified by the AAL atlas. (C) SNR of the occipital response for each subject, evaluated for CAD-, HALO-, and MC-derived calibration (in black, red, and blue respectively). (D) The location of the largest change in evoked response at 108 ms; peaks for each participant are shown for the three calibration methods. CAD shown as black circles, HALO as red dots, and MC as blue crosses. In addition, the mean position and standard deviation in each axis are shown by shaded ellipsoids.

Fig. 9.

Evoked responses. (A) Average pseudo-Z-statistical images (normalised by their maximum absolute value) taken at t = 108 ms and thresholded at 80% of the maximum. (B) the source-reconstructed evoked response in the occipital, and left and right fusiform, shown for each calibration method (CAD, HALO, and MC in black, red, and blue respectively). All data are averaged across subjects with the standard deviation shown as a shaded area. The data were reconstructed for a point at the centre of mass of the calcarine, identified by the AAL atlas. (C) SNR of the occipital response for each subject, evaluated for CAD-, HALO-, and MC-derived calibration (in black, red, and blue respectively). (D) The location of the largest change in evoked response at 108 ms; peaks for each participant are shown for the three calibration methods. CAD shown as black circles, HALO as red dots, and MC as blue crosses. In addition, the mean position and standard deviation in each axis are shown by shaded ellipsoids.

Close modal

MEG provides a sensitive means to characterise human brain function, and has significant clinical utility, for example to identify abnormal brain activity in epilepsy and determine its spatial origin (Rampp et al., 2019). However, the accuracy of MEG-derived metrics of both normal and abnormal brain function depends critically on knowing the relative locations of sensors, their sensitive axes, and their gain. If these key calibration parameters are incorrect, systematic errors in source localisation and reduced sensitivity necessarily follow. In this paper, we demonstrate two independent techniques that use electromagnetic coils to produce known magnetic fields which can be used for OPM-MEG array calibration. We have shown that both techniques enable accurate determination of sensor location, orientation, and gain; the two methods agree with each other (on average for position, orientation, and gain by just 2.0 mm; 1.2° and 1.3% respectively), and both offer improvements in signal-to-noise ratio when characterising well-known neurophysiological processes in human MEG data.

The accuracy of calibration was assessed via measurement of the agreement between CAD-derived and calibration-derived sensor locations. This measurement was used because the CAD locations were the closest available metric to a ‘ground truth’. We assessed sensor-to-sensor distance for all sensor pairs, and the difference between CAD- and calibration-derived values was 2.08 ± 0.07 mm and 2.6 ± 0.2 mm for HALO and MC, respectively. This notionally offers a quantifiable ‘error’ in the sensor localisation (and, consequently, calibration) process. However, it is likely that this scale of variation originates not only from errors due to the calibration but also from inaccuracies in the CAD model of the helmet. Indeed, the high similarity between Figure 4B and C shows that the sensor-to-sensor pairings in which the HALO calibration differed most from CAD were the same pairings for which the MC calibration differed from CAD. Further, the discrepancy between our calibration techniques (i.e., the HALO-to-MC difference of 1.97 ± 0.16 mm) was less than the discrepancy between either technique and CAD. Errors in the CAD model likely come from multiple sources including inaccuracies in 3D printing, warping of the OPM helmet over time, and inaccuracies in sensor placement in the helmet. It is likely that these issues could explain a 2-mm discrepancy. Our localisation errors on the 2-mm scale agreed well with those reported by Iivanainen et al. (2022) (3.3 mm) using a similar methodology. Both the existing CAD and the calibrated sensor locations agree with the minimum accuracy suggested by Zetter et al. (2018), of <4 mm, required to be comparable to existing cryogenic MEG systems.

Unlike the sensor locations, we have no ground-truth metric of sensor orientation, save that the orientation depends on the direction of modulation fields made by on-board coils, which we assume to be parallel with the axes of the outer casing of the sensor. We found that while the sensitive axes broadly pointed along the assumed direction, there were differences of ~17° for X-axes, ~5° for Y-axes and ~10° for Z-axes. In agreement with our sensor location results, these values were robust both across independent experiments using the same calibration procedure, and across our two (independent) calibration techniques (by 1.2° on average across all channels). It is noteworthy that Zetter et al. found that channel orientation should be known to a precision of <10° for MEG reconstruction; this would be met by most (but not all) channels using the CAD determined orientations but would be met (and bettered) in full following calibration.

For sensor gain we compared our results to the assumed values of 2.7 V/nT. There was again a significant spread in values, with a standard deviation in each axis between 6% and 10% of its mean. These differences from the assumption are large, and indeed (Nugent et al., 2022) indicated from simulation work that gain errors >5% can completely remove the capability of a dense scalp-mounted patch array of OPMs to resolve nearby sources. It is, thus, interesting that despite this significant spread of gain values our human data were not dramatically impacted when the assumed calibration was used. We suggest this is due to a larger spacing of sensors in our rigid helmets which hold sensors slightly away from the scalp surface compared to the simulation work. The system used here has been shown to offer similar performance to cryogenic devices (e.g., (Rhodes et al., 2023; Safar et al., 2024)). Tierney et al. (2020) investigated the relationship between gain errors (up to 10%) and spatial discrimination of OPM-MEG systems for shallow and deep sources, finding noticeable differences only at high SNR (100 nAm sources). As the channel density of systems is increased to answer questions about the fundamental spatial resolution that is possible with OPM-MEG, calibration will clearly become a priority.

We also note that dynamic gain changes due to field variations (from environmental or significant movement artefacts, which are not present in our data here) would not be accounted for by our calibration method. If significant field changes were expected, one could alleviate concerns with dynamic active shielding, performing calibration at regular intervals during an experiment, and by operating sensors in closed-loop mode.

In the human MEG experiments, all methods allowed sources to be localised to the expected cortical regions, demonstrating that even without calibration, MEG analysis was robust. However, several measurable improvements were afforded by performing the calibration. Firstly, the maximum pseudo-T-statistics increased, implying a better source reconstruction. Secondly, the spatial spread of peak locations (measured by the ellipsoid volumes) tended to be smaller following calibration (at least for beta and evoked responses). Perhaps most importantly, the SNR of all three measured MEG responses was increased by calibration. The improvements in source space sensitivity and spatial precision result from the beamformer algorithm’s exploitation of the improved match of the data to the MEG forward model after calibration. This may also explain why the improvement was larger for the beta band than it was for the gamma band. In the beta band, movement of a single digit likely evokes activity in a relatively small region of the motor cortex, and consequently the magnetic field generated looks dipolar in nature. In the gamma band, the visual stimulus was large and centrally presented, which likely evokes activity in both left and right hemispheres. This larger volume of activation is expected to generate a field that is less dipolar. If this departure from a dipole model is a larger source of error than that introduced by inaccurate calibration, then the improvement afforded by calibration would be relatively subtle. It is therefore possible (even likely) that calibration will not improve all MEG localisations.

Here, we showed that calibration improves beamformer localisation of induced and evoked effects. However, improvements should be maintained across other processing techniques. For example, the ability of the signal space separation method (SSS) (Taulu et al., 2005; Taulu & Kajola, 2005) and similar magnetic multipole modelling approaches (including homogeneous field correction (HFC) (Tierney et al., 2021) and adaptive multipole models (AMM) (Tierney et al., 2024)) to reject interference should be improved by accurate calibration due to an increased capability to discriminate spatially the measurable field patterns across the helmet (Nurminen et al., 2008). It has been shown that multi-axis sensors, especially triaxial sensors as used in the current generation of OPM-MEG devices, offer distinct advantages when applying such models (Nurminen et al., 2013; Tierney et al., 2022, 2024) (compared to sensors which only measure a single component of field) as the angle between the subspace of harmonics which span the neuronal signals and interference is dramatically increased (from ~10° in a conventional MEG system (Taulu et al., 2005) to ~60° in the OPM-MEG system presented here (Holmes et al., 2023a). However, temporal extensions to the purely spatial models (e.g., tSSS (Taulu & Simola, 2006)) have so far been required to fully separate brain signal from background fields, implying calibration errors hinder performance (Holmes et al., 2024; Tierney et al., 2024). Future work should look to verify this improvement.

Finally, in terms of practicality each calibration process (HALO and MC) took a similar time (c.1 min) to complete. Calibration could, therefore, easily be performed at the beginning of a scan with relatively little disruption to the participant or experimenters. However, the repeatability of results implies that calibration need not necessarily be performed prior to every experiment. Here, we showed that calibrations were stable across multiple repeats in a single day, implying one calibration per day would be sufficient. Future work may offer further insight into stability over longer periods (e.g., could calibration be performed only once per week?). However, this would only be practical if OPMs stayed in the same relative locations in a rigid helmet. In the case of a flexible (EEG-like) or adaptable OPM mounting, having a calibration procedure that can be applied every time the sensor array morphology is changed is critical. Related to this, the HALO has the advantage that it is attached to the helmet, and therefore calibration can be carried out even in the presence of large subject head movement. In contrast, the MC calibration accuracy would decrease with subject motion during the calibration. Whether such movement could be accounted for remains a topic for further study. According to the sensor localisation results, the HALO was marginally more accurate than the MC and this was generally reflected in the human MEG experiments. It is possible that the MC performed worse as the true variation of the field produced by each of the coils was not captured by the 40 x 40 x 40 cm3 10-cm resolution grid used to record the data for the MC forward model. (A finer spacing and wider volume could improve this). Furthermore, the MC are at least 1.4 m away from the sensors, meaning their spatial variation may be too slow by the time they reach the sensors to discriminate sensor locations. This said, it is noteworthy that, for the HALO, the accuracy of sensor locations drops with distance from the HALO (see Appendix), perhaps as the field amplitude from the spiral coil becomes more dipolar with distance, deviating from the nearby moment (which was calibrated with a fluxgate sensor). It is also possible that some degrees of freedom in the magnetic field could be poorly sampled due to the shared orientation of each coil in the HALO. These issues could be addressed in several ways. For example, by mounting coils into the helmet itself (rather than on a plane riding above the helmet but maintaining the advantages robust printing to get structure). For example, Adachi et al. (2019) embedded small coils into planar magnetoresistive sensor arrays developed for magnetocardiography (MCG). Coil current distributions could also be adapted taking the form of bi-planar field cancellation systems (e.g., Holmes et al. (2018); Zetter et al. (2020)) to produce orthogonal fields from coils mounted on a planar PCB. Coils for calibration could also be worn by the participant, as demonstrated by Pfeiffer et al. (2018, 2020). However, we note that with such systems the proximity of the coils to the OPMs will generate focal field patterns across the sensor array. This will necessitate an improved starting guess for the optimisation algorithm than was needed in our case to avoid finding a solution in a local minimum. For our methods, the field patterns from each coil in the HALO and the MC vary similarly across the array, allowing simple navigation of the search space by the minimisation algorithm to reach the global minimum.

We have presented two methods to achieve complete OPM-MEG system calibration (measuring sensor location, orientation, and gain). Both methods were based on accurately generating and modelling well-characterised magnetic fields over the sensor array. Our results show that these techniques offer an accurate means to calibrate an OPM array, and that the two methods agree strongly with each other (on average for position, orientation and gain by just 2.0 mm; 1.2° and 1.3% respectively). Further, when applied to human MEG experiments, both methods offer significantly improved SNR compared to an assumed calibration. As OPM-MEG matures, arrays become denser, channel count increases, and the technology becomes more widely used for clinical applications, the importance of calibration to obtain reliable data will increase. Technologies for calibration will, therefore, likely become a core part of an OPM-MEG system.

This study was approved by the University of Nottingham Faculty of Medicine and Health Sciences Research Ethics Committee (approval number H16122016). All participants provided written informed consent.

All data were acquired by the authors, and code was custom developed in-house using MATLAB. These will be shared following a formal data-sharing agreement with the authors due to subject anonymity and intellectual property.

Conceptualisation: R.M.H., R.B., M.J.B., V.S., and N.H. Methodology: R.M.H., M.J.B., and V.S., N.H. Software: R.M.H., G.R.R., A.T., H.S., C.D., J.O., D.B., L.R., and N.H. Validation: R.M.H., G.R.R., A.T., H.S., L.R., J.G., and Z.T. Formal Analysis: R.M.H., G.R.R., A.T., H.S., C.D., J.O., D.B., L.R., J.G., Z.T., and N.H. Investigation: R.M.H., G.R.R., A.T., H.S., and N.H. Resources: R.M.H., G.R.R., A.T., H.S., C.D., J.O., D.B., E.B., and N.H. Data Curation: R.M.H., G.R.R., A.T., H.S., C.D., J.O., D.B., and N.H. Writing—Original Draft: R.M.H., M.J.B., V.S., and N.H. Writing—Reviewing and Editing: All authors. Visualisation: R.M.H., M.J.B., and N.H. Supervision: E.B., R.B., M.J.B., V.S., and N.H. Project Administration: R.M.H., N.H. Funding Acquisition: E.B., R.B., M.J.B., and V.S.

This work was supported by an Innovate UK biomedical catalyst grant (10037425) awarded to Cerca Magnetics Limited. We also acknowledge support from the UK Quantum Technology Hub in Sensing and Timing, funded by EPSRC (EP/T001046/1), a Medical Research Council (MRC) Mid-Range Equipment grant (MC_PC_MR/X012263/1) and an Engineering and Physical Sciences Research Council (EPSRC) Healthcare Impact Partnership Grant (EP/V047264/1). Sensor and HALO development was made possible by funding from the National Institutes of Health (R44MH110288 & 1R43HL165987-01).

V.S. is the founding director of QuSpin, a commercial entity selling OPM magnetometers. J.O., D.B., and C.D. are employees of QuSpin. E.B. and M.J.B. are directors of Cerca Magnetics Limited, a spin-out company whose aim is to commercialise aspects of OPM-MEG technology. E.B., M.J.B., R.B., N.H., and R.H. hold founding equity in Cerca Magnetics Limited. H.S. and Z.T. are employees of Cerca Magnetics Limited. L.R. is a Scientific Advisor to Cerca Magnetics Limited. G.R.R.’s PhD studentship was in part funded by Cerca Magnetics Limited.

The authors would like to thank Peter D.D. Schwindt for extremely useful discussions regarding this work.

Adachi
,
Y.
,
Oyama
,
D.
,
Higuchi
,
M.
, &
Uehara
,
G.
(
2023
).
A spherical coil array for the calibration of whole-head magnetoencephalograph systems
.
IEEE Trans Instrum Meas
,
72
,
1
10
. https://doi.org/10.1109/TIM.2023.3265750
Adachi
,
Y.
,
Oyama
,
D.
,
Terazono
,
Y.
,
Hayashi
,
T.
,
Shibuya
,
T.
, &
Kawabata
,
S.
(
2019
).
Calibration of room temperature magnetic sensor array for biomagnetic measurement
.
IEEE Trans Magn
,
55
,
1
6
. https://doi.org/10.1109/TMAG.2019.2895355
Allred
,
J. C.
,
Lyman
,
R. N.
,
Kornack
,
T. W.
, &
Romalis
,
M. V.
(
2002
).
High-sensitivity atomic magnetometer unaffected by spin-exchange relaxation
.
Phys Rev Lett
,
89
,
130801
. https://doi.org/10.1103/PhysRevLett.89.130801
Altarev
,
I.
,
Fierlinger
,
P.
,
Lins
,
T.
,
Marino
,
M. G.
,
Nießen
,
B.
,
Petzoldt
,
G.
,
Reisner
,
M.
,
Stuiber
,
S.
,
Sturm
,
M.
,
Taggart
Singh
, J.,
Taubenheim
,
B.
,
Rohrer
,
H. K.
, &
Schläpfer
,
U.
(
2015
).
Minimizing magnetic fields for precision experiments
.
J Appl Phys
,
117
,
233903
. https://doi.org/10.1063/1.4922671
Baillet
,
S.
(
2017
).
Magnetoencephalography for brain electrophysiology and imaging
.
Nat Neurosci
,
20
,
327
339
. https://doi.org/10.1038/nn.4504
Bentin
,
S.
,
Allison
,
T.
,
Puce
,
A.
,
Perez
,
E.
, &
McCarthy
,
G.
(
1996
).
Electrophysiological studies of face perception in humans
.
J Cogn Neurosci
,
8
,
551
565
. https://doi.org/10.1162/jocn.1996.8.6.551
Borna
,
A.
,
Iivanainen
,
J.
,
Carter
,
T. R.
,
McKay
,
J.
,
Taulu
,
S.
,
Stephen
,
J.
, &
Schwindt
,
P. D. D.
(
2022
).
Cross-Axis projection error in optically pumped magnetometers and its implication for magnetoencephalography systems
.
Neuroimage
,
247
,
118818
. https://doi.org/10.1016/j.neuroimage.2021.118818
Boto
,
E.
,
Holmes
,
N.
,
Leggett
,
J.
,
Roberts
,
G.
,
Shah
,
V.
,
Meyer
,
S. S.
,
Muñoz
,
L. D.
,
Mullinger
,
K. J.
,
Tierney
,
T. M.
,
Bestmann
,
S.
,
Barnes
,
G. R.
,
Bowtell
,
R.
, &
Brookes
,
M. J.
(
2018
).
Moving magnetoencephalography towards real-world applications with a wearable system
.
Nature
,
555
,
657
661
. https://doi.org/10.1038/nature26147
Boto
,
E.
,
Shah
,
V.
,
Hill
,
R. M.
,
Rhodes
,
N.
,
Osborne
,
J.
,
Doyle
,
C.
,
Holmes
,
N.
,
Rea
,
M.
,
Leggett
,
J.
,
Bowtell
,
R.
, &
Brookes
,
M. J.
(
2022
).
Triaxial detection of the neuromagnetic field using optically-pumped magnetometry: Feasibility and application in children
.
Neuroimage
,
252
,
119027
. https://doi.org/10.1016/j.neuroimage.2022.119027
Brookes
,
M. J.
,
Leggett
,
J.
,
Rea
,
M.
,
Hill
,
R. M.
,
Holmes
,
N.
,
Boto
,
E.
, &
Bowtell
,
R.
(
2022
).
Magnetoencephalography with optically pumped magnetometers (OPM-MEG): The next generation of functional neuroimaging
.
Trends Neurosci
,
45
,
621
634
. https://doi.org/10.1016/j.tins.2022.05.008
Chella
,
F.
,
Zappasodi
,
F.
,
Marzetti
,
L.
,
Della Penna
,
S.
, &
Pizzella
,
V.
(
2012
).
Calibration of a multichannel MEG system based on the signal space separation method
.
Phys Med Biol
,
57
,
4855
4870
. https://doi.org/10.1088/0031-9155/57/15/4855
Cohen-Tannoudji
,
C.
,
Dupont-Roc
,
J.
,
Haroche
,
S.
, &
Laloë
,
F.
(
1970
).
Diverses résonances de croisement de niveaux sur des atomes pompés optiquement en champ nul. I
.
Théorie. Revue de Physique Appliquée
,
5
,
95
101
. https://doi.org/10.1051/rphysap:019700050109500
Corvilain
,
P.
,
Wens
,
V.
,
Bourguignon
,
M.
,
Capparini
,
C.
,
Fourdin
,
L.
,
Ferez
,
M.
,
Ferez
,
M.
,
Feys
,
O.
,
De Tiège
,
X.
, &
Bertels
,
J.
(
2024
).
Extending the applicability of optically pumped magnetoencephalography toward early human life
.
bioRxiv
. https://doi.org/10.1101/2023.10.28.564455
Feys
,
O.
,
Corvilain
,
P.
,
Aeby
,
A.
,
Sculier
,
C.
,
Holmes
,
N.
,
Brookes
,
M.
,
Goldman
,
S.
,
Wens
,
V.
, &
De Tiège
,
X
. (
2022
).
On-scalp optically pumped magnetometers versus cryogenic magnetoencephalography for diagnostic evaluation of epilepsy in school-aged children
.
Radiology
,
304
,
429
434
. https://doi.org/10.1148/radiol.212453
Halgren
,
E.
,
Raij
,
T.
,
Marinkovic
,
K.
,
Jousmäki
,
V.
, &
Hari
,
R.
(
2000
).
Cognitive response profile of the human fusiform face area as determined by MEG
.
Cereb Cortex
,
10
,
69
81
. https://doi.org/10.1093/cercor/10.1.69
Hämäläinen
,
M.
,
Hari
,
R.
,
Ilmoniemi
,
R. J.
,
Knuutila
,
J.
, &
Lounasmaa
,
O. V.
(
1993
).
Magnetoencephalography—Theory, instrumentation, and applications to noninvasive studies of the working human brain
.
Rev Mod Phys
,
65
,
413
497
. https://doi.org/10.1103/RevModPhys.65.413
Hammond
,
P.
(
1960
).
Electric and magnetic images
.
Proceedings of the IEE Part C: Monographs
,
107
,
306
. https://doi.org/10.1049/pi-c.1960.0047
Happer
,
W.
(
1972
).
Optical pumping
.
Rev Mod Phys
,
44
,
169
. https://doi.org/10.1103/RevModPhys.44.169
Hill
,
R. M.
,
Boto
,
E.
,
Holmes
,
N.
,
Hartley
,
C.
,
Seedat
,
Z. A.
,
Leggett
,
J.
,
Roberts
,
G.
,
Shah
,
V.
,
Tierney
,
T. M.
,
Woolrich
,
M. W.
,
Stagg
,
C. J.
,
Barnes
,
G. R.
,
Bowtell
,
R.
,
Slater
,
R.
, &
Brookes
,
M. J.
(
2019
).
A tool for functional brain imaging with lifespan compliance
.
Nat Commun
,
10
,
4785
. https://doi.org/10.1038/s41467-019-12486-x
Hill
,
R. M.
,
Boto
,
E.
,
Rea
,
M.
,
Holmes
,
N.
,
Leggett
,
J.
,
Coles
,
L. A.
,
Papastavrou
,
M.
,
Everton
,
S. K.
,
Hunt
,
B. A .E.
,
Sims
,
D.
,
Osborne
,
J.
,
Shah
,
V.
,
Bowtell
,
R.
, &
Brookes
,
M. J.
(
2020
).
Multi-channel whole-head OPM-MEG: Helmet design and a comparison with a conventional system
.
Neuroimage
,
219
,
116995
. https://doi.org/10.1016/j.neuroimage.2020.116995
Hill
,
R. M.
,
Devasagayam
,
J.
,
Holmes
,
N.
,
Boto
,
E.
,
Shah
,
V.
,
Osborne
,
J.
,
Safar
,
K.
,
Worcester
,
F.
,
Mariani
,
C.
,
Dawson
,
E.
,
Woolger
,
D.
,
Bowtell
,
R.
,
Taylor
,
M. J.
, &
Brookes
,
M. J.
(
2022
).
Using OPM-MEG in contrasting magnetic environments
.
Neuroimage
,
253
,
119084
. https://doi.org/10.1016/j.neuroimage.2022.119084
Hill
,
R. M.
,
Schofield
,
H.
,
Boto
,
E.
,
Rier
,
L.
,
Osborne
,
J.
,
Doyle
,
C.
,
Worcester
,
F.
,
Hayward
,
T.
,
Holmes
,
N.
,
Bowtell
,
R.
,
Shah
,
V.
, &
Brookes
,
M. J.
(
2024
).
Optimising the sensitivity of optically-pumped magnetometer magnetoencephalography to gamma band electrophysiological activity
.
Imaging Neuroscience
,
2
,
1
19
. https://doi.org/10.1162/imag_a_00112
Holmes
,
N.
,
Bowtell
,
R.
,
Brookes
,
M. J.
, &
Taulu
,
S.
(
2023a
).
An Iterative Implementation of the signal space separation method for magnetoencephalography systems with low channel counts
.
Sensors
,
23
,
6537
. https://doi.org/10.3390/s23146537
Holmes
,
N.
,
Leggett
,
J.
,
Boto
,
E.
,
Roberts
,
G.
,
Hill
,
R. M.
,
Tierney
,
T. M.
,
Shah
,
V.
,
Barnes
,
G. R.
,
Brookes
,
M. J.
, &
Bowtell
,
R.
(
2018
).
A bi-planar coil system for nulling background magnetic fields in scalp mounted magnetoencephalography
.
Neuroimage
,
181
,
760
774
. https://doi.org/10.1016/j.neuroimage.2018.07.028
Holmes
,
N.
,
Leggett
,
J.
,
Hill
,
R. M.
,
Rier
,
L.
,
Boto
,
E.
,
Schofield
,
H.
,
Hayward
,
T.
,
Dawson
,
E.
,
Woolger
,
D.
,
Shah
,
V.
,
Taulu
,
S.
,
Brookes
,
M. J.
, &
Bowtell
,
R.
(
2024
).
Wearable magnetoencephalography in a lightly shielded environment
.
IEEE Trans Biomed Eng
,
72
,
609
618
. https://doi.org/10.1109/TBME.2024.3465654
Holmes
,
N.
,
Rea
,
M.
,
Chalmers
,
J.
,
Leggett
,
J.
,
Edwards
,
L. J.
,
Nell
,
P.
,
Pink
,
S.
,
Patel
,
P.
,
Wood
,
J.
,
Murby
,
N.
,
Woolger
,
D.
,
Dawson
,
E.
,
Mariani
,
C.
,
Tierney
,
T. M.
,
Mellor
,
S.
,
O’Neill
,
G. C.
,
Boto
,
E.
,
Hill
,
R. M.
,
Shah
,
V.
, …
Bowtell
,
R.
(
2022
).
A lightweight magnetically shielded room with active shielding
.
Sci Rep
,
12
,
13561
. https://doi.org/10.1038/s41598-022-17346-1
Holmes
,
N.
,
Rea
,
M.
,
Hill
,
R. M.
,
Leggett
,
J.
,
Edwards
,
L. J.
,
Hobson
,
P. J.
,
Boto
,
E.
,
Tierney
,
T. M.
,
Rier
,
L.
,
Rivero
,
G. R.
,
Shah
,
V.
,
Osborne
,
J.
,
Fromhold
,
T. M.
,
Glover
,
P.
,
Brookes
,
M. J.
, &
Bowtell
,
R.
(
2023b
).
Enabling ambulatory movement in wearable magnetoencephalography with matrix coil active magnetic shielding
.
Neuroimage
,
274
,
120157
. https://doi.org/10.1016/j.neuroimage.2023.120157
Hoogenboom
,
N.
,
Schoffelen
,
J.-M.
,
Oostenveld
,
R.
,
Parkes
,
L. M.
, &
Fries
,
P.
(
2006
).
Localizing human visual gamma-band activity in frequency, time and space
.
Neuroimage
,
29
,
764
773
. https://doi.org/10.1016/j.neuroimage.2005.08.043
Iivanainen
,
J.
,
Borna
,
A.
,
Zetter
,
R.
,
Carter
,
T. R.
,
Stephen
,
J. M.
,
McKay
,
J.
,
Parkkonen
,
L.
,
Taulu
,
S.
, &
Schwindt
,
P. D. D.
(
2022
).
Calibration and localization of optically pumped magnetometers using electromagnetic coils
.
Sensors
,
22
,
3059
. https://doi.org/10.3390/s22083059
Iivanainen
,
J.
,
Stenroos
,
M.
, &
Parkkonen
,
L.
(
2017
).
Measuring MEG closer to the brain: Performance of on-scalp sensor arrays
.
Neuroimage
,
147
,
542
553
. https://doi.org/10.1016/j.neuroimage.2016.12.048
Jenkinson
,
M.
,
Bannister
,
P.
,
Brady
M.
, &
Smith
,
S.
(
2002
).
Improved optimization for the robust and accurate linear registration and motion correction of brain images
.
Neuroimage
,
17
,
825
841
. https://doi.org/10.1016/S1053-8119(02)91132-8
Jenkinson
,
M.
, &
Smith
,
S.
(
2001
).
A global optimisation method for robust affine registration of brain images
.
Med Image Anal
,
5
,
143
156
. https://doi.org/10.1016/S1361-8415(01)00036-6
Li
,
H.
,
Zhang
,
S.-L.
,
Zhang
,
C.-X.
,
Kong
,
X.-Y.
, &
Xie
,
X.-M.
(
2016
).
An efficient calibration method for SQUID measurement system using three orthogonal Helmholtz coils
.
Chinese Phys B
,
25
,
068501
. https://doi.org/10.1088/1674-1056/25/6/068501
Mellor
,
S.
,
Tierney
,
T. M.
,
Seymour
,
R. A.
,
Timms
,
R. C.
,
O’Neill
,
G. C.
,
Alexander
,
N.
,
Spedden
,
M. E.
,
Payne
,
H.
, &
Barnes
,
G. R.
(
2023
).
Real-time, model-based magnetic field correction for moving, wearable MEG
.
Neuroimage
,
278
,
120252
. https://doi.org/10.1016/j.neuroimage.2023.120252
Mellor
,
S. J.
,
Tierney
,
T.
,
O’Neill
,
G.
,
Alexander
,
N.
,
Seymour
,
R.
,
Holmes
,
N.
,
Lopez
,
J. D.
,
Hill
,
R.
,
Boto
,
E.
,
Rea
,
M.
,
Roberts
,
G.
,
Leggett
,
J.
,
Bowtell
,
R.
,
Brookes
,
M. J.
,
Maguire
,
E.
,
Walker
,
M.
, &
Barnes
,
G.
(
2021
).
Magnetic field mapping and correction for moving OP-MEG
.
IEEE Trans Biomed Eng
,
69
,
528
536
. https://doi.org/10.1109/TBME.2021.3100770
Nolte
,
G.
(
2003
).
The magnetic lead field theorem in the quasi-static approximation and its use for magnetoencephalography forward calculation in realistic volume conductors
.
Phys Med Biol
,
48
,
3637
3652
. https://doi.org/10.1088/0031-9155/48/22/002
Nugent
,
A. C.
,
Benitez
Andonegui
, A.,
Holroyd
,
T.
, &
Robinson
,
S. E.
(
2022
).
On-scalp magnetocorticography with optically pumped magnetometers: Simulated performance in resolving simultaneous sources
.
Neuroimage: Reports
,
2
,
100093
. https://doi.org/10.1016/j.ynirp.2022.100093
Nurminen
,
J.
,
Taulu
,
S.
,
Nenonen
,
J.
,
Helle
,
L.
,
Simola
,
J.
, &
Ahonen
,
A.
(
2013
).
Improving MEG performance with additional tangential sensors
.
IEEE Trans Biomed Eng
,
60
,
2559
2566
. https://doi.org/10.1109/TBME.2013.2260541
Nurminen
,
J.
,
Taulu
,
S.
, &
Okada
,
Y.
(
2008
).
Effects of sensor calibration, balancing and parametrization on the signal space separation method
.
Phys Med Biol
,
53
,
1975
1987
. https://doi.org/10.1088/0031-9155/53/7/012
Oyama
,
D.
,
Adachi
,
Y.
,
Higuchi
,
M.
, &
Uehara
,
G.
(
2022
).
Calibration of a coil array geometry using an X-ray computed tomography
.
IEEE Trans Magn
,
58
,
1
5
. https://doi.org/10.1109/TMAG.2021.3080673
Pfeiffer
,
C.
,
Andersen
,
L. M.
,
Lundqvist
,
D.
,
Hämäläinen
,
M.
,
Schneiderman
,
J. F.
, &
Oostenveld
,
R.
(
2018
).
Localizing on-scalp MEG sensors using an array of magnetic dipole coils
.
PLoS One
,
13
,
e0191111
. https://doi.org/10.1371/journal.pone.0191111
Pfeiffer
,
C.
,
Ruffieux
,
S.
,
Andersen
,
L. M.
,
Kalabukhov
,
A.
,
Winkler
,
D.
,
Oostenveld
,
R.
,
Lundqvist
,
D.
, &
Schneiderman
,
J. F.
(
2020
).
On-scalp MEG sensor localization using magnetic dipole-like coils: A method for highly accurate co-registration
.
Neuroimage
,
212
,
116686
. https://doi.org/10.1016/j.neuroimage.2020.116686
Pfurtscheller
,
G.
, &
Lopes da Silva
,
F. H
. (
1999
).
Event-related EEG/MEG synchronization and desynchronization: Basic principles
.
Clin Neurophysiol
,
110
,
1842
1857
. https://doi.org/10.1016/S1388-2457(99)00141-8
Rampp
,
S.
,
Stefan
,
H.
,
Wu
,
X.
,
Kaltenhäuser
,
M.
,
Maess
,
B.
,
Schmitt
,
F. C.
,
Wolters
,
C. H.
,
Hamer
,
H.
,
Kasper
,
B. S.
,
Schwab
,
S.
,
Doerfler
,
A.
,
Blümcke
,
I.
,
Rössler
,
K.
, &
Buchfelder
,
M.
(
2019
).
Magnetoencephalography for epileptic focus localization in a series of 1000 cases
.
Brain
,
142
,
3059
3071
. https://doi.org/10.1093/brain/awz231
Rea
,
M.
,
Holmes
,
N.
,
Hill
,
R. M.
,
Boto
,
E.
,
Leggett
,
J.
,
Edwards
,
L. J.
,
Woolger
,
D.
,
Dawson
,
E.
,
Shah
,
V.
,
Osborne
,
J.
,
Bowtell
,
R.
, &
Brookes
,
M. J.
(
2021
).
Precision magnetic field modelling and control for wearable magnetoencephalography
.
Neuroimage
,
241
,
118401
. https://doi.org/10.1016/j.neuroimage.2021.118401
Rhodes
,
N.
,
Rea
,
M.
,
Boto
,
E.
,
Rier
,
L.
,
Shah
,
V.
,
Hill
,
R. M.
,
Osborne
,
J.
,
Doyle
,
C.
,
Holmes
,
N.
,
Coleman
,
S. C.
,
Mullinger
,
K.
,
Bowtell
,
R.
, &
Brookes
,
M. J.
(
2023
).
Measurement of frontal midline theta oscillations using OPM-MEG
.
Neuroimage
,
271
,
120024
. https://doi.org/10.1016/j.neuroimage.2023.120024
Rier
,
L.
,
Rhodes
,
N.
,
Pakenham
,
D.
,
Boto
,
E.
,
Holmes
,
N.
,
Hill
,
R. M.
,
Reina Rivero
,
G.
,
Shah
,
V.
,
Doyle
,
C.
,
Osborne
,
J.
,
Bowtell
,
R.
,
Taylor
,
M. J.
, &
Brookes
,
M. J.
(
2024
).
The neurodevelopmental trajectory of beta band oscillations: An OPM-MEG study
.
Elife
. https://doi.org/10.7554/elife.94561
Robinson
,
S.
, &
Vrba
,
J.
(
1998
).
Functional neuroimaging by synthetic aperture magnetometry
.
Recent Adv Biomagnet
,
302
305
. https://doi.org/10.1109/acssc.2000.910969
Roshen
,
W. A.
(
1990
).
Effect of finite thickness of magnetic substrate on planar inductors
.
IEEE Trans Magn
,
26
,
270
275
. https://doi.org/10.1109/20.50553
Safar
,
K.
,
Vandewouw
,
M. M.
,
Sato
,
J.
,
Devasagayam
,
J.
,
Hill
,
R. M.
,
Rea
,
M.
,
Brookes
,
M. J.
, &
Taylor
,
M. J.
(
2024
).
Using optically pumped magnetometers to replicate task-related responses in next generation magnetoencephalography
.
Sci Rep
,
14
,
6513
. https://doi.org/10.1038/s41598-024-56878-6
Schofield
,
H.
,
Boto
,
E.
,
Shah
,
V.
,
Hill
,
R. M.
,
Osborne
,
J.
,
Rea
,
M.
,
Doyle
,
C.
,
Holmes
,
N.
,
Bowtell
,
R.
,
Woolger
,
D.
, &
Brookes
,
M. J.
(
2023
).
Quantum enabled functional neuroimaging: The why and how of magnetoencephalography using optically pumped magnetometers
.
Contemp Phys
,
63
,
161
179
. https://doi.org/10.1080/00107514.2023.2182950
Schofield
,
H.
,
Hill
,
R. M.
,
Feys
,
O.
,
Holmes
,
N.
,
Osborne
,
J.
,
Doyle
,
C.
,
Bobela
,
D.
,
Corvilain
,
P.
,
Wens
,
V.
,
Rier
,
L.
,
Bowtell
,
R.
,
Ferez
,
M.
,
Mullinger
,
K. J.
,
Coleman
,
S.
,
Rhodes
,
N.
,
Rea
,
M.
,
Tanner
,
Z.
,
Boto
,
E.
,
de Tiège
,
X.
,
Shah
,
V.
, &
Brookes
,
M. J.
(
2024
).
A novel, robust, and portable platform for magnetoencephalography using optically-pumped magnetometers
.
Imaging Neuroscience
,
2
,
1
22
. https://doi.org/10.1162/imag_a_00283
Schwindt
,
P. D. D.
,
Knappe
,
S.
,
Shah
,
V.
,
Hollberg
,
L.
,
Kitching
,
J.
,
Liew
,
L. A.
, &
Moreland
,
J.
(
2004
).
Chip-scale atomic magnetometer
.
Appl Phys Lett
,
85
,
6409
6411
. https://doi.org/10.1063/1.1839274
Schwindt
,
P. D. D.
,
Lindseth
,
B.
,
Knappe
,
S.
,
Shah
,
V.
,
Kitching
,
J.
, &
Liew
,
L.-A.
(
2007
).
Chip-scale atomic magnetometer with improved sensitivity by use of the Mx technique
.
Appl Phys Lett
,
90
,
081102
. https://doi.org/10.1063/1.2709532
Shah
,
V.
,
Doyle
,
C.
, &
Osborne
,
J.
(
2020
).
Zero field parametric resonance magnetometer with triaxial sensitivity
.
US10775450B1
. https://doi.org/10.1117/12.2299197
Shah
,
V.
,
Knappe
,
S.
,
Schwindt
,
P. D. D.
, &
Kitching
,
J.
(
2007
).
Subpicotesla atomic magnetometry with a microfabricated vapour cell
.
Nat Photonics
,
1
,
649
652
. https://doi.org/10.1038/nphoton.2007.201
Shah
,
V. K.
, &
Wakai
,
R. T.
(
2013
).
A compact, high performance atomic magnetometer for biomedical applications
.
Phys Med Biol
,
58
,
8153
8161
. https://doi.org/10.1088/0031-9155/58/22/8153
Taulu
,
S.
, &
Kajola
,
M.
(
2005
).
Presentation of electromagnetic multichannel data: The signal space separation method
.
J Appl Phys
,
97
,
124905
. https://doi.org/10.1063/1.1935742
Taulu
,
S.
, &
Simola
,
J.
(
2006
).
Spatiotemporal signal space separation method for rejecting nearby interference in MEG measurements
.
Phys Med Biol
,
51
,
1759
1768
. https://doi.org/10.1088/0031-9155/51/7/008
Taulu
,
S.
,
Simola
,
J.
, &
Kajola
,
M.
(
2005
).
Applications of the signal space separation method
.
IEEE Trans Signal Process
,
53
,
3359
3372
. https://doi.org/10.1109/TSP.2005.853302
Taylor
,
M. J.
,
George
,
N.
, &
Ducorps
,
A.
(
2001
).
Magnetoencephalographic evidence of early processing of direction of gaze in humans
.
Neurosci Lett
,
316
,
173
177
. https://doi.org/10.1016/S0304-3940(01)02378-3
Tierney
,
T. M.
,
Alexander
,
N.
,
Mellor
,
S.
,
Holmes
,
N.
,
Seymour
,
R.
,
O’Neill
,
G. C.
,
Maguire
,
E. A.
, &
Barnes
,
G. R.
(
2021
).
Modelling optically pumped magnetometer interference in MEG as a spatially homogeneous magnetic field
.
Neuroimage
,
244
,
118484
. https://doi.org/10.1016/j.neuroimage.2021.118484
Tierney
,
T. M.
,
Barnes
,
G. R.
,
Seedat
,
Z.
,
Pier
,
K. S.
, &
Mellor
,
S.
(
2024
).
Adaptive multipole models of optically pumped magnetometer data
.
Hum Brain Mapp
,
45
,
e26596
. https://doi.org/10.1002/hbm.26596
Tierney
,
T. M.
,
Holmes
,
N.
,
Mellor
,
S.
,
López
,
J. D.
,
Roberts
,
G.
,
Hill
,
R. M.
,
Boto
,
E.
,
Leggett
,
J.
,
Shah
,
V.
,
Brookes
,
M. J.
,
Bowtell
,
R.
, &
Barnes
,
G. R.
(
2019
).
Optically pumped magnetometers: From quantum origins to multi-channel magnetoencephalography
.
Neuroimage
,
199
,
598
608
. https://doi.org/10.1016/j.neuroimage.2019.05.063
Tierney
,
T. M.
,
Mellor
,
S.
,
O’Neill
,
G. C.
,
Holmes
,
N.
,
Boto
,
E.
,
Roberts
,
G.
,
Hill
,
R. M.
,
Leggett
,
J.
,
Bowtell
,
R.
,
Brookes
,
M. J.
, &
Barnes
,
G. R.
(
2020
).
Pragmatic spatial sampling for wearable MEG arrays
.
Sci Rep
,
10
,
21609
. https://doi.org/10.1038/s41598-020-77589-8
Tierney
,
T. M.
,
Mellor
,
S.
,
O’Neill
,
G. C.
,
Timms
,
R. C.
, &
Barnes
,
G. R.
(
2022
).
Spherical harmonic based noise rejection and neuronal sampling with multi-axis OPMs
.
Neuroimage
,
258
,
119338
. https://doi.org/10.1016/j.neuroimage.2022.119338
Tzourio-Mazoyer
,
N.
,
Landeau
,
B.
,
Papathanassiou
,
D.
,
Crivello
,
F.
,
Etard
,
O.
,
Delcroix
,
N.
,
Mazoyer
,
B.
, &
Joliot
,
M.
(
2002
).
Automated anatomical labeling of activations in SPM using a macroscopic anatomical parcellation of the MNI MRI single-subject brain
.
Neuroimage
,
15
,
273
289
. https://doi.org/10.1006/nimg.2001.0978
Vrba
,
J.
, &
Robinson
,
S. E.
(
2001
).
Signal processing in magnetoencephalography
.
Methods
,
25
,
249
271
. https://doi.org/10.1006/meth.2001.1238
Zetter
,
R.
,
Iivanainen
,
J.
,
Stenroos
,
M.
, &
Parkkonen
,
L.
(
2018
).
Requirements for coregistration accuracy in on-scalp MEG
.
Brain Topogr
,
31
,
931
948
. https://doi.org/10.1007/s10548-018-0656-5
Zetter
,
R.
,
J. Mäkinen
,
A.
,
Iivanainen
,
J.
,
Zevenhoven
,
K. C. J.
,
Ilmoniemi
,
R. J.
, &
Parkkonen
,
L.
(
2020
).
Magnetic field modeling with surface currents. Part II. Implementation and usage of bfieldtools
.
J Appl Phys
,
128
,
063905
. https://doi.org/10.1063/5.0016087

Appendix.  Sensor Localisation Accuracy with Distance from the HALO

Appendix Figure A1a shows Euclidean distance between the HALO-derived and the CAD-derived sensor locations, following application of the ICP algorithm (y-axis), plotted against the distance between the sensor and the centre of the HALO (x-axis). Appendix Figure A1b shows the equivalent data for the MC-derived sensor locations. In both cases, data from all experimental runs are overlaid. Both plots also show a line of best-fit, with an R-squared value of 0.18 for the HALO and 0.0142 for the MC. The relationship in Appendix Figure A1a suggests that sensors distal to the HALO tended to localise more poorly than proximal sensors. As would be expected, this relationship is absent for the MC-derived sensor locations.

Appendix Fig. A1

(a) Absolute distances between the CAD and HALO sensor positions following application of the ICP algorithm, plotted against distance from the sensor to the centre of the HALO. (b) Equivalent data for the MC. Note a reduction in agreement with distance for HALO calibration.

Appendix Fig. A1

(a) Absolute distances between the CAD and HALO sensor positions following application of the ICP algorithm, plotted against distance from the sensor to the centre of the HALO. (b) Equivalent data for the MC. Note a reduction in agreement with distance for HALO calibration.

Close modal

Author notes

*

Equal contribution from multiple authors

This is an open-access article distributed under the terms of the Creative Commons Attribution 4.0 International (CC BY 4.0) license, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. For a full description of the license, please visit https://creativecommons.org/licenses/by/4.0/legalcode.